Literature DB >> 16540488

Microprocessor controlled compliance monitor for eye drop medication.

M M Hermann1, M Diestelhorst.   

Abstract

BACKGROUND/AIMS: The effectiveness of a self administered eye drop medication can only be assessed if the compliance is known. The authors studied the specificity and sensitivity of a new microprocessor controlled monitoring device.
METHODS: The monitoring system was conducted by an 8 bit microcontroller for data acquisition and storage with sensors measuring applied pressure to the bottle, temperature, and vertical position. 10 devices were mounted under commercial 10 ml eye drops. Test subjects had to note down each application manually. A total of 15 applications each within 3 days was intended.
RESULTS: Manual reports confirmed 15 applications for each of the 10 bottles. The monitoring devices detected a total of 149 events; one was missed; comprising a sensitivity of 99%. Two devices registered three applications, which did not appear in the manual protocols, indicating a specificity of about 98%. Refrigerated bottles were correctly identified. The battery lifetime exceeded 60 days.
CONCLUSION: The new monitoring device demonstrated a high reliability of the collected compliance data. The important, yet often unknown, influence of compliance in patient care and clinical trials shall be illuminated by the new device. This may lead to a better adapted patient care. Studies will profit from a higher credibility and results will be less influenced by non-compliance.

Entities:  

Mesh:

Substances:

Year:  2006        PMID: 16540488      PMCID: PMC1857146          DOI: 10.1136/bjo.2005.088526

Source DB:  PubMed          Journal:  Br J Ophthalmol        ISSN: 0007-1161            Impact factor:   4.638


  19 in total

1.  The compliance-questionnaire-rheumatology compared with electronic medication event monitoring: a validation study.

Authors:  Erik de Klerk; Désirée van der Heijde; Robert Landewé; Hille van der Tempel; Sjef van der Linden
Journal:  J Rheumatol       Date:  2003-11       Impact factor: 4.666

2.  Medication monitor for ophthalmology.

Authors:  R D Yee; P M Hahn; R E Christensen
Journal:  Am J Ophthalmol       Date:  1974-11       Impact factor: 5.258

3.  A miniature compliance monitor for eyedrop medication.

Authors:  M A Kass; D W Meltzer; M Gordon
Journal:  Arch Ophthalmol       Date:  1984-10

4.  Compliance with treatment of patients with chronic open-angle glaucoma.

Authors:  J M MacKean; A R Elkington
Journal:  Br J Ophthalmol       Date:  1983-01       Impact factor: 4.638

5.  Glaucoma patients not compliant with their drug therapy: clinical and behavioural aspects.

Authors:  P A Granström
Journal:  Br J Ophthalmol       Date:  1982-07       Impact factor: 4.638

6.  Patient compliance in glaucoma.

Authors:  S Bloch; A R Rosenthal; L Friedman; P Caldarolla
Journal:  Br J Ophthalmol       Date:  1977-08       Impact factor: 4.638

7.  Persistence and adherence with topical glaucoma therapy.

Authors:  Beth L Nordstrom; David S Friedman; Essy Mozaffari; Harry A Quigley; Alexander M Walker
Journal:  Am J Ophthalmol       Date:  2005-10       Impact factor: 5.258

Review 8.  Compliance with ocular therapy.

Authors:  F S Ashburn; I Goldberg; M A Kass
Journal:  Surv Ophthalmol       Date:  1980 Jan-Feb       Impact factor: 6.048

9.  A medication monitor and fluorescein technique designed to study medication behaviour.

Authors:  S E Norell; P A Granström; R Wassen
Journal:  Acta Ophthalmol (Copenh)       Date:  1980-06

10.  Factors for glaucoma progression and the effect of treatment: the early manifest glaucoma trial.

Authors:  M Cristina Leske; Anders Heijl; Mohamed Hussein; Bo Bengtsson; Leslie Hyman; Eugene Komaroff
Journal:  Arch Ophthalmol       Date:  2003-01
View more
  4 in total

1.  [Preservative-containing eye drops and adherence in ophthalmological practice].

Authors:  I Lanzl; T Kaercher
Journal:  Ophthalmologe       Date:  2012-11       Impact factor: 1.059

Review 2.  [Glaucoma in the aged--barriers to understanding and compliance].

Authors:  T S Dietlein
Journal:  Ophthalmologe       Date:  2006-09       Impact factor: 1.059

3.  Electronic compliance monitoring of topical treatment after ophthalmic surgery.

Authors:  Manuel Marcel Hermann; Can Ustündag; Michael Diestelhorst
Journal:  Int Ophthalmol       Date:  2010-04-07       Impact factor: 2.031

4.  Inkjet Printed Textile Force Sensitive Resistors for Wearable and Healthcare Devices.

Authors:  Beomjun Ju; Inhwan Kim; Braden M Li; Caitlin G Knowles; Amanda Mills; Landon Grace; Jesse S Jur
Journal:  Adv Healthc Mater       Date:  2021-07-01       Impact factor: 11.092

  4 in total