Complications at screw removal in slipped capital femoral epiphysis treated by cannulated titanium screws.

T Ilchmann1, K Parsch.   

Abstract

INTRODUCTION: Various modes of fixation are proposed for the treatment of slips of the capital femoral epiphysis (SCFE). We describe our experience with the use and removal of a new, cannulated titanium screw (Asnis III, Stryker, Howmedica). PATIENTS AND METHODS: Single cannulated titanium screws had been inserted in 101 hips of 65 patients in the 3-year period from 2001 to 2003. These pins were used for in situ fixation of minor chronic slips in 41 patients and for prophylactic fixation on the contralateral side in patients with open physis in 60 patients.
RESULTS: The insertion of these screws was achieved without any real problem. The mean surgical time was 25 min (13-46 min). Problems came up when we started to remove the pins. Hardware retrieval was attempted in 27 patients with 43 pins. The mean surgical time for removal was double the average time of insertion with 51 min (26-107 min). The hexagonal Allen sockets proved to be too weak to overcome the necessary torque for loosening the pin from bone and applying the reverse-cutting-force, necessary to extract the pin. Eleven patients needed extensive chiselling. Two adolescents sustained a subtrochanteric fracture 5 and 7 weeks after hardware removal. Seven pins could not be totally removed.
CONCLUSION: Due to the considerable disadvantages encountered in our series we conclude that Asnis III cannulated screws should be suspended from further use in SCFE.

Entities:  

Mesh:

Substances:

Year:  2006        PMID: 16721619     DOI: 10.1007/s00402-006-0126-5

Source DB:  PubMed          Journal:  Arch Orthop Trauma Surg        ISSN: 0936-8051            Impact factor:   3.067


  7 in total

Review 1.  [Early and late complications and their management in slipped capital femoral epiphysis].

Authors:  F F Fernandez; O Eberhardt; T Wirth
Journal:  Orthopade       Date:  2019-08       Impact factor: 1.087

2.  Complications and associated risk factors at screw removal in slipped capital femoral epiphysis treated by cannulated stainless steel screws.

Authors:  Juan Pretell-Mazzini; Victor Rodriguez-Vega; Jorge Muñoz-Ledesma; Eva María Andrés-Esteban; Rafael Marti-Ciruelos; José Luis González-López; Angel Curto-De la Mano
Journal:  J Child Orthop       Date:  2012-08-01       Impact factor: 1.548

3.  [Slipped capital femoral epiphysis].

Authors:  C Zilkens; M Jäger; B Bittersohl; Y-J Kim; M B Millis; R Krauspe
Journal:  Orthopade       Date:  2010-10       Impact factor: 1.087

4.  Optimum Configuration of Cannulated Compression Screws for the Fixation of Unstable Femoral Neck Fractures: Finite Element Analysis Evaluation.

Authors:  Jiantao Li; Menglin Wang; Jianfeng Zhou; Lin Han; Hao Zhang; Chen Li; Lianting Li; Ming Hao
Journal:  Biomed Res Int       Date:  2018-12-09       Impact factor: 3.411

Review 5.  Role of the prophylactic fixation of contralateral unaffected hip in paediatric unilateral slipped capital femoral epiphysis: a systematic review.

Authors:  Filippo Maria Anghilieri; Ilaria Morelli; Giuseppe M Peretti; Fabio Verdoni; Domenico Curci
Journal:  EFORT Open Rev       Date:  2022-02-16

6.  Management of Cannulated Screw Failure and Recurrent SCFE Displacement - Case Report.

Authors:  Nathan A Jacobson; Siegfried P Feierabend; Christopher L Lee
Journal:  J Orthop Case Rep       Date:  2014 Jan-Mar

7.  Outcomes after slipped capital femoral epiphysis: a population-based study with three-year follow-up.

Authors:  B Herngren; M Stenmarker; K Enskär; G Hägglund
Journal:  J Child Orthop       Date:  2018-10-01       Impact factor: 1.548

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.