Literature DB >> 16618845

What happens if you delay scanning Digora phosphor storage plates (PSPs) for up to 4 hours?

M G B Q Martins1, E J Whaites, G M B Ambrosano, F Haiter Neto.   

Abstract

OBJECTIVES: To assess, both objectively and subjectively, the effects of different combinations of storage conditions and delaying scanning for up to 4 h on digital images captured using Digora phosphor storage plates (PSPs).
METHOD: Standardized images were obtained of an aluminium step wedge and dry mandible in acrylic using GE 1000 and phosphor storage plates (PSPs) from the Digora (Soredex) digital system. 12 plates were exposed and immediately scanned to produce the baseline gold standard. The plates were re-exposed and stored using three different storage combinations--(A) ambient temperature, (R) refrigeration and (S) low humidity, and then scanned after 10 min, 30 min, 1 h, 2 h, 3 h or 4 h. The objective analysis was carried out by pixel density measurements and the data analysed statistically using analysis of variance. Subjective analysis was carried out by three oral radiologists and the results analysed using the Mann Whitney U-test.
RESULTS: Objective analysis showed loss of pixel density after 4 h using all storage combinations. Subjectively, this loss of density was not evident.
CONCLUSION: There is a loss of image density at 4 h when using Digora PSPs which can not be detected clinically, but could compromise multisite research.

Mesh:

Year:  2006        PMID: 16618845     DOI: 10.1259/dmfr/29710762

Source DB:  PubMed          Journal:  Dentomaxillofac Radiol        ISSN: 0250-832X            Impact factor:   2.419


  7 in total

1.  The effect of delayed scanning of storage phosphor plates on occlusal caries detection.

Authors:  E Sogur; B G Baksi; A Mert
Journal:  Dentomaxillofac Radiol       Date:  2012-01-26       Impact factor: 2.419

2.  Analysis of the deterioration of photostimulable phosphor plates.

Authors:  Gustavo Nascimento de Souza-Pinto; Gustavo Machado Santaella; Amanda Achkar Coli; Anne Caroline Oenning; Francisco Haiter-Neto
Journal:  Dentomaxillofac Radiol       Date:  2020-04-29       Impact factor: 2.419

3.  Effect of beam quality and readout direction in the edge profile on the modulation transfer function of photostimulable phosphor systems via the edge method.

Authors:  Shinya Takarabe; Taku Kuramoto; Yusuke Shibayama; Hiroki Tsuru; Masato Tatsumi; Toyoyuki Kato; Kazutoshi Okamura; Kazunori Yoshiura
Journal:  J Med Imaging (Bellingham)       Date:  2021-07-20

4.  Evaluation of the Effect of Light and Scanning Time Delay on The Image Quality of Intra Oral Photostimulable Phosphor Plates.

Authors:  Amir Eskandarloo; Arman Yousefi; Setareh Soheili; Karim Ghazikhanloo; Payam Amini; Haniyeh Mohammadpoor
Journal:  Open Dent J       Date:  2017-12-29

5.  Diagnosis of Approximal Caries after Delayed Scanning of Photostimulable Phosphor Plates.

Authors:  Zahra Ghoncheh; Yasaman Kheirandish; Hanieh Kaviani; Mohammad Javad Kharazi Fard; Kurosh Sadeghi
Journal:  J Dent (Shiraz)       Date:  2017-09

6.  A comparative study of the physicochemical properties of hesperidin, MTA-Angelus and calcium hydroxide as pulp capping materials.

Authors:  Ebtesam O Abo El-Mal; Ashraf M Abu-Seida; Salma H El Ashry
Journal:  Saudi Dent J       Date:  2018-09-15

7.  Effect of differences in pixel size on image characteristics of digital intraoral radiographic systems: a physical and visual evaluation.

Authors:  Taku Kuramoto; Shinya Takarabe; Kazutoshi Okamura; Kenshi Shiotsuki; Yusuke Shibayama; Hiroki Tsuru; Hiroshi Akamine; Masato Tatsumi; Toyoyuki Kato; Junji Morishita; Kazunori Yoshiura
Journal:  Dentomaxillofac Radiol       Date:  2020-05-07       Impact factor: 2.419