Different patterns of the fall of impedance as the result of heating during ostial pulmonary vein ablation: implications for power titration.

Christopher Reithmann1, Thomas Remp, Ellen Hoffmann, Tomas Matis, Reza Wakili, Gerhard Steinbeck.   

Abstract

BACKGROUND: A variety of strategies have been proposed to avoid the risks of pulmonary vein ablation for atrial fibrillation. The fall of impedance during radiofrequency catheter ablation can be used as a real time measure of tissue heating. The aim of this study was to analyze the impedance fall during ostial pulmonary vein ablation and to evaluate whether adjusting power to the fall of impedance may contribute to a reduction of the risk of complications.
METHODS: Analysis of biophysical parameters of ablation and determination of ostial diameters during follow-up were performed in 70 patients undergoing impedance-guided segmental ostial pulmonary vein ablation. Repeat radiographic angiography, local electrograms, and baseline impedance were the criteria to define the position of the 4-mm electrode tip at atrial sites or inside the proximal pulmonary veins.
RESULTS: Energy application inside the proximal pulmonary veins led to an increased impedance fall inside the first 5-10 mm of the pulmonary veins (1.1 +/- 0.5 Omega/W) as compared to ablation at atrial sites (0.7 +/- 0.3 Omega/W) (P < 0.01). The analysis of temperature and impedance fall during ostial ablation demonstrated an increased impedance fall with heating at sites inside the proximal pulmonary veins (1.5 +/- 0.6 Omega/ degrees C) as compared to atrial sites (1.2 +/- 0.5 Omega/ degrees C) (P < 0.001). The regression lines analyzing these correlations indicated that adjusting power to a maximum impedance fall of 20 Omega would limit heating at pulmonary venous sites to lower temperatures (average maximum temperature: 48 degrees C) than at atrial sites (average maximum temperature: 63 degrees C). The ablation strategy used for segmental ostial ablation in 70 patients, which involved power limitation to a maximum impedance fall of 20 Omega, allowed isolation of 89% of targeted pulmonary veins with a low rate of impedance rises (0.3% of applications). No pulmonary vein stenoses >30% were detected by follow-up computed tomography analysis.
CONCLUSIONS: An increased impedance fall as the result of heating during ostial ablation was found inside the proximal pulmonary veins as compared to atrial sites. Adjusting power to the fall of impedance during segmental ostial pulmonary vein ablation contributes to the prevention of overheating inside the pulmonary veins and may lower the risk of coagulum formation and pulmonary vein stenosis.

Entities:  

Mesh:

Year:  2005        PMID: 16403160     DOI: 10.1111/j.1540-8159.2005.00269.x

Source DB:  PubMed          Journal:  Pacing Clin Electrophysiol        ISSN: 0147-8389            Impact factor:   1.976


  4 in total

1.  Benefit of Contact Force-Guided Catheter Ablation for Treating Premature Ventricular Contractions.

Authors:  Ziming Zhao; Xiaowei Liu; Lianjun Gao; Yutao Xi; Qi Chen; Dong Chang; Xianjie Xiao; Jie Cheng; Yanzong Yang; Yunlong Xia; Xiaomeng Yin
Journal:  Tex Heart Inst J       Date:  2020-02-01

2.  No severe pulmonary vein stenosis after extensive encircling pulmonary vein isolation: 12-month follow-up with 3D computed tomography.

Authors:  Shingo Maeda; Yoshito Iesaka; Kiyoshi Otomo; Kikuya Uno; Yasutoshi Nagata; Kenji Suzuki; Hitoshi Hachiya; Masahiko Goya; Atsushi Takahashi; Hideomi Fujiwara; Mitsuaki Isobe
Journal:  Heart Vessels       Date:  2010-12-04       Impact factor: 2.037

3.  Impact of real-time contact force and impedance measurement in pulmonary vein isolation procedures for treatment of atrial fibrillation.

Authors:  Reza Wakili; Sebastian Clauss; Viola Schmidt; Michael Ulbrich; Anton Hahnefeld; Franziska Schüssler; Johannes Siebermair; Stefan Kääb; Heidi L Estner
Journal:  Clin Res Cardiol       Date:  2013-10-06       Impact factor: 5.460

Review 4.  Impedance-guided Radiofrequency Ablation: Using Impedance to Improve Ablation Outcomes.

Authors:  Jason S Chinitz; Gregory F Michaud; Kent Stephenson
Journal:  J Innov Card Rhythm Manag       Date:  2017-10-15
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.