Literature DB >> 16636604

Biocompatibility of a new bioabsorbable radiopaque stent material (BaSO4 containing poly-L,D-lactide) in the rat pancreas.

Teemu Lämsä1, Haitao Jin, Joonas Mikkonen, Johanna Laukkarinen, Juhani Sand, Isto Nordback.   

Abstract

BACKGROUND/AIM: During recent years, we have been developing bioabsorbable biliary stents with promising experimental results. In developing pancreatic stents before long-term experiments, the acute toxicity to the pancreas of a bioabsorbable, radiopaque polylactide (PLA 96-barium sulfate, BaSO4) stent material was investigated.
METHODS: The pancreas of 65 Sprague-Dawley rats was exposed either to radiopaque stent material [PLA 96 with 25% (w/w) of BaSO4], radiolucent stent material (PLA 96), or inert steel by inserting a 5-mm-long (diameter 0.3 mm) fiber/stick of material into the pancreas after laparotomy under general anesthesia. Pancreatic tissue specimens and blood samples were taken after 1, 3, 7, and 21 days for histological examination and amylase activity measurements. Samples were also taken from 5 baseline (control) rats without exposing to any materials.
RESULTS: The baseline serum amylase activity was normal, and no histological changes in the pancreas were observed. A significant increase (mean +/- SE) in the serum amylase activity was observed only on day 1 in the animals having radiopaque stent material (PLA 96-BaSO4; 5,845 +/- 1,135 U/l), steel (4,946 +/- 667 U/l), or radiolucent stent material (PLA 96; 7,684 +/- 667 U/l) inserted. There was slightly more acinar cell necrosis on day 7 in the steel group than in the radiopaque stent (PLA 96-BaSO(4)) group (p = 0.028).
CONCLUSIONS: Radiopaque stent material (PLA 96-BaSO4) was not more toxic than the reference steel material in the rat pancreas during the 21-day observation period and is thus applicable for further in vivo experiments when developing pancreatic bioabsorbable stents. Copyright (c) 2006 S. Karger AG, Basel and IAP.

Entities:  

Mesh:

Substances:

Year:  2006        PMID: 16636604     DOI: 10.1159/000092772

Source DB:  PubMed          Journal:  Pancreatology        ISSN: 1424-3903            Impact factor:   3.996


  5 in total

Review 1.  Biocompatible Polymer Materials with Antimicrobial Properties for Preparation of Stents.

Authors:  Kateřina Škrlová; Kateřina Malachová; Alexandra Muñoz-Bonilla; Dagmar Měřinská; Zuzana Rybková; Marta Fernández-García; Daniela Plachá
Journal:  Nanomaterials (Basel)       Date:  2019-10-31       Impact factor: 5.076

Review 2.  Radiopaque nano and polymeric materials for atherosclerosis imaging, embolization and other catheterization procedures.

Authors:  Li Tian; Linfeng Lu; James Feng; Marites P Melancon
Journal:  Acta Pharm Sin B       Date:  2018-03-27       Impact factor: 11.413

3.  Radiopaque Fully Degradable Nanocomposites for Coronary Stents.

Authors:  Hui Ying Ang; Daniel Toong; Wei Shoon Chow; Welly Seisilya; Wei Wu; Philip Wong; Subbu S Venkatraman; Nicolas Foin; Yingying Huang
Journal:  Sci Rep       Date:  2018-11-27       Impact factor: 4.379

4.  Design and Analysis of a Biodegradable Polycaprolactone Flow Diverting Stent for Brain Aneurysms.

Authors:  Kaitlyn Tidwell; Seth Harriet; Vishal Barot; Andrew Bauer; Melville B Vaughan; Mohammad R Hossan
Journal:  Bioengineering (Basel)       Date:  2021-11-12

5.  Crystallization Behavior and Mechanical Properties of Poly(ε-caprolactone) Reinforced with Barium Sulfate Submicron Particles.

Authors:  Hegoi Amestoy; Paul Diego; Emilio Meaurio; Jone Muñoz; Jose-Ramon Sarasua
Journal:  Materials (Basel)       Date:  2021-05-02       Impact factor: 3.623

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.