| Literature DB >> 16574443 |
Kevin M Koch1, Scott McIntyre, Terence W Nixon, Douglas L Rothman, Robin A de Graaf.
Abstract
Dynamic alteration of shim settings during a multi-slice imaging experiment can improve static magnetic-field homogeneity over extended volumes. In this report, a pre-emphasized dynamic shim updating (DSU) system capable of rapidly updating all non-degenerate zeroth through second-order shims is presented and applied to high-field multi-slice imaging studies on the human brain. DSU is utilized in both non-oblique and oblique slicing geometries while updating in-plane and through-slice shims. Image-based magnetic-field maps are used to quantify homogeneity improvements and comparisons are made on a slice-specific basis between static global shimming and increasing orders of shim inclusion utilized DSU. The influence of oblique slicing geometry on DSU-utilized global homogeneity is also quantified computationally. Finally, the effect of DSU on susceptibility artifact reduction in single-shot axial-sliced EPI is analyzed using experimental acquisitions.Entities:
Mesh:
Year: 2006 PMID: 16574443 DOI: 10.1016/j.jmr.2006.03.007
Source DB: PubMed Journal: J Magn Reson ISSN: 1090-7807 Impact factor: 2.229