Intraocular lens power calculation using ray tracing following excimer laser surgery.

T M Rabsilber1, A J Reuland, M P Holzer, G U Auffarth.   

Abstract

PURPOSE: To evaluate intraocular lens (IOL) power calculation using ray tracing in patients presenting with cataract after excimer laser surgery.
METHODS: Ten eyes of seven consecutive patients who presented for cataract surgery following excimer laser treatment without any pre-refractive biometry data were enrolled in this prospective clinical study. Preoperatively, IOL power calculation was performed using a ray tracing software called OKULIX. Keratometry data (C-Scan) were imported and axial length (IOLMaster) was entered manually. Accuracy of IOL power calculation was investigated by subtracting attempted and achieved spherical equivalent.
RESULTS: Mean spherical equivalent was -3.51+/-2.77 D (range -10.38 to -0.5 D) preoperatively and -1.01+/-1.08 D (range -2.5 to +0.75 D) postoperatively. Mean error was 0.31+/-0.84 D, mean absolute error was 0.74+/-0.46 D, and IOL calculation errors ranged from -1.39 to +1.47 D. A total of 40% of eyes were within +/-0.5 D, 70% within +/-1.0 D, and 100% within +/-1.5 D. Three eyes with corneal radii over 10 mm showed calculation errors exceeding +/-1.0 D. Mean best-corrected visual acuity increased from 20/60 to 20/30 postoperatively.
CONCLUSIONS: IOL power calculation after excimer laser surgery can be difficult, especially when pre-refractive keratometry values are not available. In these cases, ray tracing combined with corneal topography measurements provides reliable and satisfactory postoperative results. However, it is advisable to be careful when calculating IOL power for eyes with corneal radii exceeding 10 mm because of slightly higher prediction errors.

Entities:  

Mesh:

Year:  2006        PMID: 16498433     DOI: 10.1038/sj.eye.6702300

Source DB:  PubMed          Journal:  Eye (Lond)        ISSN: 0950-222X            Impact factor:   3.775


  8 in total

1.  Ray tracing software for intraocular lens power calculation after corneal excimer laser surgery.

Authors:  Megumi Saiki; Kazuno Negishi; Naoko Kato; Hidemasa Torii; Murat Dogru; Kazuo Tsubota
Journal:  Jpn J Ophthalmol       Date:  2014-02-25       Impact factor: 2.447

2.  Optical coherence tomography-based corneal power measurement and intraocular lens power calculation following laser vision correction (an American Ophthalmological Society thesis).

Authors:  David Huang; Maolong Tang; Li Wang; Xinbo Zhang; Rebecca L Armour; Devin M Gattey; Lorinna H Lombardi; Douglas D Koch
Journal:  Trans Am Ophthalmol Soc       Date:  2013-09

3.  Comparison of OKULIX ray-tracing software with SRK-T and Hoffer-Q formula in intraocular lens power calculation.

Authors:  Mohammad Ghoreyshi; Ahmadreza Khalilian; Mohammadreza Peyman; Mohaddeseh Mohammadinia; Alireza Peyman
Journal:  J Curr Ophthalmol       Date:  2017-10-27

4.  Development of a New Method for Calculating Intraocular Lens Power after Myopic Laser In Situ Keratomileusis by Combining the Anterior-Posterior Ratio of the Corneal Radius of the Curvature with the Double-K Method.

Authors:  Yoshihiko Iida; Kimiya Shimizu; Nobuyuki Shoji
Journal:  J Clin Med       Date:  2022-01-20       Impact factor: 4.241

5.  IOL Power Calculations and Cataract Surgery in Eyes with Previous Small Incision Lenticule Extraction.

Authors:  Roman Lischke; Walter Sekundo; Rainer Wiltfang; Martin Bechmann; Thomas C Kreutzer; Siegfried G Priglinger; Martin Dirisamer; Nikolaus Luft
Journal:  J Clin Med       Date:  2022-07-29       Impact factor: 4.964

Review 6.  Intraocular lens power calculation following laser refractive surgery.

Authors:  Christopher Hodge; Colm McAlinden; Michael Lawless; Colin Chan; Gerard Sutton; Aifric Martin
Journal:  Eye Vis (Lond)       Date:  2015-04-02

7.  Improving accuracy of corneal power measurement with partial coherence interferometry after corneal refractive surgery using a multivariate polynomial approach.

Authors:  Michele Lanza; Robert Koprowski; Mario Bifani Sconocchia
Journal:  Biomed Eng Online       Date:  2018-08-13       Impact factor: 2.819

Review 8.  Intraocular lens power calculation in eyes with previous corneal refractive surgery.

Authors:  Giacomo Savini; Kenneth J Hoffer
Journal:  Eye Vis (Lond)       Date:  2018-07-08