Feasibility study using a Ni-Ti stent and electronic portal imaging to localize the prostate during radiotherapy.

Jesper Carl1, Bente Lund, Erik Hoejkjaer Larsen, Jane Nielsen.   

Abstract

BACKGROUND AND
PURPOSE: A new method for localization of the prostate during external beam radiotherapy is presented. The method is based on insertion of a thermo-expandable Ni-Ti stent. The stent is originally developed for treatment of bladder outlet obstruction caused by benign hyperplasia. The radiological properties of the stent are used for precise prostate localization during treatment using electronic portal images. PATIENTS AND METHODS: Patients referred for intended curative radiotherapy and having a length of their prostatic urethra in the range from 25 to 65 mm were included. Pairs of isocentric orthogonal portal images were used to determine the 3D position at eight different treatment sessions for each patient.
RESULTS: Fourteen patients were enrolled in the study. The data obtained demonstrated that the stent position was representative of the prostate location. The stent may also improve delineation of the prostate GTV, and prevent obstruction of bladder outlet during treatment. Precision in localization of the stent was less than 1 mm. Random errors in stent position were left-right 1.6 mm, cranial-caudal 2.2 mm and anterior-posterior 3.2 mm. In four of 14 patients a dislocation of the stent to the bladder occurred. Dislocation only occurred in patients with length of prostatic urethra less than 40 mm.
CONCLUSIONS: A new method for radiological high precision localization of the prostate during radiotherapy is presented. The method is based on insertion of a standard Ni-Ti thermo-expandable stent, designed for treatment of benign prostate hyperplasia.

Entities:  

Mesh:

Substances:

Year:  2006        PMID: 16413623     DOI: 10.1016/j.radonc.2005.11.015

Source DB:  PubMed          Journal:  Radiother Oncol        ISSN: 0167-8140            Impact factor:   6.280


  2 in total

1.  A method for robust segmentation of arbitrarily shaped radiopaque structures in cone-beam CT projections.

Authors:  Per Rugaard Poulsen; Walther Fledelius; Paul J Keall; Elisabeth Weiss; Jun Lu; Emily Brackbill; Geoffrey D Hugo
Journal:  Med Phys       Date:  2011-04       Impact factor: 4.071

2.  Comparison of manual and automatic MR-CT registration for radiotherapy of prostate cancer.

Authors:  Anne Sofie Korsager; Jesper Carl; Lasse Riis Østergaard
Journal:  J Appl Clin Med Phys       Date:  2016-05-08       Impact factor: 2.102

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.