Epicardial beating heart cryoablation using a novel argon-based cryoclamp and linear probe.

Federico Milla1, Nikolaos Skubas, William M Briggs, Leonard N Girardi, Leonard Y Lee, Wilson Ko, Anthony J Tortolani, Karl H Krieger, O Wayne Isom, Charles A Mack.   

Abstract

OBJECTIVE: Epicardial, beating heart cryoablation for the treatment of atrial fibrillation may be limited by heat from intracardiac blood flow. We therefore evaluated the ability to create cryolesions using an argon-based cryoclamp device, which temporarily occludes blood flow and facilitates transmurality.
METHODS: Six mongrel dogs underwent sternotomy. A clamp employing a 10-cm argon-based linear cryoablation device was used epicardially to isolate the pulmonary veins and left atrial appendage. After clamping of lesions, the probe was removed from the cryoclamp device, and the remaining linear lesions, analogous to the Cox maze III, were performed. Pulmonary vein stenosis was evaluated with the use of magnetic resonance imaging. Left atrial function and pulmonary venous flow velocities were assessed with transesophageal echocardiography. Transmurality was confirmed both electrically and histologically. Animals were then put to death at 30 days.
RESULTS: All acute and chronic cryoclamp lesions produced conduction block. There was no change in right (RPV) or left pulmonary vein (LPV) diameter on the basis of magnetic resonance imaging at baseline and at planned death (RPV-1, 19.6 +/- 2.9 mm vs 16.9 +/- 2.8 mm, P = .22; RPV-2, 13.2 +/- 2.0 mm vs 11.8 +/- 1.6 mm, P = .22; and LPV, 12.2 +/- 2.4 mm vs 11.2 +/- 1.9 mm, P = .30). Left atrial function and pulmonary venous flow velocities were unchanged. Tissue sections determined transmurality in 93% of cryoclamp lesions and 84% of linear ablations performed with the 10-cm malleable probe.
CONCLUSIONS: Epicardial application of this cryoclamp device on the beating heart produced transmural lesions, which persisted 30 days. Linear epicardial cryoablation was not as effective as the cryoclamp device at producing consistent transmural lesions. This novel, versatile device may be useful in treating patients with atrial fibrillation on the beating heart without cardiopulmonary bypass.

Entities:  

Mesh:

Year:  2006        PMID: 16434271     DOI: 10.1016/j.jtcvs.2005.10.048

Source DB:  PubMed          Journal:  J Thorac Cardiovasc Surg        ISSN: 0022-5223            Impact factor:   5.209


  8 in total

Review 1.  Ablation technology for the surgical treatment of atrial fibrillation.

Authors:  Spencer J Melby; Richard B Schuessler; Ralph J Damiano
Journal:  ASAIO J       Date:  2013 Sep-Oct       Impact factor: 2.872

2.  Biatrial lesion sets.

Authors:  Ralph J Damiano; Rochus K Voeller
Journal:  J Interv Card Electrophysiol       Date:  2007-12       Impact factor: 1.900

Review 3.  Experimental cryosurgery investigations in vivo.

Authors:  A A Gage; J M Baust; J G Baust
Journal:  Cryobiology       Date:  2009-10-13       Impact factor: 2.487

Review 4.  Surgical ablation devices for atrial fibrillation.

Authors:  Shelly C Lall; Ralph J Damiano
Journal:  J Interv Card Electrophysiol       Date:  2007-12       Impact factor: 1.900

5.  Cardiac function in relation to rhythm outcome after intraoperative epicardial left atrial cryoablation.

Authors:  Birgitta Johansson; Birgitta Houltz; Nils Edvardsson; Henrik Scherstén; Thomas Karlsson; Birger Wandt; Eva Berglin
Journal:  Scand Cardiovasc J       Date:  2011-08-05       Impact factor: 1.589

6.  The influence of epicardial and endocardial use of cryoenergy on the completeness of lesions in surgical ablation of atrial fibrillation.

Authors:  Vojtech Kurfirst; Julia Csanady; Ales Mokracek; Jiri Hanis; Alan Bulava; Ladislav Pesl
Journal:  Kardiochir Torakochirurgia Pol       Date:  2020-04-07

7.  Evaluation of a novel cryoablation system: in vivo testing in a chronic porcine model.

Authors:  Timo Weimar; Anson M Lee; Shuddhadeb Ray; Richard B Schuessler; Ralph J Damiano
Journal:  Innovations (Phila)       Date:  2012 Nov-Dec

8.  Evaluation of a novel cryoablation system: in vitro testing of heat capacity and freezing temperatures.

Authors:  Timo Weimar; Anson M Lee; Shuddhadeb Ray; Richard B Schuessler; Ralph J Damiano
Journal:  Innovations (Phila)       Date:  2012 Nov-Dec
  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.