Literature DB >> 16685674

Observation and quantification of cavitation on a mechanical heart valve with an electro-hydraulic total artificial heart.

H Lee1, Y Taenaka.   

Abstract

In previous studies, we investigated the cavitation phenomenon in a mechanical heart valve using an electro-hydraulic total artificial heart. With this system, a 50% glycerin solution kept at 37 degrees C was used as the working fluid. We reported that most of the cavitation bubbles were observed near the valve stop and were caused by the squeeze flow. However, in these studies, the effect of the partial pressure of CO(2) on the mechanical heart valve cavitation was neglected. In this study, in order to investigate the effect of the partial pressure of CO(2) on mechanical heart valve cavitation using an electro-hydraulic total artificial heart, we controlled the partial pressure of the CO(2) in vitro. A 25-mm Medtronic Hall valve was installed in the mitral position of an electro-hydraulic total artificial heart. In order to quantify the mechanical heart valve cavitation, we used a high-speed camera. Even though cavitation intensity slightly increased with increases in the PCO(2) at heart rates of 60, 70 and 100 bpm, throughout the experiment, there was no significant difference between the PCO(2) and cavitation intensity.

Entities:  

Mesh:

Year:  2006        PMID: 16685674     DOI: 10.1177/039139880602900308

Source DB:  PubMed          Journal:  Int J Artif Organs        ISSN: 0391-3988            Impact factor:   1.595


  1 in total

1.  Observation of cavitation pits on mechanical heart valve surfaces in an artificial heart used in in vitro testing.

Authors:  Hwansung Lee; Akihiko Homma; Eisuke Tatsumi; Yoshiyuki Taenaka
Journal:  J Artif Organs       Date:  2010-02-13       Impact factor: 1.731

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.