The ability of two apex locators to locate the apical foramen: an in vitro study.

Fábio Luiz Cunha D'Assunção1, Diana Santana de Albuquerque, Linalda Correia de Queiroz Ferreira.   

Abstract

The objective of this study was to compare the accuracy of the Root-ZX and Novapex electronic apex locators (EALs) in locating the apical foramen. Forty extracted human teeth were used. The coronal portion of each canal was flared using Gates Glidden drill. Canals were irrigated with 2.5% sodium hypochlorite. The actual length (AL) and electronic length (EL) measurements were made on each specimen separately with both devices with an aid of a K-type file. The results obtained with each EAL were compared with the corresponding control length. The statistical analysis of the results showed EAL reliability in detecting the apical foramen to 89.7% for the Root-ZX and 82.1% for the Novapex, taking the tolerance of +/-0.5 mm into account. A paired sample t test showed that there was no statistically significant difference between the accuracy of the devices (p = 0.4305). The results of this study indicate that the Root-ZX and Novapex are useful and accurate devices for the apical foramen location.

Entities:  

Mesh:

Year:  2006        PMID: 16728251     DOI: 10.1016/j.joen.2005.11.011

Source DB:  PubMed          Journal:  J Endod        ISSN: 0099-2399            Impact factor:   4.171


  8 in total

1.  The accuracy of two electronic apex locators on effect of preflaring and file size: An in vitro study.

Authors:  V Saritha; H Raghu; Twino H Kumar; Shrishail Totad; Laxmikant Kamatagi; Prahlad A Saraf
Journal:  J Conserv Dent       Date:  2021-07-05

2.  Effect of pre-flaring and file size on the accuracy of two electronic apex locators.

Authors:  Manoel Brito-Júnior; Carla Cristina Camilo; Gil Moreira-Júnior; Jesus Djalma Pecora; Manoel Damião Sousa-Neto
Journal:  J Appl Oral Sci       Date:  2012 Sep-Oct       Impact factor: 2.698

3.  Comparative evaluation of accuracy of two electronic apex locators in the presence of various irrigants: An in vitro study.

Authors:  Saru Jain; Ravi Kapur
Journal:  Contemp Clin Dent       Date:  2012-09

4.  Evaluation of Dentaport ZX and Raypex 6 electronic apex locators: an in vivo study.

Authors:  S Moscoso; K Pineda; J Basilio; C Alvarado; M Roig; F Duran-Sindreu
Journal:  Med Oral Patol Oral Cir Bucal       Date:  2014-03-01

5.  In vitro evaluation of Root ZX and Raypex 6 in teeth with different apical diameters.

Authors:  Ugur Aydin; Emrah Karataslioglu; Fatih Aksoy; Cihan Yildirim
Journal:  J Conserv Dent       Date:  2015 Jan-Feb

6.  Evaluation of Conventional Radiography and an Electronic Apex Locator in Determining the Working Length in C-shaped Canals.

Authors:  Hamid Jafarzadeh; Masoud Beyrami; Maryam Forghani
Journal:  Iran Endod J       Date:  2017

7.  In vitro detection of simulated apical root perforation with two electronic apex locators.

Authors:  Kiumars Nazari Moghaddam; Shahrzad Nazari; Leila Shakeri; Kiamars Honardar; Farshid Mirmotalebi
Journal:  Iran Endod J       Date:  2010-02-20

8.  Comparison Between the Accuracy of NovApex Apex Locator and Radiographs in Determining Radiographic Apex.

Authors:  Shahriar Shahi; Saeed Rahimi; Amin Salem Milani; Mohammad Asghari Jafarabadi; Gholam Reza Djoorabbaf Naghsh
Journal:  Iran Endod J       Date:  2011-05-15