Intraoperative testing of the implantable cardioverter-defibrillator: how much is enough?

Luis A Pires1, Katrina M Johnson.   

Abstract

INTRODUCTION: Defibrillation testing of the implantable cardioverter-defibrillator (ICD) is considered a standard and required practice at the time of implantation. How much testing, if any in some cases, should be performed, however, remains unknown. METHODS AND
RESULTS: Included in this retrospective analysis were 835 patients (77% men; age 65 +/- 13 years) who received transvenous ICDs between January 1996 and December 2003. One hundred twenty-nine (15.5%) had intraoperative defibrillation threshold (DFT) testing, 503 (60.2%) had limited defibrillation safety margin testing, and 203 (24.3%) had no defibrillation testing. We compared the outcome (success of ICD therapies against spontaneous VT/VF events and survival) of the three groups of patients, who in some respects had important clinical differences. The success of the first delivered shocks against VT/VF was similar for DFT (91%), safety margin testing (91%), and no-testing (92%) groups; and the second shocks terminated the remaining episodes in all three groups. Sudden-death-free survival rates were similar in the three groups, however, the overall long-term survival rate was significantly lower in the no-testing group (58%) than in the DFT (74%) and safety margin testing (69%) groups (P < 0.0005). Multivariate analysis found no strong predictors of sudden death, but there were several independent predictors of overall mortality including lack of ICD testing (HR: 2.031, CI: 1.253-3.290, P = 0.004).
CONCLUSION: In this select patient cohort, success of ICD therapies and sudden-death-free survival were similar in patients who had DFT, safety margin testing, and no testing, but overall survival was significantly lower in the no-testing group. Thus in the absence of prospective mortality data, a minimum of safety margin ICD testing should remain standard practice.

Entities:  

Mesh:

Year:  2006        PMID: 16533250     DOI: 10.1111/j.1540-8167.2005.00294.x

Source DB:  PubMed          Journal:  J Cardiovasc Electrophysiol        ISSN: 1045-3873


  14 in total

1.  [Is intraoperative ICD-testing still necessary?].

Authors:  C Mewis; H-R Neuberger; A Buob
Journal:  Herzschrittmacherther Elektrophysiol       Date:  2010-06

Review 2.  Complications of defibrillation testing. A review of the literature.

Authors:  Bishoy Deif; Jeff S Healey
Journal:  Herzschrittmacherther Elektrophysiol       Date:  2015-11-30

3.  Incidence of very high defibrillation thresholds (DFT) and efficacy of subcutaneous (SQ) array insertion during implantable cardioverter defibrillator (ICD) implantation.

Authors:  Atul Verma; Alexander J Kaplan; Bradley Sarak; Richard Oosthuizen; Marianne Beardsall; Jan Higgenbottam; Zaev Wulffhart; Yaariv Khaykin
Journal:  J Interv Card Electrophysiol       Date:  2010-09-24       Impact factor: 1.900

4.  ICD implantation without intraoperative testing does not increase the rate of system modifications and does not impair defibrillation efficacy tested in follow-up.

Authors:  Dirk Bastian; Stefan Kracker; Matthias Pauschinger; Konrad Göhl
Journal:  Herzschrittmacherther Elektrophysiol       Date:  2013-06-07

5.  Defibrillation threshold testing fails to show clinical benefit during long-term follow-up of patients undergoing cardiac resynchronization therapy defibrillator implantation.

Authors:  Yoav Michowitz; Nicolas Lellouche; Tahmeed Contractor; Tara Bourke; Isaac Wiener; Eric Buch; Noel Boyle; Malcolm Bersohn; Kalyanam Shivkumar
Journal:  Europace       Date:  2011-01-19       Impact factor: 5.214

6.  How to Manage a High Defibrillation Threshold in ICD Patients: and Does it Really Matter?

Authors:  Maria Vittoria Matassini; Jeff S Healey
Journal:  Curr Treat Options Cardiovasc Med       Date:  2013-08

7.  Defibrillation Testing During Defibrillator Implantation.

Authors:  Dietmar Bänsch
Journal:  Arrhythm Electrophysiol Rev       Date:  2012-09

8.  Intra-operative defibrillation testing and clinical shock efficacy in patients with implantable cardioverter-defibrillators: the NORDIC ICD randomized clinical trial.

Authors:  Dietmar Bänsch; Hendrik Bonnemeier; Johan Brandt; Frank Bode; Jesper Hastrup Svendsen; Miloš Táborský; Stefan Kuster; Carina Blomström-Lundqvist; Angelika Felk; Tino Hauser; Anna Suling; Karl Wegscheider
Journal:  Eur Heart J       Date:  2015-06-25       Impact factor: 29.983

9.  Intraoperative defibrillation threshold testing and postoperative long-term efficacy of cardioverter-defibrillator implantation.

Authors:  Tianyi Gan; Xiaozhi Cao; Zhang Yu; Baopeng Tang; Jinxin Li; Guojun Xu; Xianhui Zhou; Yanyi Zhang; Yaodong Li; Jianghua Zhang
Journal:  Exp Ther Med       Date:  2012-11-05       Impact factor: 2.447

10.  Defibrillation testing of the implantable cardioverter defibrillator: when, how, and by whom?

Authors:  Luis A Pires
Journal:  Indian Pacing Electrophysiol J       Date:  2007-08-01
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.