Literature DB >> 16446099

Improving the quality of the ion beam exiting a quadrupole ion guide.

Vadym D Berkout1, Vladimir M Doroshenko.   

Abstract

The effect of fringing fields on the divergence of the ion beam exiting an RF quadrupole ion guide was studied using a computer simulation. It was shown that reducing the strength of the RF field towards the ion guide exit reduces ion beam divergence. Further improvement was demonstrated when creating a DC gradient towards the exit. The results of the numerical simulation were verified experimentally using a time-of-flight (TOF) mass analyzer with orthogonal acceleration. Decreasing the ion beam divergence resulted in considerably improved mass resolution of the instrument.

Mesh:

Substances:

Year:  2006        PMID: 16446099      PMCID: PMC1839038          DOI: 10.1016/j.jasms.2005.12.002

Source DB:  PubMed          Journal:  J Am Soc Mass Spectrom        ISSN: 1044-0305            Impact factor:   3.109


  3 in total

Review 1.  An introduction to quadrupole-time-of-flight mass spectrometry.

Authors:  I V Chernushevich; A V Loboda; B A Thomson
Journal:  J Mass Spectrom       Date:  2001-08       Impact factor: 1.982

2.  Mass selective axial ion ejection from a linear quadrupole ion trap.

Authors:  F A Londry; James W Hager
Journal:  J Am Soc Mass Spectrom       Date:  2003-10       Impact factor: 3.109

3.  Collisional focusing effects in radio frequency quadrupoles.

Authors:  D J Douglas; J B French
Journal:  J Am Soc Mass Spectrom       Date:  1992-05       Impact factor: 3.109

  3 in total
  4 in total

1.  Characterization of the ion beam focusing in a mass spectrometer using an IonCCD™ detector.

Authors:  Grant E Johnson; Omar Hadjar; Julia Laskin
Journal:  J Am Soc Mass Spectrom       Date:  2011-05-19       Impact factor: 3.109

2.  Optimized Electrostatic Linear Ion Trap for Charge Detection Mass Spectrometry.

Authors:  Joanna A Hogan; Martin F Jarrold
Journal:  J Am Soc Mass Spectrom       Date:  2018-07-09       Impact factor: 3.109

3.  Dramatic Improvement in Sensitivity with Pulsed Mode Charge Detection Mass Spectrometry.

Authors:  Aaron R Todd; Martin F Jarrold
Journal:  Anal Chem       Date:  2019-10-07       Impact factor: 6.986

4.  Current literature in mass spectrometry.

Authors: 
Journal:  J Mass Spectrom       Date:  2007-01       Impact factor: 1.982

  4 in total