Skin entrance radiation dose in an interventional radiology procedure.

Robert Y L Chu1, George Thomas, Feroz Maqbool.   

Abstract

Monitoring of skin entrance radiation exposure in lengthy interventional procedures has been recommended because of the potential for skin injury. Fluoroscopy duration and dose-area product (DAP) are readily available real-time measurements. It would be of interest to study the correlation of these parameters and skin entrance radiation. Twenty neurological interventional procedures performed through the aortic arch were monitored. Two pieces of GafChromic XR Type R film were placed between the patient and the examination table. An observer recorded the fluoroscopy duration and DAP for each phase of the procedure. Each film was scanned post-procedure in RBG mode, and then the image was analyzed for peak skin entrance radiation dose (in air kerma). All DAP values were corrected according to a calibration with an ion chamber. With the DAP values for the respective phases of a procedure, the effective dose in a Reference Man was calculated. For these twenty cases, the means and standard deviations were 17.2+/-6.4 min for x ray on-time, 256+/-65 Gy cm (-2) for DAP, 94+/-34 cGy for peak skin entrance dose in air kerma, and 19.2+/-5.0 mSv for effective dose, respectively. The peak skin entrance dose was correlated to fluoroscopy duration, DAP, and effective dose with the r(2)-values of 0.48, 0.46, and 0.09, respectively. The correlation with DAP or fluoroscopy duration was not sufficiently strong to infer skin entrance dose from either of these parameters. Therefore, skin entrance dose should be determined directly.

Entities:  

Mesh:

Year:  2006        PMID: 16775479     DOI: 10.1097/01.HP.0000198784.94210.1f

Source DB:  PubMed          Journal:  Health Phys        ISSN: 0017-9078            Impact factor:   1.316


  2 in total

1.  How to set up and apply reference levels in fluoroscopy at a national level.

Authors:  A Aroua; H Rickli; J-C Stauffer; P Schnyder; P R Trueb; J-F Valley; P Vock; F R Verdun
Journal:  Eur Radiol       Date:  2006-10-27       Impact factor: 5.315

2.  Entrance skin dose measured with MOSFETs in children undergoing interventional radiology procedures.

Authors:  Diana Glennie; Bairbre L Connolly; Christopher Gordon
Journal:  Pediatr Radiol       Date:  2008-08-19
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.