OBJECTIVE: L-shell x-ray fluorescence (LXRF) is a non-invasive approach to lead (Pb) concentration measurements in the human bone. The first studies were published in the early 1980s. In the same period the K-shell x-ray fluorescence (KXRF) method using a Cd-109 radionuclide source was developed and later improved and refined. Lower sensitivity and calibration difficulties associated with the LXRF method led the KXRF to be the most adopted method for in vivo human bone Pb studies. In the present study a microbeam-based grazing-incidence approach to Pb LXRF measurements was investigated. APPROACH: The microbeam produced by an integrated x-ray tube and polycapillary x-ray lens (PXL) unit was used to excite cylindrical plaster-of-Paris (poP) bone phantoms doped with Pb in seven concentrations: 0, 8, 16, 29, 44, 59, and 74 µg g-1. Two 1 mm- and 3 mm-thick cylindrical shell soft tissue phantoms were made out of polyoxymethylene (POM) plastic. Three bone-soft tissue phantom sets corresponding to the 0, 1, and 3 mm POM thickness values resulted. Each phantom was placed between the microbeam and the detector; its position was controlled using a positioning stage. Small steps (0.1-0.5 mm) and short 30 s x-ray spectra acquisitions were used to find the optimal phantom position according to the maximum observed Sr Kα peak height. At the optimal geometry, five 180 s x-ray spectra were acquired for each phantom set. Calibration lines were obtained using the fitted peak heights of the two observed Pb Lα and Pb Lβ peaks. MAIN RESULTS: The lowest detection limit (DL) values were (2.9 ± 0.2), (4.9 ± 0.3), and (23 ± 3) µg g-1, respectively. The order of magnitude of the absorbed radiation dose in the POM plastic for the 180 s irradiation was estimated to be <1 mGy. SIGNIFICANCE: The results are superior to a relatively recently published LXRF phantom study and show promise for future designs of in vivo LXRF measurements.
OBJECTIVE: L-shell x-ray fluorescence (LXRF) is a non-invasive approach to lead (Pb) concentration measurements in the human bone. The first studies were published in the early 1980s. In the same period the K-shell x-ray fluorescence (KXRF) method using a Cd-109 radionuclide source was developed and later improved and refined. Lower sensitivity and calibration difficulties associated with the LXRF method led the KXRF to be the most adopted method for in vivo human bone Pb studies. In the present study a microbeam-based grazing-incidence approach to Pb LXRF measurements was investigated. APPROACH: The microbeam produced by an integrated x-ray tube and polycapillary x-ray lens (PXL) unit was used to excite cylindrical plaster-of-Paris (poP) bone phantoms doped with Pb in seven concentrations: 0, 8, 16, 29, 44, 59, and 74 µg g-1. Two 1 mm- and 3 mm-thick cylindrical shell soft tissue phantoms were made out of polyoxymethylene (POM) plastic. Three bone-soft tissue phantom sets corresponding to the 0, 1, and 3 mm POM thickness values resulted. Each phantom was placed between the microbeam and the detector; its position was controlled using a positioning stage. Small steps (0.1-0.5 mm) and short 30 s x-ray spectra acquisitions were used to find the optimal phantom position according to the maximum observed Sr Kα peak height. At the optimal geometry, five 180 s x-ray spectra were acquired for each phantom set. Calibration lines were obtained using the fitted peak heights of the two observed Pb Lα and Pb Lβ peaks. MAIN RESULTS: The lowest detection limit (DL) values were (2.9 ± 0.2), (4.9 ± 0.3), and (23 ± 3) µg g-1, respectively. The order of magnitude of the absorbed radiation dose in the POM plastic for the 180 s irradiation was estimated to be <1 mGy. SIGNIFICANCE: The results are superior to a relatively recently published LXRF phantom study and show promise for future designs of in vivo LXRF measurements.
