Literature DB >> 16394343

In vivo investigation of a new 109Cd gamma-ray induced K-XRF bone lead measurement system.

Huiling Nie1, David Chettle, Liqiang Luo, Joanne O'Meara.   

Abstract

A new 109Cd gamma-ray induced K-XRF bone lead measurement system using an array of four detectors has been developed. Previous results from Monte Carlo (MC) simulations and experiments with phantoms predicted that it would be about three times more sensitive than the conventional system, albeit using a more active source. A dosimetry study has been performed for this system and it demonstrated that the dose delivered to the measured individuals is acceptable even for 5-year-old children. Approval to apply this system to human studies has been received from the Research Ethics Board. In this study, 20 adult volunteers, 10 male, 10 female, were recruited to have their tibia measured with both the conventional system and the new system. The result confirmed the improvement predicted by the MC simulations and the in vitro measurements. Two other interesting points were discovered from the data. One is that the data from the new system showed a significant positive correlation between age and tibia lead concentration, while the data from the conventional system do not. The other is that 85% of the tibia lead concentrations were under the minimum detection limit when measured by the conventional system, and the proportion reduced to 50% when measured by the new system.

Entities:  

Mesh:

Substances:

Year:  2006        PMID: 16394343     DOI: 10.1088/0031-9155/51/2/011

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  9 in total

1.  Blood lead levels and cumulative blood lead index (CBLI) as predictors of late neurodevelopment in lead poisoned children.

Authors:  Linda H Nie; Robert O Wright; David C Bellinger; Javed Hussain; Chitra Amarasiriwardena; David R Chettle; Ana Pejović-Milić; Alan Woolf; Michael Shannon
Journal:  Biomarkers       Date:  2011-08-09       Impact factor: 2.658

2.  Theoretical modeling of a portable x-ray tube based KXRF system to measure lead in bone.

Authors:  Aaron J Specht; Marc G Weisskopf; Linda Huiling Nie
Journal:  Physiol Meas       Date:  2017-02-07       Impact factor: 2.833

3.  In vivo quantification of lead in bone with a portable x-ray fluorescence system--methodology and feasibility.

Authors:  L H Nie; S Sanchez; K Newton; L Grodzins; R O Cleveland; M G Weisskopf
Journal:  Phys Med Biol       Date:  2011-01-17       Impact factor: 3.609

Review 4.  Biomarkers of manganese intoxication.

Authors:  Wei Zheng; Sherleen X Fu; Ulrike Dydak; Dallas M Cowan
Journal:  Neurotoxicology       Date:  2010-10-12       Impact factor: 4.294

5.  Feasibility of a portable X-ray fluorescence device for bone lead measurements of condor bones.

Authors:  Aaron J Specht; Chris N Parish; Emma K Wallens; Rick T Watson; Linda H Nie; Marc G Weisskopf
Journal:  Sci Total Environ       Date:  2017-10-04       Impact factor: 7.963

6.  Blood lead, bone lead and child attention-deficit-hyperactivity-disorder-like behavior.

Authors:  Yanfen Lin; Lihua Huang; Jian Xu; Aaron J Specht; Chonghuai Yan; Hongquan Geng; Xiaoming Shen; Linda H Nie; Howard Hu
Journal:  Sci Total Environ       Date:  2018-12-15       Impact factor: 7.963

7.  Portable XRF Technology to Quantify Pb in Bone In Vivo.

Authors:  Aaron James Specht; Marc Weisskopf; Linda Huiling Nie
Journal:  J Biomark       Date:  2014-11-27

8.  A microbeam grazing-incidence approach to L-shell x-ray fluorescence measurements of lead concentration in bone and soft tissue phantoms.

Authors:  Mihai Raul Gherase; Summer Al-Hamdani
Journal:  Physiol Meas       Date:  2018-03-29       Impact factor: 2.833

9.  Methodology using a portable X-ray fluorescence device for on-site and rapid evaluation of heavy-atom contamination in wounds: a model study for application to plutonium contamination.

Authors:  Hiroshi Yoshii; Kouta Yanagihara; Hitoshi Imaseki; Tsuyoshi Hamano; Hirokuni Yamanishi; Masayo Inagaki; Yasuhiro Sakai; Nobuyuki Sugiura; Osamu Kurihara; Kazuo Sakai
Journal:  PLoS One       Date:  2014-07-10       Impact factor: 3.240

  9 in total