Removal torque of miniscrews used for orthodontic anchorage--a preliminary report.

Yi-Jane Chen1, Yuan-Hou Chen, Li-Deh Lin, Chung-Chen Jane Yao.   

Abstract

PURPOSE: Implant anchors such as miniscrews and miniplates have been loaded immediately for anchorage during orthodontic treatment. The purpose of this study was to measure the removal torque of immediately loaded miniscrews after clinical usage and to determine the possible factors associated with this value.
MATERIALS AND METHODS: From 29 patients with malocclusions, 46 miniscrews were removed, and removal torque was measured with a torque gauge. Removal torque values were subjected to statistical analysis for possible association of different clinical characteristics.
RESULTS: The mean removal torque value was 1.10 kg x cm, and removal torques for 50% of the implants were greater than 0.89 kg x cm (8.7 N x cm). Removal torque values were significantly higher in the mandible than in the maxilla. The removal torques of 15-mm and 17-mm miniscrews were significantly higher than those of 13-mm miniscrews. Therefore, the site of implantation and miniscrew length were important factors associated with removal torque. However, there was no significant correlation between the removal torque value and age, gender, healing time, or time in function. DISCUSSION: When miniscrews are used as anchorage for uprighting tipped molars, excessive torque in a counterclockwise direction may loosen them. From the measurements obtained in this investigation, miniscrews can sufficiently sustain an uprighting moment.
CONCLUSION: The removal torque values of the majority of miniscrews in this study population when loaded immediately as orthodontic anchorage were greater than 0.89 kg x cm, and this was sufficient for these implants to fulfill their purpose as anchors in 3-dimensional tooth movements.

Entities:  

Mesh:

Year:  2006        PMID: 16634500

Source DB:  PubMed          Journal:  Int J Oral Maxillofac Implants        ISSN: 0882-2786            Impact factor:   2.804


  7 in total

1.  In vivo degradation of orthodontic miniscrew implants: surface analysis of as-received and retrieved specimens.

Authors:  Masahiro Iijima; Takeshi Muguruma; Masahiro Kawaguchi; Yoshitaka Yasuda; Itaru Mizoguchi
Journal:  J Mater Sci Mater Med       Date:  2015-01-29       Impact factor: 3.896

2.  Accuracy of torque-limiting devices for mini-implant removal: an in vitro study.

Authors:  A Pauls; M Nienkemper; D Drescher
Journal:  J Orofac Orthop       Date:  2013-05-08       Impact factor: 1.938

3.  Mechanical properties of self-drilling orthodontic micro-implants with different diameters.

Authors:  Yan Chen; Hee Moon Kyung; Limin Gao; Won-Jae Yu; Eun-Ju Bae; Sang-Min Kim
Journal:  Angle Orthod       Date:  2010-09       Impact factor: 2.079

4.  A processing method for orthodontic mini-screws reuse.

Authors:  Saeed Noorollahian; Shiva Alavi; Mohammad Monirifard
Journal:  Dent Res J (Isfahan)       Date:  2012-07

5.  The effect of multiple processing and re-use on orthodontic mini-screw torque values.

Authors:  Saeed Noorollahian; Shiva Alavi; Elahe Rafiei
Journal:  Dent Res J (Isfahan)       Date:  2015 May-Jun

6.  Drilling Capability of Orthodontic Miniscrews: In Vitro Study.

Authors:  Alessandra Marchi; Matteo Camporesi; Maurizio Festa; Luis Salvatierra; Sara Izadi; Giampietro Farronato
Journal:  Dent J (Basel)       Date:  2020-12-21

7.  Factors influencing the removal torque of palatal implant used for orthodontic anchorage.

Authors:  Marc Andreas Schätzle; Monika Hersberger-Zurfluh; Raphael Patcas
Journal:  Prog Orthod       Date:  2021-06-07       Impact factor: 2.750

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.