PURPOSE: The aim of the present study was to compare the mechanical behavior of some devices used for femoral fixation of doubled hamstring tendon graft in anterior cruciate ligament (ACL) reconstruction when included in a graft fixation complex (GFC).
METHODS: An ACL reconstruction was performed on 90 porcine knees. The graft used was the doubled lateral extensor of toes (DLET). Nine different femoral fixation devices were tested and classified according to their fixation mechanism: compression (Bioscrew and RCI screw); expansion (Rigidfix); cortical suspension (Ligament Anchor, EndoButton-CL, and Swing Bridge); cancellous suspension (Linx-HT); and cortical-cancellous suspension (Transfix and Bio-Transfix). All GFC were subjected to a cyclic loading test, then to a load-to-failure test. Graft elongation after 1,000 load cycles, failure load, and stiffness were calculated for each device.
RESULTS: Regarding graft elongation, Bioscrew and RCI screws showed the highest mean values. All the other GFC showed no significant differences between them when delta elongation (elongation(1,000 cycles) - elongation(20 cycles)) was considered. For failure load, the highest mean values were observed for Bio-Transfix, Transfix, and Swing Bridge; a homogeneous subset with the lowest mean values was formed by Ligament Anchor, RCI screw, Bioscrew, and Linx-HT. For stiffness, the greatest values were observed for Bio-Transfix, Transfix, and Swing Bridge; all other groups showed no significant differences between them.
CONCLUSIONS: Cortical-cancellous suspension fixation seemed to offer the best and most predictable results in terms of elongation, fixation strength, and stiffness. For both compression and suspension, the weakest fixation was attained with cancellous fixation devices. Cortical suspension devices showed a greatly variable mechanical behavior, according to their design. CLINICAL RELEVANCE: Transcondylar devices offer the best structural properties for femoral fixation of doubled hamstring tendon graft in ACL reconstruction.

Entities:  

Mesh:

Year:  2006        PMID: 16762706     DOI: 10.1016/j.arthro.2006.04.082

Source DB:  PubMed          Journal:  Arthroscopy        ISSN: 0749-8063            Impact factor:   4.772


  66 in total

1.  Anterior cruciate ligament reconstruction using biodegradable transfemoral fixation at 5-year follow-up: clinical and magnetic resonance imaging evaluation.

Authors:  Sven Nebelung; Gregor Deitmer; Rolf Gebing; Frank Reichwein; Wolfgang Nebelung
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2012-03-06       Impact factor: 4.342

2.  The fixation strength of a novel ACL soft-tissue graft fixation device compared with conventional interference screws: a biomechanical study in vitro.

Authors:  Camilla Halewood; Michael T Hirschmann; Simon Newman; Jaffar Hleihil; Gershon Chaimski; Andrew A Amis
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2010-09-14       Impact factor: 4.342

3.  What is the best femoral fixation of hamstring autografts in anterior cruciate ligament reconstruction?: a meta-analysis.

Authors:  Alexis Colvin; Charu Sharma; Michael Parides; Jonathan Glashow
Journal:  Clin Orthop Relat Res       Date:  2011-04       Impact factor: 4.176

4.  Breakage of bioabsorbable femoral transfix screw after arthroscopic anterior cruciate ligament repair: a report of two cases.

Authors:  A Raviraj; Ashish Anand; K Santosh
Journal:  Eur J Orthop Surg Traumatol       Date:  2012-02-04

5.  Transtibial double bundle posterior cruciate ligament reconstruction using TransFix tibial fixation.

Authors:  Yong Seuk Lee; Jin Hwan Ahn; Young Bok Jung; Joon Ho Wang; Jae Chul Yoo; Ho Joong Jung; Bun Jung Kang
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2007-02-16       Impact factor: 4.342

6.  Double-bundle anterior cruciate ligament reconstruction using two different suspensory femoral fixation: a technical note.

Authors:  Yong Seuk Lee; Sung Kon Kim; Jung Ho Park; Jong Woong Park; Joon Ho Wang; Young Bok Jung; Jin Hwan Ahn
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2007-05-12       Impact factor: 4.342

7.  Arthroscopic posterior cruciate ligament reconstruction using bioabsorbable cross-pin femoral fixation: a technical note.

Authors:  Nam-Hong Choi; Kyung-Mo Son; Brian N Victoroff
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2008-05       Impact factor: 4.342

8.  A pitfall of transfix fixation during anterior cruciate ligament reconstruction.

Authors:  Nam-Hong Choi; Kyung-Mo Son; Brian N Victoroff
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2008-05       Impact factor: 4.342

9.  Comparative biomechanical study of the Ligament Plate and other fixation devices in ACL reconstruction.

Authors:  Jae Ang Sim; Ji Hoon Kwak; Sang Hoon Yang; Beom Koo Lee
Journal:  Int Orthop       Date:  2008-10-16       Impact factor: 3.075

10.  Determinants of Femoral Tunnel Length in Anterior Cruciate Ligament Reconstruction: CT Analysis of the Influence of Tunnel Orientation on the Length.

Authors:  Geethan Iyyampillai; Easwar Thirunellai Raman; David Vaithyalingam Rajan; Ajith Krishnamoorthy; Santhosh Sahanand
Journal:  Knee Surg Relat Res       Date:  2013-11-29
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.