Oscillometric blood pressure measurement used for calibration of the arterial tonometry method contributes significantly to error.

S Hansen1, M Staber.   

Abstract

BACKGROUND AND
OBJECTIVE: We have investigated the accuracy of the Colin tonometry module, a non-invasive alternative to intra-arterial blood pressure (BP) measurement. The claims of previous publications range between good agreement to significant discrepancies between the Colin tonometry and intra-arterial methods. The Colin method relies on calibration of the tonometry module using an oscillometric method. Our aim was to look at the effect of this calibration on the accuracy.
METHODS: Ten conscious patients who already had an indwelling arterial catheter were studied. The Colin monitor was set to perform ten calibration cycles first with the oscillometric cuff on the same arm as the tonometry module and secondly on the opposite arm. Simultaneous recording of BP wave forms from the intra-arterial BP transducer and the Colin monitor allowed comparison of beat-by-beat systolic, diastolic and mean BP.
RESULTS: There was considerable inter- and intra-patient variation. In the worst case, the error range was 41 mmHg over ten calibrations. Analysis of variance showed that contra- and ipsilateral calibrations gave a significantly different bias while the multiple calibrations accounted for a significant proportion of the variability in systolic BP error.
CONCLUSIONS: The Colin Tonometry method is not accurate enough to be used with confidence in clinical practice. The main reason for this is its reliance on an oscillometric method for calibration of the tonometry module. Single BP measurements, using either manual or semiautomatic instruments may vary considerably from the 'true' BP due to short-term perturbations of BP.

Entities:  

Mesh:

Year:  2006        PMID: 16723049     DOI: 10.1017/S0265021506000688

Source DB:  PubMed          Journal:  Eur J Anaesthesiol        ISSN: 0265-0215            Impact factor:   4.330


  14 in total

Review 1.  Non-invasive continuous blood pressure monitoring: a review of current applications.

Authors:  Elena Chung; Guo Chen; Brenton Alexander; Maxime Cannesson
Journal:  Front Med       Date:  2013-01-23       Impact factor: 4.592

2.  Ensemble Methodology for Confidence Interval in Oscillometric Blood Pressure Measurements.

Authors:  Soojeong Lee; Gaseong Lee
Journal:  J Med Syst       Date:  2020-03-17       Impact factor: 4.460

3.  Smartphone-based blood pressure monitoring via the oscillometric finger-pressing method.

Authors:  Anand Chandrasekhar; Chang-Sei Kim; Mohammed Naji; Keerthana Natarajan; Jin-Oh Hahn; Ramakrishna Mukkamala
Journal:  Sci Transl Med       Date:  2018-03-07       Impact factor: 17.956

4.  Toward Ubiquitous Blood Pressure Monitoring via Pulse Transit Time: Theory and Practice.

Authors:  Ramakrishna Mukkamala; Jin-Oh Hahn; Omer T Inan; Lalit K Mestha; Chang-Sei Kim; Hakan Töreyin; Survi Kyal
Journal:  IEEE Trans Biomed Eng       Date:  2015-06-05       Impact factor: 4.538

5.  On using maximum a posteriori probability based on a Bayesian model for oscillometric blood pressure estimation.

Authors:  Soojeong Lee; Gwanggil Jeon; Gangseong Lee
Journal:  Sensors (Basel)       Date:  2013-10-10       Impact factor: 3.576

6.  Cuffless Blood Pressure Estimation Using Pressure Pulse Wave Signals.

Authors:  Zeng-Ding Liu; Ji-Kui Liu; Bo Wen; Qing-Yun He; Ye Li; Fen Miao
Journal:  Sensors (Basel)       Date:  2018-12-02       Impact factor: 3.576

7.  Statistical Approaches Based on Deep Learning Regression for Verification of Normality of Blood Pressure Estimates.

Authors:  Soojeong Lee; Gangseong Lee; Gwanggil Jeon
Journal:  Sensors (Basel)       Date:  2019-05-08       Impact factor: 3.576

8.  Cuffless blood pressure estimation based on haemodynamic principles: progress towards mobile healthcare.

Authors:  Takehiro Yamakoshi; Peter Rolfe; Ken-Ichi Yamakoshi
Journal:  PeerJ       Date:  2021-05-25       Impact factor: 2.984

9.  Two-Step Pseudomaximum Amplitude-Based Confidence Interval Estimation for Oscillometric Blood Pressure Measurements.

Authors:  Soojeong Lee; Gwanggil Jeon; Seokhoon Kang
Journal:  Biomed Res Int       Date:  2015-10-04       Impact factor: 3.411

10.  Cuffless blood pressure estimation using only a smartphone.

Authors:  Kenta Matsumura; Peter Rolfe; Sogo Toda; Takehiro Yamakoshi
Journal:  Sci Rep       Date:  2018-05-08       Impact factor: 4.379

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.