Axial cyclic behavior of the bone-screw interface.

Serkan Inceoğlu1, Mike Ehlert, Atilla Akbay, Robert F McLain.   

Abstract

Screw fixation strength is investigated by using a pullout test. Despite many screw pullout studies, the effects of loading rate on the pullout behavior of pedicle screws are not known. The objective of this study was to assess the effects of loading rate on the pullout stiffness and strength of pedicle screws. Sixty pedicle screws were inserted in foam blocks and pulled out at four different rates: 0.1, 1, 5 and 50 mm/min. Twenty of these 60 screws were cycled non-destructively at four different rates sequentially, i.e., 0.1, 1, 5 and 50 mm/min prior to pullout. Ten additional pedicle screws were inserted in five calf lumbar vertebrae, cycled as in foam group, and pulled out at a rate of either 0.1 or 50 mm/min. The results showed that the stiffness was higher at all rates compared to 0.1 mm/min in foam model but in bone model only 1 and 5 mm/min groups were higher compared to 0.1 mm/min. The pullout strength in 50 mm/min group was higher than that in 0.1 mm/min group in both foam and bone model. The results suggested that loading rate influenced the mechanics of the bone-screw interface. Therefore, a fair comparison between the pullout studies can be achieved under same loading rate conditions. Moreover, the cycling of the pedicle screws in axial direction within a pre-yield region showed an unusual hysteresis curve. Further studies are needed for a better understanding of the mechanics of the screw-bone interface.

Entities:  

Mesh:

Year:  2006        PMID: 16458568     DOI: 10.1016/j.medengphy.2005.12.009

Source DB:  PubMed          Journal:  Med Eng Phys        ISSN: 1350-4533            Impact factor:   2.242


  4 in total

1.  The contribution of the cortical shell to pedicle screw fixation.

Authors:  Matthew Henry Pelletier; Nicky Bertollo; Darweesh Al-Khawaja; William Robert Walsh
Journal:  J Spine Surg       Date:  2017-06

2.  Effect of pilot hole on biomechanical and in vivo pedicle screw-bone interface.

Authors:  Patrícia Silva; Rodrigo César Rosa; Antonio Carlos Shimano; Helton L A Defino
Journal:  Eur Spine J       Date:  2013-05-08       Impact factor: 3.134

3.  Vertebral Trabecular Bone Mechanical Properties Vary Among Functional Groups of Cetaceans.

Authors:  D N Ingle; M E Porter
Journal:  Integr Org Biol       Date:  2022-01-07

4.  Effects of the Insertion Type and Depth on the Pedicle Screw Pullout Strength: A Finite Element Study.

Authors:  K Jendoubi; Y Khadri; M Bendjaballah; N Slimane
Journal:  Appl Bionics Biomech       Date:  2018-07-26       Impact factor: 1.781

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.