Abstract

BACKGROUND: Nonpermanent caval filters are placed in critical thromboembolic situations in which anticoagulation therapy is transiently contraindicated, ineffective, or the source of complications. The purpose of this study was to assess the safety and effectiveness of a second-generation long-duration temporary caval filter in these situations and compare its utility with that of other temporary filters.
METHODS: This prospective study, including patients who underwent temporary caval filtration with the Tempofilter II, was conducted in nine European centers. All filters were successfully implanted. The filter was removed when the indication for caval filtration ceased.
RESULTS: A total of 104 filters were inserted in 103 patients with an average age of 60 +/- 15.5 years (range, 22-92 years). Most patients (85%) had pulmonary embolism, deep venous thrombosis, or both. The main indications for caval filter placement were complications of or contraindications to anticoagulation therapy (n = 85; 82.5%) or for ineffectiveness of anticoagulation therapy (n = 12; 11.7%). The average duration of implantation was 29.5 +/- 14.0 days (range, 2-86 days). One filter migrated in the right atrium, followed by pulmonary embolism. No other case of pulmonary embolism or of infectious or mechanical complications related to the filter was observed. Thrombus was trapped within the filter in 24 cases (23.3%). All filters but one were removed, regardless of whether thrombus had been trapped. Retrieval was always successful after implantation periods up to 12 weeks. In 16 cases (15.5%), the filter was replaced by a permanent filter.
CONCLUSIONS: The Tempofilter II is safe, effective, and useful in critical thromboembolic situations. It offers a valuable alternative to retrievable optional filters.

Entities:  

Mesh:

Year:  2006        PMID: 16765244     DOI: 10.1016/j.jvs.2006.02.004

Source DB:  PubMed          Journal:  J Vasc Surg        ISSN: 0741-5214            Impact factor:   4.268


  9 in total

1.  Surgical removal of a tilted temporary inferior vena cava filter incorporated with a large thrombus: report of a case.

Authors:  Takaki Sugimoto; Takuo Nomura; Takashi Kitade
Journal:  Surg Today       Date:  2008-02-29       Impact factor: 2.549

2.  Pulmonary embolism after IVC filter.

Authors:  M K Urban; K Jules-Elysee; C R MacKenzie
Journal:  HSS J       Date:  2007-12-21

3.  Tempofilter II implantation in patients with lower extremity fractures and proximal deep vein thrombosis.

Authors:  Wei Jia; Jianlong Liu; Xuan Tian; Peng Jiang
Journal:  Diagn Interv Radiol       Date:  2014 May-Jun       Impact factor: 2.630

4.  Rates and predictors of plans for inferior vena cava filter retrieval in hospitalized patients.

Authors:  John F Mission; Robert K Kerlan; Justin H Tan; Margaret C Fang
Journal:  J Gen Intern Med       Date:  2010-01-20       Impact factor: 5.128

Review 5.  Using inferior vena cava filters to prevent pulmonary embolism.

Authors:  John Chung; Richard J T Owen
Journal:  Can Fam Physician       Date:  2008-01       Impact factor: 3.275

6.  Bedside placement of removable vena cava filters guided by intravascular ultrasound in the critically injured.

Authors:  Konstantinos Spaniolas; George C Velmahos; Christopher Kwolek; Alice Gervasini; Marc De Moya; Hasan B Alam
Journal:  World J Surg       Date:  2008-07       Impact factor: 3.352

7.  Retrograde Tempofilter II™ placement within the superior vena cava in a patient with acute upper extremity deep venous thrombosis: the filter stands on its head.

Authors:  Nam Yeol Yim; Nam Kyu Chang; Jae Hoon Lim; Jae Kyu Kim
Journal:  Korean J Radiol       Date:  2011-01-03       Impact factor: 3.500

8.  Fracture of a Tempofilter II: an initial case report.

Authors:  Hyung Jun Kim; Nam Kyu Chang; Jae Hoon Lim; Jae Kyu Kim
Journal:  Korean J Radiol       Date:  2011-08-24       Impact factor: 3.500

9.  Caval filters in intensive care: a retrospective study.

Authors:  F Ferraro; T L Di Gennaro; A Torino; J Petruzzi; A d'Elia; P Fusco; R Marfella; B Lettieri
Journal:  Drug Des Devel Ther       Date:  2014-11-06       Impact factor: 4.162

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.