Literature DB >> 16563998

Quality assurance of treatment plans for interstitial and intracavitary high-dose-rate brachytherapy.

Rupak K Das1, Kristin A Bradley, Ian A Nelson, Rakesh Patel, Bruce R Thomadsen.   

Abstract

PURPOSE: Quality assurance for complex high-dose-rate (HDR) treatment planning has always been a challenge to the physics community because of the time constraint between HDR planning and the delivery of the treatment. This study proposes an efficient, precise, and easy method for checking the complex computer calculation. METHODS AND MATERIALS: Posttreatment, three-dimensional dose-volume study was performed for 98 patients with 128 new treatment plans along with 30 library plans. Volumes covered by the 100% isodose line, source activity (Ci), total dwell time (s), and the prescription dose (100%) were recorded. Variation of R(V) defined as (irradiated time x activity/elongation factor x prescribed dose) with volume was studied for different catheter systems.
RESULTS: Parametric fit of R(V) with volume for three different systems that cover most of the interstitial and intracavitary brachytherapy implants agrees within +/-6%.
CONCLUSIONS: The excellent agreement of R(V) derived from this simplistic point source model with three-dimensional dose calculations for individual HDR treatment plans clearly establishes that for an implant with known number of catheters, the time needed to deliver a prescribed dose to a given prescription volume can be easily predicted.

Entities:  

Mesh:

Year:  2006        PMID: 16563998     DOI: 10.1016/j.brachy.2005.11.002

Source DB:  PubMed          Journal:  Brachytherapy        ISSN: 1538-4721            Impact factor:   2.362


  4 in total

1.  Need of independent dose verification in brachytherapy.

Authors:  A Shanta
Journal:  J Med Phys       Date:  2008-07

2.  A Japanese prospective multi-institutional feasibility study on accelerated partial breast irradiation using interstitial brachytherapy: treatment planning and quality assurance.

Authors:  Yuki Otani; Takayuki Nose; Takushi Dokiya; Toshiaki Saeki; Yu Kumazaki; Shuuji Asahi; Iwao Tsukiyama; Ichirou Fukuda; Hiroshi Sekine; Naoto Shikama; Takao Takahashi; Ken Yoshida; Tadayuki Kotsuma; Norikazu Masuda; Eisaku Yoden; Kazutaka Nakashima; Taisei Matsumura; Shino Nakagawa; Seiji Tachiiri; Yoshio Moriguchi; Jun Itami; Masahiko Oguchi
Journal:  Radiat Oncol       Date:  2015-06-04       Impact factor: 3.481

3.  High-dose rate intracavitary brachytherapy pretreatment dwell position verification using a transparent applicator.

Authors:  Yuki Otani; Iori Sumida; Takayuki Nose; Shigetoshi Shimamoto; Hirofumi Okubo; Kazuhiko Ogawa
Journal:  J Appl Clin Med Phys       Date:  2018-06-30       Impact factor: 2.102

4.  The usefulness of an independent patient-specific treatment planning verification method using a benchmark plan in high-dose-rate intracavitary brachytherapy for carcinoma of the uterine cervix.

Authors:  Yutaka Takahashi; Masahiko Koizumi; Iori Sumida; Fumiaki Isohashi; Toshiyuki Ogata; Yuichi Akino; Yasuo Yoshioka; Shintaro Maruoka; Shinichi Inoue; Koji Konishi; Kazuhiko Ogawa
Journal:  J Radiat Res       Date:  2012-07-17       Impact factor: 2.724

  4 in total