Quantifying the effect of IV contrast media on integrated PET/CT: clinical evaluation.

Osama Mawlawi1, Jeremy J Erasmus, Reginald F Munden, Tinsu Pan, Amy E Knight, Homer A Macapinlac, Donald A Podoloff, Marvin Chasen.   

Abstract

OBJECTIVES: The use of IV contrast media in PET/CT can result in an overestimation of PET attenuation factors that potentially can affect interpretation. The objective of this study was to quantify the effect of IV contrast media in PET/CT and assess its impact on patients with intrathoracic malignancies.
MATERIALS AND METHODS: Nine patients had CTs performed with and without IV contrast media followed by (18)F-FDG PET. PET images were reconstructed using contrast-enhanced and unenhanced CT. To quantify the effect of contrast media on standardized uptake values (SUV), similar regions of interest (ROIs) were drawn on the subclavian vein, heart, liver, spleen, and site of malignancy on both CT and corresponding reconstructed PET images, and the mean and maximum values were compared. In addition, two physicians blinded to the imaging parameters that were used evaluated the reconstructed PET images to assess whether IV contrast media had an effect on clinical interpretation.
RESULTS: For all patient studies, the subclavian vein region on the ipsilateral side of contrast media administration had the highest increase in CT numbers with a corresponding average SUV(max) increase of 27.1%. Similarly, ROIs of the heart and at the site of malignancy showed an increase in the maximum attenuation value with a corresponding average SUV(max) increase of 16.7% and 8.4%, respectively. Other locations had relatively small attenuation value differences with a correspondingly negligible SUV variation.
CONCLUSION: Although there is a significant increase in SUV in regions of high-contrast concentration when contrast-enhanced CT is used for attenuation correction, this increase is clinically insignificant. Accordingly, in PET/CT, IV contrast-enhanced CT can be used in combination with the PET to evaluate patients with cancer.

Entities:  

Mesh:

Substances:

Year:  2006        PMID: 16423932     DOI: 10.2214/AJR.04.1740

Source DB:  PubMed          Journal:  AJR Am J Roentgenol        ISSN: 0361-803X            Impact factor:   3.959


  31 in total

1.  PET/CT in lung cancer: Influence of contrast medium on quantitative and clinical assessment.

Authors:  Florian F Behrendt; Yavuz Temur; Frederik A Verburg; Moritz Palmowski; Thomas Krohn; Hubertus Pietsch; Christiane K Kuhl; Felix M Mottaghy
Journal:  Eur Radiol       Date:  2012-06-04       Impact factor: 5.315

2.  Multiphase contrast-enhanced CT with highly concentrated contrast agent can be used for PET attenuation correction in integrated PET/CT imaging.

Authors:  Philip Aschoff; Christian Plathow; Thomas Beyer; Matthias P Lichy; Gunter Erb; Mehmet Ö Öksüz; Claus D Claussen; Christina Pfannenberg
Journal:  Eur J Nucl Med Mol Imaging       Date:  2011-11-29       Impact factor: 9.236

3.  Comparative assessment of energy-mapping approaches in CT-based attenuation correction for PET.

Authors:  Mohammad R Ay; Maryam Shirmohammad; Saeed Sarkar; Arman Rahmim; Habib Zaidi
Journal:  Mol Imaging Biol       Date:  2011-02       Impact factor: 3.488

4.  Assessment of liver metastases from colorectal adenocarcinoma following chemotherapy: SPIO-MRI versus FDG-PET/CT.

Authors:  L Bacigalupo; S Aufort; M C Eberlé; E Assenat; M Ychou; B Gallix
Journal:  Radiol Med       Date:  2010-06-23       Impact factor: 3.469

5.  Quantitative effects of contrast enhanced CT attenuation correction on PET SUV measurements.

Authors:  Tira Bunyaviroch; Timothy G Turkington; Terence Z Wong; John W Wilson; James G Colsher; R Edward Coleman
Journal:  Mol Imaging Biol       Date:  2007-12-21       Impact factor: 3.488

6.  Contrast medium injection protocol adjusted for body surface area in combined PET/CT.

Authors:  Florian F Behrendt; Marilou Rebière; Andreas Goedicke; Hubertus Pietsch; Karin Palmowski; Christiane K Kuhl; Felix M Mottaghy; Frederik A Verburg
Journal:  Eur Radiol       Date:  2013-02-20       Impact factor: 5.315

Review 7.  Accuracy and precision of radioactivity quantification in nuclear medicine images.

Authors:  Eric C Frey; John L Humm; Michael Ljungberg
Journal:  Semin Nucl Med       Date:  2012-05       Impact factor: 4.446

8.  Low-dose non-enhanced CT versus full-dose contrast-enhanced CT in integrated PET/CT studies for the diagnosis of uterine cancer recurrence.

Authors:  Kazuhiro Kitajima; Kayo Suzuki; Yuji Nakamoto; Yumiko Onishi; Setsu Sakamoto; Michio Senda; Masato Kita; Kazuro Sugimura
Journal:  Eur J Nucl Med Mol Imaging       Date:  2010-04-13       Impact factor: 9.236

9.  Performance of integrated FDG-PET/contrast-enhanced CT in the diagnosis of recurrent pancreatic cancer: comparison with integrated FDG-PET/non-contrast-enhanced CT and enhanced CT.

Authors:  Kazuhiro Kitajima; Koji Murakami; Erena Yamasaki; Yasushi Kaji; Mitsugi Shimoda; Keiichi Kubota; Narufumi Suganuma; Kazuro Sugimura
Journal:  Mol Imaging Biol       Date:  2009-12-01       Impact factor: 3.488

Review 10.  PET/CT imaging: what radiologists need to know.

Authors:  M Benamor; L Ollivier; H Brisse; G Moulin-Romsee; V Servois; S Neuenschwander
Journal:  Cancer Imaging       Date:  2007-10-01       Impact factor: 3.909

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.