Lead (Pb) is a well-known toxic element. While its toxicity has been known
for centuries, extensive research in the past several decades has revealed the
serious negative effects that various Pb exposure levels have on human health.
Increasing worldwide awareness by the public, health, and governmental institutions
and organizations has led to systematic Pb removal from many commonly-used chemicals
such as paints and gasoline. However, occupational Pb exposure and implementation of
Pb-free standards and regulations in developing countries remain important public
health concerns. Also, published studies in the last two decades demonstrated
correlations between developmental problems and even low Pb blood concentration
(<10 µg/dL) in children (Canfield
, Lanphear
, Jusko
) as well as links between Pb exposure and
various health problems in adults (Schaumberg
, Navas-Acien , 2008, 2009, Shih , Weisskopf ). While blood Pb
concentration measurement remains the metric of choice for quantitative clinical
assessments of individual Pb exposure, it has been known for a long time that most
of the Pb in the adult human body resides in the bone (Barry and Mossman 1970, Barry
1975). Moreover, the stable Pb isotope early study of Rabinowitz
estimated a biological half-life of blood Pb of about 30 d, which is much shorter
than the occupational or environmental Pb exposures spanning years or decades.
Therefore, in vivo bone Pb concentration measurements are more
representative of the cumulative effects of the prolonged or continuous humanPb
exposure. The history of in vivo bone Pb concentration measurements
using the x-ray fluorescence (XRF) emission process spans four decades and the data
provided by the in vivo bone Pb concentration measurements taken
during this period significantly contributed to the knowledge of Pb kinetics in the
human body during and after the Pb exposure as demonstrated in the review article by
Chettle (2005).Excitation of Pb atom innermost K-shell electrons by photoelectric absorption
requires high energy x-ray photons: the K-edge is 88.006 keV (Deslattes ). The in
vivo K-shell XRF (KXRF) method developed by Somervaille used the tibia bone
Pb excitation by the 88.04 keV gamma-ray photons from 109Cd radio-nuclide
(Dryak and Kovar 2006) and an internal
calibration method based on the ratio between the observed Pb KXRF peaks and the 88
keV elastic scattering peak. This method, typically employing acquisition times of
30 min and a 180° backscatter geometry, was most widely adopted for
in vivo bone Pb studies. The detection limit (DL) of Pb bone
concentration measured using the KXRF method improved considerably over a span of
two decades from 16–20 µg Pb/g-bone mineral in the
mid-1980s to around 2 µg Pb/g-bone mineral using
four-detector systems (Nie , Fleming and Mills 2007). A
recent general population survey data showed that tibia bone Pb concentration KXRF
measurements varied from below the DL to as high as 20 µg
Pb/g-bone mineral for the tibia bone (Behinaein
).The L-shell XRF (LXRF) tibia bone Pb measurement was first demonstrated in
the early 1980s using an 125I radionuclide (Wielopolski ) and later using the
partly polarized x-rays from an x-ray tube with a silver (Ag) target (Wielopolski ). The
x-ray tubes have the advantages of the turn off/on option for its x-ray beam and,
with the more recent development of portable x-ray spectrometers, there is also the
added advantages of size and portability as well as shorter acquisitions times
(Fleming ,
Nie ). The
original LXRF Pb measurement method developed by Wielopolski was closely scrutinized by
Todd (2002a). In summary, Todd’s
analysis indicated that the ultrasound soft tissue thickness measurements combined
with oversimplifying assumptions regarding the soft tissue x-ray attenuation
properties can lead to significant systematic uncertainties in the Pb concentration
measurements. Todd also experimentally explored LXRF system optimization (Todd 2002b) using various x-ray beam polarizers
and an x-ray tube with molybdenum (Mo) target. The optimal LXRF system was then used
to measure Pb concentration ex vivo in adult human cadaver tibiae.
