AIM: To compare the photographic screening for retinopathy of prematurity (ROP) using RetCam 120 with binocular indirect ophthalmoscope (BIO), which is the current gold standard. SETTING AND
DESIGN: Prospective, comparative study.
MATERIALS AND METHODS: A total of 87 RetCam examinations were performed on 27 premature babies. They were stored in a separate file after deleting the identifying information. At the same visit using the BIO with scleral depression, an experienced vitreoretinal surgeon evaluated the fundus in detail. A masked examiner then evaluated the RetCam photographs for presence or absence of ROP, the stage and zone of the disease and the presence or absence of plus disease. These data were then compared with the BIO findings to determine the sensitivity, specificity and the positive and negative predictive values of the method.
RESULTS: ROP was detected in 63 of 87 examinations by BIO and in 56 of 87 RetCam examinations. Nine RetCam examinations were false-negative and two were false-positive. Sensitivity of RetCam was 85.71% (54/63) and specificity was 91.66% (22/24). The positive and negative predictive values were 96.43% and 70.97% respectively.
CONCLUSION: Nine cases having ROP were missed by the RetCam. All these cases were either in zone 3 or the outer part of zone 2, which later regressed. These were missed mostly because of the restricted mobility of the camera head caused by its size and the barrier caused by the lid speculum arms. No case of threshold ROP was missed. RetCam may replace BIO for screening of ROP.

Entities:  

Mesh:

Year:  2006        PMID: 16531668     DOI: 10.4103/0301-4738.21612

Source DB:  PubMed          Journal:  Indian J Ophthalmol        ISSN: 0301-4738            Impact factor:   1.848


  11 in total

Review 1.  Telemedicine for retinopathy of prematurity diagnosis: evaluation and challenges.

Authors:  Grace M Richter; Steven L Williams; Justin Starren; John T Flynn; Michael F Chiang
Journal:  Surv Ophthalmol       Date:  2009-08-08       Impact factor: 6.048

2.  Development of 3D Printed Smartphone-Based Multi-Purpose Fundus Camera (MultiScope) for Retinopathy of Prematurity.

Authors:  Arivazhagan Pugalendhi; Rajesh Ranganathan
Journal:  Ann Biomed Eng       Date:  2021-11-12       Impact factor: 3.934

3.  Stanford University Network for Diagnosis of Retinopathy of Prematurity (SUNDROP): 18-month experience with telemedicine screening.

Authors:  Ruwan A Silva; Yohko Murakami; Atul Jain; Jarel Gandhi; Eleonora M Lad; Darius M Moshfeghi
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2008-09-11       Impact factor: 3.117

4.  Speed of telemedicine vs ophthalmoscopy for retinopathy of prematurity diagnosis.

Authors:  Grace M Richter; Grace Sun; Thomas C Lee; R V Paul Chan; John T Flynn; Justin Starren; Michael F Chiang
Journal:  Am J Ophthalmol       Date:  2009-04-18       Impact factor: 5.258

5.  Retinopathy of prematurity management using single-image vs multiple-image telemedicine examinations.

Authors:  Alexandra Lajoie; Susan Koreen; Lu Wang; Steven A Kane; Thomas C Lee; David J Weissgold; Audina M Berrocal; Yunling E Du; Osode Coki; John T Flynn; Justin Starren; Michael F Chiang
Journal:  Am J Ophthalmol       Date:  2008-06-10       Impact factor: 5.258

6.  Large spot transpupillary thermotherapy: a quicker laser for treatment of high risk prethreshold retinopathy of prematurity - a randomized study.

Authors:  Parag K Shah; V Narendran; N Kalpana
Journal:  Indian J Ophthalmol       Date:  2011 Mar-Apr       Impact factor: 1.848

7.  Diagnostic accuracy of a digital fundus photographic system for detection of retinopathy of prematurity requiring treatment (ROP-RT).

Authors:  Phanthipha Wongwai; Sirinya Suwannaraj; Somkiat Asawaphureekorn
Journal:  PLoS One       Date:  2018-07-31       Impact factor: 3.240

8.  MII RetCam assisted smartphone based fundus imaging for retinopathy of prematurity.

Authors:  T Lekha; S Ramesh; Ashish Sharma; G Abinaya
Journal:  Indian J Ophthalmol       Date:  2019-06       Impact factor: 1.848

9.  The KIDROP model of combining strategies for providing retinopathy of prematurity screening in underserved areas in India using wide-field imaging, tele-medicine, non-physician graders and smart phone reporting.

Authors:  Anand Vinekar; Clare Gilbert; Mangat Dogra; Mathew Kurian; Gangadharan Shainesh; Bhujang Shetty; Noel Bauer
Journal:  Indian J Ophthalmol       Date:  2014-01       Impact factor: 1.848

10.  Screening of retinopathy of prematurity: a neglected public health issue.

Authors:  S Puri; Ss Sarpal; M Ashat
Journal:  Ann Med Health Sci Res       Date:  2014-03
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.