Bassem Hiba1, Nathalie Richard, Marc Janier, Pierre Croisille.
Abstract
ECG-gated cardiac MRI in the mouse is hindered by many technical difficulties in ECG signal recording inside static and variable high magnetic scanner fields. The present study proposes an alternative robust method of acquiring auto-gated cardiac and respiratory cine images in mouse heart. In our approach, a motion synchronization signal is extracted from the echo peak MR signal of a non-triggered radial acquisition. This signal is then used for both cardiac and respiratory retrospective gating before cine image reconstruction. Highly asymmetric echoes were acquired to achieve the radial k-space sampling in order to avoid radial acquisition related artifacts and to increase auto-gating robustness. In vivo experiments demonstrated the feasibility and robustness of self-gated cine-MRI in the mouse heart at 7T. The signal-to-noise and contrast-to-noise ratios of the self-gated and ECG-gated images were comparable, all parameters being equal. Magn Reson Med, 2006. (c) 2006 Wiley-Liss, Inc.Entities:
Mesh:
Year: 2006 PMID: 16463350 DOI: 10.1002/mrm.20815