Santiago Echeverri1, Pierre-Francois Leyvraz, Pierre-Yves Zambelli, Brigitte M Jolles.
Abstract
Acetabular cup orientation is a key factor determining hip stability, and standard mechanical guides have shown little help in improving alignment. An in vitro study was carried out to compare the accuracy and precision of a new gravity-assisted guidance system with a standard mechanical guide. Three hundred ten cups were impacted by 5 surgeons, and the final cup orientation was measured. With the new guide, the average error in anteversion was 0.4 degrees , compared with 10.4 degrees with the standard guide and 0.3 degrees and -4.7 degrees , respectively, for abduction angles. The average time required for orienting the cups was similar for both guides. The accuracy and reproducibility obtained with the new guide were better (P < .0001). These good results would require a clinical validation.Mesh:
Year: 2006 PMID: 16627152 DOI: 10.1016/j.arth.2005.04.015
Source DB: PubMed Journal: J Arthroplasty ISSN: 0883-5403 Impact factor: 4.757