| Literature DB >> 16769071 |
N A Osei1, B M Bradley, P Culpan, J B Mitchell, M Barry, K E Tanner.
Abstract
The wire-bolt interface in an Ilizarov frame has been mechanically tested. The optimal torque to be applied to the frame locking-bolts during physiological loading has been defined. The set-up configuration was as is used clinically except a copper tube was used to simulate bone. The force-displacement curves of the Ilizarov wires are not altered by locking-bolt torque. The force in the bone model at which pre-tension is lost increases as the locking-bolts are tightened to 14 Nm torque, but decreases if torque exceeds 14 Nm. Thus, 14 Nm is the optimal locking-bolt torque in frame. The relationship between pre-tension versus load for different locking-bolt torques arises because at low and high clamping torques poor wire holding and plastic deformation respectively occur. Wire damage was seen under light and electron microscopy. Clinically, over or under-tightening locking-bolts will cause loss of pre-tension, reduction in frame stiffness and excessive movement at the fracture site, which may be associated with delayed union.Entities:
Mesh:
Year: 2006 PMID: 16769071 DOI: 10.1016/j.injury.2006.02.019
Source DB: PubMed Journal: Injury ISSN: 0020-1383 Impact factor: 2.586