Corneal endothelium evaluation with 2 noncontact specular microscopes and their semiautomated methods of analysis.

Ugo de Sanctis1, Federica Machetta, Luca Razzano, Paola Dalmasso, Federico M Grignolo.   

Abstract

PURPOSE: Noncontact specular microscopes and semiautomated methods of endothelial analysis are widely used in the clinical practice and for research purposes. In this study, the interexaminer reproducibility between two noncontact specular microscopes using different semiautomated methods was assessed in normal corneas.
METHODS: Average cell size (ACS), endothelial cell density (ECD), coefficient of variation (CV), and hexagonality values were independently calculated by two examiners with the Topcon SP 2000P Image-NET retraced method and the Konan CC7000 center method in 49 normal subjects. Interexaminer reproducibility and correlation between instruments/methods were assessed through 95% limits of agreement, intra-class correlation coefficient (ICC), Pearson correlation coefficient, and Spearman rank test.
RESULTS: Interexaminer reproducibility was good for ACS and ECD measurements (ICC > 0.85) but weak for CV and hexagonality (ICC < 0.65). Significantly higher ACS and lower ECD values (P < 0.05) were obtained with the Topcon in comparison with the Konan method by both examiners. ACS and ECD mean differences were respectively, +26 microm2 and -184 cell/mm2 for examiner 1 and +36 microm2 and -228 cell/mm2 for examiner 2. No significant difference (P > 0.05) between instruments/methods was found in CV or hexagonality values; however, the correlation for these parameters was low (r < 0.20).
CONCLUSIONS: Either instrument or method allows reproducible and accurate ACS and ECD measurements on normal corneas, but the difference between the two instruments is systematic and significant. The difference found were not clinically meaningful however for research/longitudinal study purposes, the data collected with the two systems cannot be used interchangeably.

Entities:  

Mesh:

Year:  2006        PMID: 16783136     DOI: 10.1097/01.ico.0000214217.83119.af

Source DB:  PubMed          Journal:  Cornea        ISSN: 0277-3740            Impact factor:   2.651


  9 in total

1.  Measurements of central corneal thickness and endothelial parameters with three different non-contact specular microscopy devices.

Authors:  Ozgur Cakici; Remzi Karadag; Huseyin Bayramlar; Efe Koyun
Journal:  Int Ophthalmol       Date:  2016-05-24       Impact factor: 2.031

2.  Comparison of semi-automated center-dot and fully automated endothelial cell analyses from specular microscopy images.

Authors:  Sachiko Maruoka; Shunsuke Nakakura; Naoko Matsuo; Kayo Yoshitomi; Chikako Katakami; Hitoshi Tabuchi; Taiichiro Chikama; Yoshiaki Kiuchi
Journal:  Int Ophthalmol       Date:  2017-10-30       Impact factor: 2.031

3.  Two single descriptors of endothelial polymegethism and pleomorphism.

Authors:  Jose Manuel González-Méijome; Jorge Jorge; Antonio Queirós; Sofia C Peixoto-de-Matos; Manuel A Parafita
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2010-03-12       Impact factor: 3.117

4.  Representation of Central Endothelial Cell Density by Analysis of Single Best Specular Microscopy Image Regardless of Cell Size Variance.

Authors:  Yue Shi; Jianyan Huang; Elmira Baghdasaryan; Ping Huang; Xiwen Huang; Srinivas R Sadda; Olivia L Lee
Journal:  Transl Vis Sci Technol       Date:  2019-05-29       Impact factor: 3.283

5.  Retropupillary Iris Fixation of an Artisan Myopia Lens for Intraocular Lens Dislocation and Aphakia in Eyes with Extremely High Myopia: A Case Series and a Literature Review.

Authors:  Tae Keun Yoo; Seung Min Lee; Hansang Lee; Eun Young Choi; Min Kim
Journal:  Ophthalmol Ther       Date:  2022-03-15

6.  Corneal endothelial cell density in healthy Caucasian population.

Authors:  Rahmi Duman; Mediha Tok Çevik; Sadık Görkem Çevik; Reşat Duman; İrfan Perente
Journal:  Saudi J Ophthalmol       Date:  2016-11-02

7.  Corneal Endothelial Morphology and Thickness Changes in Patients with Gout

Authors:  Pınar Kösekahya; Cemile Üçgül Atılgan; Kadir Gökhan Atılgan; Mustafa Koç; Kemal Tekin; Mehtap Çağlayan; Yasin Şakir Göker
Journal:  Turk J Ophthalmol       Date:  2019-09-03

8.  Improved Interchangeability with Different Corneal Specular Microscopes for Quantitative Endothelial Cell Analysis.

Authors:  Gwyneth A van Rijn; C Jasper F Wijnen; Bart Th van Dooren; Yanny Yy Cheng; Jan-Willem M Beenakker; Gregorius Pm Luyten
Journal:  Clin Ophthalmol       Date:  2020-01-13

9.  Comparison of corneal endothelial cell measurements by two non-contact specular microscopes.

Authors:  Laura Gasser; Thomas Reinhard; Daniel Böhringer
Journal:  BMC Ophthalmol       Date:  2015-07-29       Impact factor: 2.209

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.