Anthropometric precision and accuracy of digital three-dimensional photogrammetry: comparing the Genex and 3dMD imaging systems with one another and with direct anthropometry.

Seth M Weinberg1, Sybill Naidoo, Daniel P Govier, Rick A Martin, Alex A Kane, Mary L Marazita.   

Abstract

A variety of commercially available three-dimensional (3D) surface imaging systems are currently in use by craniofacial specialists. Little is known, however, about how measurement data generated from alternative 3D systems compare, specifically in terms of accuracy and precision. The purpose of this study was to compare anthropometric measurements obtained by way of two different digital 3D photogrammetry systems (Genex and 3dMD) as well as direct anthropometry and to evaluate intraobserver precision across these three methods. On a sample of 18 mannequin heads, 12 linear distances were measured twice by each method. A two-factor repeated measures analysis of variance was used to test simultaneously for mean differences in precision across methods. Additional descriptive statistics (e.g., technical error of measurement [TEM]) were used to quantify measurement error magnitude. Statistically significant (P < 0.05) mean differences were observed across methods for nine anthropometric variables; however, the magnitude of these differences was consistently at the submillimeter level. No significant differences were noted for precision. Moreover, the magnitude of imprecision was determined to be very small, with TEM scores well under 1 mm, and intraclass correlation coefficients ranging from 0.98 to 1. Results indicate that overall mean differences across these three methods were small enough to be of little practical importance. In terms of intraobserver precision, all methods fared equally well. This study is the first attempt to simultaneously compare 3D surface imaging systems directly with one another and with traditional anthropometry. Results suggest that craniofacial surface data obtained by way of alternative 3D photogrammetric systems can be combined or compared statistically.

Entities:  

Mesh:

Year:  2006        PMID: 16770184     DOI: 10.1097/00001665-200605000-00015

Source DB:  PubMed          Journal:  J Craniofac Surg        ISSN: 1049-2275            Impact factor:   1.046


  69 in total

1.  Reliability of periocular anthropometry using three-dimensional digital stereophotogrammetry.

Authors:  Yongwei Guo; Alexander C Rokohl; Friederike Schaub; Xiaoyi Hou; Jinhua Liu; Yue Ruan; Renbing Jia; Konrad R Koch; Ludwig M Heindl
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2019-08-12       Impact factor: 3.117

2.  Three-dimensional prediction of nose morphology in Chinese young adults: a pilot study combining cone-beam computed tomography and 3dMD photogrammetry system.

Authors:  Guang Chu; Jia-Min Zhao; Meng-Qi Han; Qing-Nan Mou; Ling-Ling Ji; Hong Zhou; Teng Chen; Shao-Yi Du; Yu-Cheng Guo
Journal:  Int J Legal Med       Date:  2020-07-09       Impact factor: 2.686

3.  Prenatal sex hormones, digit ratio, and face shape in adult males.

Authors:  S M Weinberg; T E Parsons; Z D Raffensperger; M L Marazita
Journal:  Orthod Craniofac Res       Date:  2014-09-25       Impact factor: 1.826

4.  Head orthesis therapy in infants with unilateral positional plagiocephaly: an interdisciplinary approach to broadening the range of orthodontic treatment.

Authors:  P Meyer-Marcotty; H Böhm; C Linz; F Kunz; N Keil; A Stellzig-Eisenhauer; T Schweitzer
Journal:  J Orofac Orthop       Date:  2012-03-04       Impact factor: 1.938

5.  Heritability of Face Shape in Twins: A Preliminary Study using 3D Stereophotogrammetry and Geometric Morphometrics.

Authors:  Seth M Weinberg; Trish E Parsons; Mary L Marazita; Brion S Maher
Journal:  Dent 3000       Date:  2013

6.  3D stereophotogrammetry versus traditional craniofacial anthropometry: Comparing measurements from the 3D facial norms database to Farkas's North American norms.

Authors:  Seth M Weinberg
Journal:  Am J Orthod Dentofacial Orthop       Date:  2019-05       Impact factor: 2.650

7.  A 3D anthropometric analysis of the orolabial region in Chinese young adults.

Authors:  Yasas S N Jayaratne; Curtis K Deutsch; Roger A Zwahlen
Journal:  Br J Oral Maxillofac Surg       Date:  2013-10-03       Impact factor: 1.651

8.  Stereophotogrammetry-based facial depth measurements: a novel method for quantifying facial projection.

Authors:  Yasas S N Jayaratne; Curtis K Deutsch; Roger A Zwahlen
Journal:  Surg Innov       Date:  2013-02-18       Impact factor: 2.058

9.  Repair of bilateral cleft lip and its variants.

Authors:  John B Mulliken
Journal:  Indian J Plast Surg       Date:  2009-10

10.  Metric precision via soft-tissue landmarks in three-dimensional structured-light scans of human faces.

Authors:  M Fink; J Medelnik; K Strobel; U Hirschfelder; E Hofmann
Journal:  J Orofac Orthop       Date:  2014-03-02       Impact factor: 1.938

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.