Evaluation of a new blood glucose monitoring system with auto-calibration for home and hospital bedside use.

Charles Kilo1, William T Dickey, Judy O Joynes, Mary B Pinson, John M Baum, Joan L Parkes, Donald R Parker.   

Abstract

A new self-calibrating blood glucose monitoring system (BGMS) was evaluated in a series of clinical studies with both ambulatory subjects and with hospitalized patients. The new BGMS requires a 0.6microL sample volume, provides results in 15s, and uses a glucose dehydrogenase chemistry that is oxygen independent. In the first study, Ascensia Contour meters calibrated to whole blood were tested by health care professionals (HCP) and lay users at two clinical sites. Both HCPs and lay users obtained results that fulfilled the ISO 15197:2003 criteria that 95% of self-monitoring blood glucose (SMBG) measurements should fall within +/-20% (for blood glucose (BG) concentrations> or =4.2mmol/L or +/-0.83mmol/L for BG concentrations<4.2mmol/L) of the laboratory value. Lay users and HCPs obtained 97.2 and 96.7% of glucose results within ISO criteria, respectively. In a second study, HCPs assayed blood samples from patients at the hospital bedside using meters calibrated to give whole blood glucose and meters calibrated to give plasma glucose results. Overall, 94.7% of the measurements met the ISO 15197:2003 criteria. Most lay subjects rated the BGMS as either excellent or very good in a questionnaire, and were able to use it properly without training. These findings indicate that this new BGMS is a convenient and accurate instrument system suitable for both hospital bedside use by HCPs and for SMBG by people who routinely monitor their blood glucose.

Entities:  

Mesh:

Substances:

Year:  2006        PMID: 16644056     DOI: 10.1016/j.diabres.2006.02.012

Source DB:  PubMed          Journal:  Diabetes Res Clin Pract        ISSN: 0168-8227            Impact factor:   5.602


  3 in total

1.  Calibration-Free Electrochemical Biosensors Supporting Accurate Molecular Measurements Directly in Undiluted Whole Blood.

Authors:  Hui Li; Philippe Dauphin-Ducharme; Gabriel Ortega; Kevin W Plaxco
Journal:  J Am Chem Soc       Date:  2017-08-02       Impact factor: 15.419

2.  New Criteria for Assessing the Accuracy of Blood Glucose Monitors meeting, October 28, 2011.

Authors:  John Walsh; Ruth Roberts; Robert A Vigersky; Frank Schwartz
Journal:  J Diabetes Sci Technol       Date:  2012-03-01

Review 3.  Analytical and clinical performance of blood glucose monitors.

Authors:  Suzanne Austin Boren; William L Clarke
Journal:  J Diabetes Sci Technol       Date:  2010-01-01
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.