A method for automatic identification of reliable heart rates calculated from ECG and PPG waveforms.

Chenggang Yu1, Zhenqiu Liu, Thomas McKenna, Andrew T Reisner, Jaques Reifman.   

Abstract

OBJECTIVE: The development and application of data-driven decision-support systems for medical triage, diagnostics, and prognostics pose special requirements on physiologic data. In particular, that data are reliable in order to produce meaningful results. The authors describe a method that automatically estimates the reliability of reference heart rates (HRr) derived from electrocardiogram (ECG) waveforms and photoplethysmogram (PPG) waveforms recorded by vital-signs monitors. The reliability is quantitatively expressed through a quality index (QI) for each HRr.
DESIGN: The proposed method estimates the reliability of heart rates from vital-signs monitors by (1) assessing the quality of the ECG and PPG waveforms, (2) separately computing heart rates from these waveforms, and (3) concisely combining this information into a QI that considers the physical redundancy of the signal sources and independence of heart rate calculations. The assessment of the waveforms is performed by a Support Vector Machine classifier and the independent computation of heart rate from the waveforms is performed by an adaptive peak identification technique, termed ADAPIT, which is designed to filter out motion-induced noise.
RESULTS: The authors evaluated the method against 158 randomly selected data samples of trauma patients collected during helicopter transport, each sample consisting of 7-second ECG and PPG waveform segments and their associated HRr. They compared the results of the algorithm against manual analysis performed by human experts and found that in 92% of the cases, the algorithm either matches or is more conservative than the human's QI qualification. In the remaining 8% of the cases, the algorithm infers a less conservative QI, though in most cases this was because of algorithm/human disagreement over ambiguous waveform quality. If these ambiguous waveforms were relabeled, the misclassification rate would drop from 8% to 3%.
CONCLUSION: This method provides a robust approach for automatically assessing the reliability of large quantities of heart rate data and the waveforms from which they are derived.

Entities:  

Mesh:

Year:  2006        PMID: 16501184      PMCID: PMC1513657          DOI: 10.1197/jamia.M1925

Source DB:  PubMed          Journal:  J Am Med Inform Assoc        ISSN: 1067-5027            Impact factor:   4.497


  7 in total

Review 1.  Trauma scoring systems: a review.

Authors:  C K Senkowski; M G McKenney
Journal:  J Am Coll Surg       Date:  1999-11       Impact factor: 6.113

2.  The principles of software QRS detection.

Authors:  Bert-Uwe Köhler; Carsten Hennig; Reinhold Orglmeister
Journal:  IEEE Eng Med Biol Mag       Date:  2002 Jan-Feb

3.  Combat medical informatics: present and future.

Authors:  Reed W Hoyt; Jaques Reifman; Trinka S Coster; Mark J Buller
Journal:  Proc AMIA Symp       Date:  2002

4.  Support vector machine-based expert system for reliable heartbeat recognition.

Authors:  Stanislaw Osowski; Linh Tran Hoai; Tomasz Markiewicz
Journal:  IEEE Trans Biomed Eng       Date:  2004-04       Impact factor: 4.538

5.  Manual vital signs reliably predict need for life-saving interventions in trauma patients.

Authors:  John B Holcomb; Jose Salinas; John M McManus; Charles C Miller; William H Cooke; Victor A Convertino
Journal:  J Trauma       Date:  2005-10

6.  Heart rate variability and its association with mortality in prehospital trauma patients.

Authors:  William H Cooke; Jose Salinas; Victor A Convertino; David A Ludwig; Denise Hinds; James H Duke; Fredrick A Moore; John B Holcomb
Journal:  J Trauma       Date:  2006-02

7.  Prehospital Index: a scoring system for field triage of trauma victims.

Authors:  J J Koehler; L J Baer; S A Malafa; M S Meindertsma; N R Navitskas; J E Huizenga
Journal:  Ann Emerg Med       Date:  1986-02       Impact factor: 5.721

  7 in total
  6 in total

1.  Deep learning approaches for plethysmography signal quality assessment in the presence of atrial fibrillation.

Authors:  Tania Pereira; Cheng Ding; Kais Gadhoumi; Nate Tran; Rene A Colorado; Karl Meisel; Xiao Hu
Journal:  Physiol Meas       Date:  2019-12-27       Impact factor: 2.833

2.  Development and validation of a portable platform for deploying decision-support algorithms in prehospital settings.

Authors:  A T Reisner; M Y Khitrov; L Chen; A Blood; K Wilkins; W Doyle; S Wilcox; T Denison; J Reifman
Journal:  Appl Clin Inform       Date:  2013-08-21       Impact factor: 2.342

3.  Automatic pre-hospital vital signs waveform and trend data capture fills quality management, triage and outcome prediction gaps.

Authors:  Colin F Mackenzie; Peter Hu; Ayan Sen; Rick Dutton; Steve Seebode; Doug Floccare; Tom Scalea
Journal:  AMIA Annu Symp Proc       Date:  2008-11-06

4.  Robust detection of heart beats in multimodal data.

Authors:  Ikaro Silva; Benjamin Moody; Joachim Behar; Alistair Johnson; Julien Oster; Gari D Clifford; George B Moody
Journal:  Physiol Meas       Date:  2015-07-28       Impact factor: 2.833

5.  Treatment of rabbit elastase-induced aneurysm models by flow diverters: development of quantifiable indexes of device performance using digital subtraction angiography.

Authors:  Chander Sadasivan; Liliana Cesar; Jaehoon Seong; Ajay K Wakhloo; Baruch B Lieber
Journal:  IEEE Trans Med Imaging       Date:  2009-01-19       Impact factor: 10.048

6.  A novel diversity method for smartphone camera-based heart rhythm signals in the presence of motion and noise artifacts.

Authors:  Fatemehsadat Tabei; Rifat Zaman; Kamrul H Foysal; Rajnish Kumar; Yeesock Kim; Jo Woon Chong
Journal:  PLoS One       Date:  2019-06-19       Impact factor: 3.240

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.