Use of the CDI blood parameter monitoring system 500 for continuous blood gas measurement during extracorporeal membrane oxygenation simulation.

Aaron Schreur1, Scott Niles, James Ploessl.   

Abstract

UNLABELLED: The Oximetrix III Opticath (Abbott Critical Care Systems) is used for continuous measurement of venous saturation in a variety of applications, including extracorporeal membrane oxygenation (ECMO), despite clinical reports that have presented data showing poor accuracy of these devices. The CDI Blood Parameter Monitoring System 500 (Terumo) is an inline blood gas monitoring tool commonly used during cardiopulmonary bypass procedures to continuously assess oxygen saturation, blood gases, potassium, and bicarbonate. The purpose of this experiment was to compare the Opticath and the CDI 500 in trending venous blood saturation during a simulation of ECMO. An ECMO simulation circuit consisting of a silicone rubber membrane oxygenator and a stainless steel heat exchanger was constructed, and a standard venous reservoir bag was used to represent the patient. The CDI and the Opticath were incorporated side by side into a shunt that originated just before the oxygenator and returned flow to the venous line. The circuit was primed with fresh porcine blood and conditioned with the addition of CO2 to simulate typical venous blood under ECMO conditions. After an initial calibration procedure, samples were drawn and analyzed by an AVL Opti CCA (Roche/Osmetech) every 4-8 hours for a period of 7 days, with calibration of each device at sample intervals. The data were plotted, and a least squares regression line was calculated. The average error for venous saturation of the CDI and Opticath after 72 hours was 3.86 and 9.51 respectively. At 168 hours, error for the CDI was 8.37, and the Opticath had an error of 14.78. A correlation analysis of the CDI and AVL CCA analyzer yielded a correlation coefficient of r = .88 at 72 hours and r = .84 at 168 hours. Correlation between the Opticath and the AVL CCA yielded a correlation coefficient of r = .77 at 72 hours and r = .55 at 168 hours. Based on these findings, the CDI 500 is an effective tool for monitoring venous blood saturation under simulated conditions of ECMO. KEYWORDS: CDI 500, Opticath, extracorporeal membrane oxygenation, venous saturation.

Entities:  

Mesh:

Year:  2005        PMID: 16524156      PMCID: PMC4680830     

Source DB:  PubMed          Journal:  J Extra Corpor Technol        ISSN: 0022-1058


  4 in total

1.  Clinical evaluation of a new in-line continuous blood gas monitor.

Authors:  R Southworth; R Sutton; S Mize; A H Stammers; L W Fristoe; D Cook; D Hostetler; W E Richenbacher
Journal:  J Extra Corpor Technol       Date:  1998-12

2.  The effects of continuous blood gas monitoring during cardiopulmonary bypass: a prospective, randomized study--Part I.

Authors:  C C Trowbridge; M Vasquez; A H Stammers; K Glowgowski; K Tremain; K Niimi; M Muhle; T Yiang
Journal:  J Extra Corpor Technol       Date:  2000-09

3.  CDI Blood Parameter Monitoring System 500--a new tool for the clinical perfusionist.

Authors:  D W Fried; J J Leo; G J Mattioni; H Mohamed; R P Rector; F Weber; T L Zombolas
Journal:  J Extra Corpor Technol       Date:  2000-03

4.  A laboratory comparison of three pulmonary artery oximetry catheters.

Authors:  P E Scuderi; D A MacGregor; D L Bowton; R L James
Journal:  Anesthesiology       Date:  1994-07       Impact factor: 7.892

  4 in total
  3 in total

1.  In vitro comparison of the new in-line monitor BMU 40 versus a conventional laboratory analyzer.

Authors:  F Oliver Grosse; David Holzhey; Volkmar Falk; Jan Schaarschmidt; Klaus Kraemer; Friedrich W Mohr
Journal:  J Extra Corpor Technol       Date:  2010-03

2.  Spectrum Medical Quantum or Terumo CDI 500: Which Device Measures Hemoglobin and Oxygen Saturation Most Accurately When Compared to a Benchtop Blood Analyzer?

Authors:  James A Reagor; Zhiqian Gao; James S Tweddell
Journal:  J Extra Corpor Technol       Date:  2021-09

3.  Assessment of a continuous blood gas monitoring system in animals during circulatory stress.

Authors:  Sandro Gelsomino; Roberto Lorusso; Ugolino Livi; Stefano Romagnoli; Salvatore Mario Romano; Rocco Carella; Fabiana Lucà; Giuseppe Billè; Francesco Matteucci; Attilio Renzulli; Gil Bolotin; Giuseppe De Cicco; Pierluigi Stefàno; Jos Maessen; Gian Franco Gensini
Journal:  BMC Anesthesiol       Date:  2011-01-11       Impact factor: 2.217

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.