Literature DB >> 16516777

Comparative clinical trial of ultrasound phacoemulsification with and without the WhiteStar system.

William Fishkind1, Brock Bakewell, Eric D Donnenfeld, Aron D Rose, L Andrew Watkins, Randall J Olson.   

Abstract

PURPOSE: To compare the postoperative outcomes of cataract surgery performed with the Sovereign 4.0 system or the Sovereign system with WhiteStar power modulation.
SETTING: Ambulatory surgery centers at 4 sites in the United States.
METHODS: This was a 3-month, open-label, randomized, parallel-group comparative clinical trial in patients with visually interfering cataract. Surgeons used the divide-and-conquer phacoemulsification technique for nuclear removal. An intraocular lens was implanted using the recommended insertion system. Primary operative outcome measures were equivalent phaco time (EPT), percentage of phaco power, amount of balanced salt solution used, and surgical complications. Patients were seen 1 day and 3 months after surgery. Postoperative outcome measures were the change in endothelial cell count from the preoperative visit to 3 months, corneal clarity, inflammation, and corneal thickness.
RESULTS: The Sovereign with WhiteStar group had 48 patients and the Sovereign 4.0 group, 49 patients. Mean EPT and mean percentage of power were significantly lower in the Sovereign with WhiteStar group. Mean EPT was 6.67 seconds +/- 8.2 (SD) in the Sovereign with WhiteStar group and 8.59 +/- 9.3 seconds (P = .01) in the Sovereign 4.0 group. Mean percentage of phaco power was 6.41% +/- 3.3% in the Sovereign with WhiteStar group and 8.51% +/- 4.9% in the Sovereign 4.0 group (P = .01). The Sovereign with WhiteStar group lost significantly fewer endothelial cells (-319.6 +/- 634.2 cells/mm(2)) than the Sovereign 4.0 group (-430.3 +/- 594.6 cells/mm(2)) (P = .01). Corneal clarity, cells and flare, and pachymetry were comparable with the exception of the 3-month visit. The mean change in baseline central pachymetry showed significantly less corneal thickening in the Sovereign with WhiteStar group.
CONCLUSIONS: The Sovereign with WhiteStar power modulation system provides effective lens removal at lower levels of phaco power and ultrasound energy than the Sovereign 4.0 system. Lower ultrasound levels may reduce the risk for endothelial cell loss during phacoemulsification.

Entities:  

Mesh:

Year:  2006        PMID: 16516777     DOI: 10.1016/j.jcrs.2005.10.026

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  8 in total

1.  Comparative clinical study of Whitestar Signature phacoemulsification system with standard and Ellips FX handpieces.

Authors:  Daniele Tognetto; Rossella D'Aloisio; Paolo Cecchini; Marta Di Nicola; Giuseppe Di Martino
Journal:  Int Ophthalmol       Date:  2017-07-13       Impact factor: 2.031

2.  Comparison of corneal endothelial changes following phacoemulsification with transversal and torsional phacoemulsification machines.

Authors:  Mustafa Ataş; Süleyman Demircan; Arzu Seyhan Karatepe Haşhaş; Ahmet Gülhan; Gökmen Zararsız
Journal:  Int J Ophthalmol       Date:  2014-10-18       Impact factor: 1.779

3.  Comparison of surgical parameters using different lens fragmentation patterns in eyes undergoing laser-assisted cataract surgery.

Authors:  Harvey S Uy; Pik Sha Chan; Raquel Gil-Cazorla; Sunil Shah
Journal:  Int Ophthalmol       Date:  2019-03-04       Impact factor: 2.031

4.  Differences in energy and corneal endothelium between femtosecond laser-assisted and conventional cataract surgeries: prospective, intraindividual, randomized controlled trial.

Authors:  Lucia Bascaran; Txomin Alberdi; Itziar Martinez-Soroa; Cristina Sarasqueta; Javier Mendicute
Journal:  Int J Ophthalmol       Date:  2018-08-18       Impact factor: 1.779

5.  An evaluation of intraoperative and postoperative outcomes of torsional mode versus longitudinal ultrasound mode phacoemulsification: a Meta-analysis.

Authors:  Pia Leon; Ingrid Umari; Alessandro Mangogna; Andrea Zanei; Daniele Tognetto
Journal:  Int J Ophthalmol       Date:  2016-06-18       Impact factor: 1.779

6.  Comparison of Femtosecond Laser-Assisted and Ultrasound-Assisted Cataract Surgery with Focus on Endothelial Analysis.

Authors:  Anna Schroeter; Martina Kropp; Zeljka Cvejic; Gabriele Thumann; Bojan Pajic
Journal:  Sensors (Basel)       Date:  2021-02-02       Impact factor: 3.576

7.  Transverse vs torsional ultrasound: prospective randomized contralaterally controlled study comparing two phacoemulsification-system handpieces.

Authors:  Kerry K Assil; Lindsay Harris; Jeannie Cecka
Journal:  Clin Ophthalmol       Date:  2015-08-03

8.  All-laser bladeless cataract surgery, combining femtosecond and nanosecond lasers: a novel surgical technique.

Authors:  Anastasios John Kanellopoulos
Journal:  Clin Ophthalmol       Date:  2013-09-13
  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.