B1 field, SAR, and SNR comparisons for birdcage, TEM, and microstrip coils at 7T.

Chunsheng Wang1, Gary X Shen.   

Abstract

PURPOSE: To compare the performance of birdcage, transverse electromagnetic (TEM) and microstrip volume coils at 7T under the same geometric conditions.
MATERIALS AND METHODS: Birdcage, TEM, and microstrip coils are modeled with the same dimensions. The finite difference time domain (FDTD) method is adopted to calculate the electromagnetic fields of the coils. Further, B(1) field, specific absorption rate (SAR) and signal-to-noise ratio (SNR) are calculated for these coils.
RESULTS: In the unloaded case, within the central axial plane, the variation of B(1) field magnitude over 18-cm distance is about 15% for the birdcage coil, 23% for the TEM coil, and 38% for the microstrip coil. In the loaded case, the percentages of the samples on the central axial plane, which have B(1) field magnitude within +/-20% of the average B(1) field magnitude, are about 57% for the birdcage, 72% for the TEM, and 59% for the microstrip coil. Average SAR levels are 11.4% and 42.9% higher in the birdcage than those in the TEM and microstrip coils, respectively. The average relative SNR on the central axial plane for the shielded birdcage, TEM, and microstrip coils are 1, 1.07, and 1.48, respectively.
CONCLUSION: The birdcage coil has the best unloaded B(1) field homogeneity, and the TEM coil has the best loaded B(1) field homogeneity and the lowest radiation loss; while the microstrip coil is better in SAR and SNR at 7T than the birdcage and TEM coils.

Mesh:

Year:  2006        PMID: 16786582     DOI: 10.1002/jmri.20635

Source DB:  PubMed          Journal:  J Magn Reson Imaging        ISSN: 1053-1807            Impact factor:   4.813


  7 in total

1.  Comparison of three commercially available radio frequency coils for human brain imaging at 3 Tesla.

Authors:  Ralf Mekle; Wietske van der Zwaag; Andreas Joosten; Rolf Gruetter
Journal:  MAGMA       Date:  2008-01-10       Impact factor: 2.310

2.  A loop resonator for slice-selective in vivo EPR imaging in rats.

Authors:  Hiroshi Hirata; Guanglong He; Yuanmu Deng; Ildar Salikhov; Sergey Petryakov; Jay L Zweier
Journal:  J Magn Reson       Date:  2007-10-30       Impact factor: 2.229

3.  Evaluation of a New 1H/31P Dual-Tuned Birdcage Coil for 31P Spectroscopy.

Authors:  Wm Potter; L Wang; Kk McCully; Q Zhao
Journal:  Concepts Magn Reson Part B Magn Reson Eng       Date:  2013-08-01       Impact factor: 1.176

4.  Quantitative study of liver magnetic resonance spectroscopy quality at 3T using body and phased array coils with physical analysis and clinical evaluation.

Authors:  Li Xu; Shiyong Gu; Qianjin Feng; Changhong Liang; Sherman Xuegang Xin
Journal:  PLoS One       Date:  2015-04-16       Impact factor: 3.240

5.  Computational and experimental evaluation of the Tic-Tac-Toe RF coil for 7 Tesla MRI.

Authors:  Narayanan Krishnamurthy; Tales Santini; Sossena Wood; Junghwan Kim; Tiejun Zhao; Howard J Aizenstein; Tamer S Ibrahim
Journal:  PLoS One       Date:  2019-01-10       Impact factor: 3.240

6.  Optimized (31)P MRS in the human brain at 7 T with a dedicated RF coil setup.

Authors:  Bart L van de Bank; Stephan Orzada; Frits Smits; Miriam W Lagemaat; Christopher T Rodgers; Andreas K Bitz; Tom W J Scheenen
Journal:  NMR Biomed       Date:  2015-10-07       Impact factor: 4.044

7.  Extended RF shimming: Sequence-level parallel transmission optimization applied to steady-state free precession MRI of the heart.

Authors:  Arian Beqiri; Anthony N Price; Francesco Padormo; Joseph V Hajnal; Shaihan J Malik
Journal:  NMR Biomed       Date:  2017-02-14       Impact factor: 4.044

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.