The effect of sterilisation on a poly(dimethylsiloxane)/poly(hexamethylene oxide) mixed macrodiol-based polyurethane elastomer.

Anne Simmons1, Jari Hyvarinen, Laura Poole-Warren.   

Abstract

The effect of various forms of sterilisation on a novel thermoplastic polyurethane elastomer synthesised using poly(hexamethylene oxide) (PHMO) and poly(dimethylsiloxane) (PDMS) macrodiols has been studied. The five sterilisation methods investigated were ethylene oxide (EtO) (single and multiple cycles), gas plasma, steam, vapour phase liquid chemical and gamma-irradiation (single and multiple cycles). Following sterilisation, scanning electron microscopy (SEM) and attenuated total reflectance Fourier transform infrared spectroscopy (ATR-FTIR) were used to assess changes in the surface chemical structure and morphology, and gel permeation chromatography (GPC) and tensile testing were used to examine changes in bulk characteristics. Biostability was assessed using subcutaneous implantation of strained samples in sheep for 6 weeks. The results showed that the properties of the commercially available control material, Pellethane 2363-80A, were significantly affected by exposure to gamma-irradiation, steam and multiple cycles of EtO with aging and implantation compounding the effect. Exposure to a gas plasma sterilisation process resulted in significant degradation in both polyurethanes. A vapour phase liquid chemical sterilisation process caused minimal adverse effects. Sterilisation of the PDMS-based polyurethane using EtO, gamma-irradiation and autoclaving resulted in no significant changes in properties. The material's biostability was also unaffected by exposure to each of these sterilisation processes followed by short-term implantation suggesting that this material is a potential candidate for use in a wide range of implantable medical devices sterilised using commercially available processes. Further biostability studies should be performed to assess the longer-term in vivo biostability of the PDMS-based material sterilised using autoclaving and gamma-irradiation.

Entities:  

Mesh:

Substances:

Year:  2006        PMID: 16690122     DOI: 10.1016/j.biomaterials.2006.04.017

Source DB:  PubMed          Journal:  Biomaterials        ISSN: 0142-9612            Impact factor:   12.479


  8 in total

1.  Exploiting novel sterilization techniques for porous polyurethane scaffolds.

Authors:  Serena Bertoldi; Silvia Farè; Håvard Jostein Haugen; Maria Cristina Tanzi
Journal:  J Mater Sci Mater Med       Date:  2015-04-17       Impact factor: 3.896

2.  Sterilization of exopolysaccharides produced by deep-sea bacteria: impact on their stability and degradation.

Authors:  Emilie Rederstorff; Ahmed Fatimi; Corinne Sinquin; Jacqueline Ratiskol; Christophe Merceron; Claire Vinatier; Pierre Weiss; Sylvia Colliec-Jouault
Journal:  Mar Drugs       Date:  2011-02-10       Impact factor: 6.085

3.  Evaluation of Sterilisation Techniques for Regenerative Medicine Scaffolds Fabricated with Polyurethane Nonbiodegradable and Bioabsorbable Nanocomposite Materials.

Authors:  Michelle Griffin; Naghmeh Naderi; Deepak M Kalaskar; Edward Malins; Remzi Becer; Catherine A Thornton; Iain S Whitaker; Ash Mosahebi; Peter E M Butler; Alexander M Seifalian
Journal:  Int J Biomater       Date:  2018-10-03

4.  Effects of sterilization treatments on bulk and surface properties of nanocomposite biomaterials.

Authors:  Maqsood Ahmed; Geoffrey Punshon; Arnold Darbyshire; Alexander M Seifalian
Journal:  J Biomed Mater Res B Appl Biomater       Date:  2013-10       Impact factor: 3.368

5.  Biodegradable and Toughened Composite of Poly(Propylene Carbonate)/Thermoplastic Polyurethane (PPC/TPU): Effect of Hydrogen Bonding.

Authors:  Dongmei Han; Guiji Chen; Min Xiao; Shuanjin Wang; Shou Chen; Xiaohua Peng; Yuezhong Meng
Journal:  Int J Mol Sci       Date:  2018-07-13       Impact factor: 5.923

Review 6.  Sustained Release Drug Delivery Applications of Polyurethanes.

Authors:  Michael B Lowinger; Stephanie E Barrett; Feng Zhang; Robert O Williams
Journal:  Pharmaceutics       Date:  2018-05-09       Impact factor: 6.321

7.  Effect of Ethylene Oxide and Gamma (γ-) Sterilization on the Properties of a PLCL Polymer Material in Balloon Implants.

Authors:  Moran Haim Zada; Awanish Kumar; Omar Elmalak; Guy Mechrez; Abraham J Domb
Journal:  ACS Omega       Date:  2019-12-03

Review 8.  A critical review of fibrous polyurethane-based vascular tissue engineering scaffolds.

Authors:  Reza Rahbarghazi; Soodabeh Davaran; Sonia Fathi-Karkan; Behnaz Banimohamad-Shotorbani; Sepideh Saghati
Journal:  J Biol Eng       Date:  2022-03-24       Impact factor: 4.355

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.