Multi-center evaluation of analytical performance of the microparticle enzyme immunoassay for sirolimus.

D Wilson1, F Johnston, D Holt, M Moreton, J Engelmayer, J-M Gaulier, H Luthe, P Marquet, D Moscato, M Oellerich, R Mosso, F Streit, M Brunet, C Fillee, R Schmid, P Wallemacq, G Barnes.   

Abstract

OBJECTIVES: This study evaluated the analytical characteristics of the new Abbott microparticle enzyme immunoassay (MEIA) for sirolimus. DESIGN AND METHODS: The protocol consisted of nine sections: evaluation of antibody specificity, linearity, detection limit, quantification limit, endogenous interferents, exogenous interferents, precision, proficiency testing panel, and method comparison.
RESULTS: The mean analytical detection limit was 0.68 microg/L. The sirolimus concentration corresponding to a total CV of 20% was 1.5 microg/L. Linearity of response was demonstrated across the dynamic range of the assay. Total precision (CVs) at QC control levels from 5 to 22 microg/L ranged from 5.7 to 12.6%. Assay standardization was found to be in good agreement with LC/MS/MS as compared with target values for spiked sirolimus proficiency samples from an international sirolimus proficiency testing program. Good correlations (R values) of the immunoassay were observed in comparisons to LC/MS/MS. R values tended to be lower in comparisons with LC/UV methods. Across both LC-based methods and all study sites, there was approximately 25% overall positive slope bias due to cross reactivity of the MEIA antibody to metabolites of sirolimus. The assay cross-reactivity to metabolites of sirolimus parent drug ranged from 6 to 63%. Assay interferences were minimal with the exception of hematocrit, which presented a negative relationship to measured sirolimus concentration.
CONCLUSIONS: The MEIA demonstrated acceptable analytical characteristics for use for routine monitoring of sirolimus immunosuppressive therapy, and is a viable alternative to HPLC-based methods for sirolimus monitoring.

Entities:  

Mesh:

Substances:

Year:  2006        PMID: 16545357     DOI: 10.1016/j.clinbiochem.2006.01.017

Source DB:  PubMed          Journal:  Clin Biochem        ISSN: 0009-9120            Impact factor:   3.281


  2 in total

1.  Development and validation of an HPLC method for the analysis of sirolimus in drug products.

Authors:  Ziba Islambulchilar; Saeed Ghanbarzadeh; Shahram Emami; Hadi Valizadeh; Parvin Zakeri-Milani
Journal:  Adv Pharm Bull       Date:  2012-05-20

2.  Determination of blood sirolimus concentrations in liver and kidney transplant recipients using the Innofluor fluorescence polarization immunoassay: comparison with the microparticle enzyme immunoassay and high-performance liquid chromatography-ultraviolet method.

Authors:  Lorena Bouzas; Jesús Hermida; J Carlos Tutor
Journal:  Ups J Med Sci       Date:  2009       Impact factor: 2.384

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.