Incidence of atrial tachyarrhythmias in pacemaker patients: results from the Balanced Evaluation of Atrial Tachyarrhythmias in Stimulated patients (BEATS) study.

Carsten W Israel1, Horst Neubauer, Hans-Georg Olbrich, Wolfgang Hartung, Sven Treusch, Stefan H Hohnloser.   

Abstract

OBJECTIVES: The prospective Balanced Evaluation of Atrial Tachyarrhythmias in Stimulated patients (BEATS) study compared atrial tachyarrhythmia (AT) detection by means of serial ECG recordings versus device detection.
BACKGROUND: The annual incidence of AT in patients with dual-chamber pacemakers may be significantly underestimated based on ECG and Holter recordings.
METHODS: A DDD(R) device capable of AT-triggered dual-channel electrogram (EGM) storage was implanted in 254 patients (70 +/- 11 years, 159 men) with a class I pacing indication. Patients were seen at 6, 26, and 52 weeks after pacemaker implantation. At all visits, symptoms were checked, surface ECGs were recorded including a 24-hour Holter recording at 6 weeks, and the pacemakers were interrogated. Primary study endpoint was AT documentation by surface ECG/Holter versus stored EGMs. Secondary endpoints consisted of the association between patients' symptoms and AT documentation, and of the AT incidence depending on pacing indication and a history of AT.
RESULTS: ATs were documented by ECG/Holter recordings in 37 patients (15%) and by stored EGMs in 137 patients (54%) (P < 0.0001). Symptoms were absent in 108 of 137 patients (79%) with device-documented AT but present in 70 of 117 patients (60%) without AT documentation. AT documentation was more frequent in patients with a history of AT but not in patients with sinus node compared to AV node disease.
CONCLUSION: ATs occur in pacemaker patients significantly more frequently than estimated by ECG/Holter recordings. Only the analysis of device-stored EGMs allows reliable assessment of the AT burden.

Entities:  

Mesh:

Year:  2006        PMID: 16784423     DOI: 10.1111/j.1540-8159.2006.00405.x

Source DB:  PubMed          Journal:  Pacing Clin Electrophysiol        ISSN: 0147-8389            Impact factor:   1.976


  5 in total

Review 1.  Electrical cardioversion of patients with implanted pacemaker or cardioverter-defibrillator: results of a survey of german centers and systematic review of the literature.

Authors:  J Lüker; A Sultan; T Plenge; J van den Bruck; C-H Heeger; S Meyer; K Mischke; R R Tilz; D Vollmann; G Nölker; B Schäffer; S Willems; D Steven
Journal:  Clin Res Cardiol       Date:  2017-11-18       Impact factor: 5.460

2.  [Stored electrograms in pacemakers and ICDs].

Authors:  C W Israel
Journal:  Herzschrittmacherther Elektrophysiol       Date:  2010-03

3.  Pacemaker detected prolonged atrial high rate episodes - Incidence, predictors and implications; a retrospective observational study.

Authors:  Pratheesh George Mathen; David Chase
Journal:  J Saudi Heart Assoc       Date:  2020-05-16

4.  Evaluation Of Two Thromboembolism (TE) Risk Methods And Oral Anticoagulation Use Among Cardiac Device Patients With Atrial Tachyarrhythmias (AT).

Authors:  Heather S Kennedy; Theresa M Wadas; Gwenda Stewart Reyes; Manoj M Panday
Journal:  J Atr Fibrillation       Date:  2014-12-31

5.  Relevance of Monitoring Atrial Fibrillation in Clinical Practice.

Authors:  Thorsten Lewalter; Giuseppe Boriani
Journal:  Arrhythm Electrophysiol Rev       Date:  2012-09
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.