Comparison with atomic absorption spectrometry Pb concentration measurements showed
a poor correlation between the two measurement methods, particularly for intact leg
Pb measurements (Todd ).The Pb LXRF signal can be enhanced by optimizing the trade-off between the
x-ray excitation of the 13.0 keV-bound Pb L-shell electrons and the spectral overlap
between the Compton scattered peak and the Pb Lα and Pb
Lβ peaks. The optimization was the object of a recent
study by Gherase using three incident photon energies (15.8, 16.6, and 17.5 keV) of
a synchrotron-generated x-ray beam and two excitation-detection geometries
corresponding to the 90° and 135° scattering angles. While the
calibration line slope data for the bare bone phantoms clearly favored the lower
15.8 keV energy, the addition of the soft tissue phantom complicated the analysis
due to the significant spectral overlap aforementioned.In the present study, the authors explored Pb detectability in
plaster-of-Paris (poP) bone phantoms using an x-ray tube integrated with a
polycapillary x-ray lens (PXL) in a grazing-incidence geometry. In conventional XRF
geometry, large angles (>10°) between the incident x-ray beam and
sample surface are employed. In a grazingincidence XRF geometry, a small angle
(<10°) between the incident x-ray beam and sample surface is
typically employed to minimize the scatter background in the acquired x-ray
spectra.Various XRF and x-ray diffraction (XRD) applications using PXLs were
summarized by MacDonald and Gibson (2003).
Studies involving PXLs in medical imaging research such as mammography (Abreu ), detection of
gold nanoparticles used as a contrast agent for x-ray imaging (Ricketts ), and optical luminescence
techniques targeting small-animal imaging (Zhang
), were also reported. Overall, biological
or medical applications involving PXLs were limited to various research labs, and,
to our best knowledge, in vivo XRF measurements of trace elements
in the human body using PXLs were not reported.The PXL produced a highly collimated and focused x-ray beam which is herein
referred to as the microbeam. There were two microbeam-related factors that
motivated the study. First, the microbeam reduces the inelastic (Compton) and
elastic (Rayleigh) scattering angle range of the photons reaching the detector by
minimizing the volume of first-interactions. Second, the bone scatter component can
be minimized by a grazing-angle approach in which the x-ray microbeam is incident
only on the outermost layer of the bone. This approach was supported by the
histological distribution of Pb within the human bone: a higher Pb concentration in
the outer layer of the cortical bone than within its inner volume and trabecular
bone (Zoeger ).The poP and polyoxymethylene (POM,
(CH2O)) were used as phantom materials
for the human cortical bone and the overlying soft tissue, respectively. Calibration
lines were obtained for the bare bone, 1 mm, and 3 mm POM thickness phantom sets
using the measured peak height of the observed Pb Lα and Pb
Lβ peaks. The measurements were performed at the
optimal grazing-angle geometry. This geometry was determined based on the maximum
observed peak height of strontium (Sr) in several short 30 s spectra acquired at
different positions of the bone phantoms relative to the incident x-ray beam. Sr is
a well-known contaminant of commercial poP products with estimates of its
concentration at a few mg per g of calcium (Ca) (Pejović-Milić ). Sr was not
of interest in this study and it was assumed to be uniformly distributed within the
poP bone phantoms. Therefore, constant peak height measurements of the Sr
Kα and Sr Kβ for different poP
bone phantoms were expected given the same XRF experimental conditions. Instead,
variations of the Sr Kα peak height measurements were
observed for the seven Pb-doped samples at the optimal grazing-angle geometry. These
variations were likely caused by inherent density differences between the bone
phantoms due to the porous microstructure of the poP material. Further, Sr
Kα peak height measurements were used to correct the Pb
Lα and Pb Lβ peak height data.
