Relationship between locking-bolt torque and load pre-tension in the Ilizarov frame.

N A Osei1, B M Bradley, P Culpan, J B Mitchell, M Barry, K E Tanner.   

Abstract

The wire-bolt interface in an Ilizarov frame has been mechanically tested. The optimal torque to be applied to the frame locking-bolts during physiological loading has been defined. The set-up configuration was as is used clinically except a copper tube was used to simulate bone. The force-displacement curves of the Ilizarov wires are not altered by locking-bolt torque. The force in the bone model at which pre-tension is lost increases as the locking-bolts are tightened to 14 Nm torque, but decreases if torque exceeds 14 Nm. Thus, 14 Nm is the optimal locking-bolt torque in frame. The relationship between pre-tension versus load for different locking-bolt torques arises because at low and high clamping torques poor wire holding and plastic deformation respectively occur. Wire damage was seen under light and electron microscopy. Clinically, over or under-tightening locking-bolts will cause loss of pre-tension, reduction in frame stiffness and excessive movement at the fracture site, which may be associated with delayed union.

Entities:  

Mesh:

Year:  2006        PMID: 16769071     DOI: 10.1016/j.injury.2006.02.019

Source DB:  PubMed          Journal:  Injury        ISSN: 0020-1383            Impact factor:   2.586


  3 in total

1.  Towards a solution of the wires' slippage problem of the Ilizarov external fixator.

Authors:  C Bairaktari; G Athanassiou; E Panagiotopoulos; D Deligianni
Journal:  Eur J Orthop Surg Traumatol       Date:  2014-08-20

2.  Direct and indirect loading of the Ilizarov external fixator: the effect on the interfragmentary movements and compressive loads.

Authors:  Jan Gessmann; Hinnerk Baecker; Birger Jettkant; Gert Muhr; Dominik Seybold
Journal:  Strategies Trauma Limb Reconstr       Date:  2011-02-11

3.  The influence of a weight-bearing platform on the mechanical behavior of two Ilizarov ring fixators: tensioned wires vs. half-pins.

Authors:  Jan Gessmann; Mustafa Citak; Birger Jettkant; Thomas A Schildhauer; Dominik Seybold
Journal:  J Orthop Surg Res       Date:  2011-12-12       Impact factor: 2.359

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.