These corrections led to statistically significant improvements in the
goodness-of-fit of the linear fitting procedures which determined the Pb calibration
lines.The detection limit (DL) value of Pb was calculated as the ratio between
three times the null Pb concentration peak height uncertainty divided by the
calibration line slope value. The lowest Pb DLs for the three overlying POM
thickness values from 0 to 3 mm were 2.9 ± 0.2, 4.9 ± 0.3, and 23
± 3 µg g−1. While the 3 mm POM
thickness DL value remains too high for in vivo human bone Pb
measurements, the DL values for a soft tissue thickness below 2 mm are within the
range of such applications. Bone Pb surveys of children are particularly attractive
since their soft tissue thickness is, in general, lower than that of adults.We did not directly measure the corresponding radiation dose. However, our
approximate calculations based on the x-ray beam measurements gave a dose estimate
below 1 mGy for the 180 s irradiations used in this study. The background values
corresponding to the Pb LXRF peaks were also extracted from the spectra analysis.
The data showed no significant background increase from the 1 mm to the 3 mm POM
thickness experiments. In all, the results and discussion sections of this study
indicate that further improvements of the Pb LXRF approach are still possible.
Combined with viable novel calibration methods the effort may lead to future
in vivo humanPb bone measurements using microbeam-based
methods.
2. Methods
2.1. Sample preparation
Human bone phantoms were made of calcium sulphate hemihydrate ( ) which is also known as plaster-of-Paris (poP).
The poP powder form (Sigma-Aldrich, St. Louis, MO) was mixed with distilled
water and doped with known quantities of lead (Pb) using pipette-measured
volumes of Pb standard atomic absorption solution (Sigma-Aldrich, St. Louis,
MO). The precise chemical form of Pb was not known. The solvent was a diluted
water-based nitric acid (HNO3) solution (2% w/w). The viscous
mixture was then poured into aluminum cylindrical molds. The bone phantoms were
rigid cylinders with a 29 mm diameter following solidification. Their masses
were measured and the final Pb concentration was calculated based on the initial
Pb standard solution volume measurements. The resulting poP density for each
phantom was also calculated based on measured masses and volumes. The poP
density value and its uncertainty provided in table 1 were calculated as the average and standard deviation of the
seven bone phantoms, respectively. Seven bone phantoms were made with varying Pb
concentrations of 0, 8, 16, 29, 44, 59, and 74 µg
g−1.
Table 1
Table of phantom materials and their relevant properties.
Material
Chemical formula
Density (g cm−3)
Plaster-of-Paris (poP)
CaSO4·12H2O
1.05 ± 0.05
Polyoxymethylene (POM)
(CH2O)n
1.42 ± 0.14
Soft tissue phantoms were made of polyoxymethylene (POM,
(CH2O)). POM density and its
uncertainty were calculated based on volume water displacement and mass
measurements. Relevant properties of the poP and POM phantom materials are
summarized in table 1. X-ray linear
attenuation coefficients were calculated based on the material mass density and
elemental composition and using the x-ray mass attenuation coefficients
available through the XCOM (Berger ) online database of the National Institute of
Standards and Technology (NIST). Two cylindrical shell-shaped phantoms with a 30
mm inner diameter and 1 and 3 mm shell thickness, respectively, were machined
out of a larger diameter solid cylindrical rod. To mimic in
vivo LXRF bone Pb measurements, cylindrical bone phantoms were
inserted in the POM soft tissue phantoms. The combination resulted in three
phantoms sets.During the measurements the air gaps between the bone and soft tissue
phantoms due to slightly mismatched phantom diameters were eliminated by placing
folded paper on the opposite side of the x-ray beam incidence. A larger diameter
solid cylindrical base of the POM phantoms secured mechanical stability of the
bone-soft tissue phantom assembly.
2.2. Experimental setup
The schematic of the experimental setup used in the Pb LXRF measurements
is shown in figure 1. In the figure,
cylindrical sample refers to one of the three bone/soft tissue phantom
combinations. The sample excitation was achieved using the microbeam produced by
an integrated x-ray tube and polycapillary x-ray lens (PXL) (Polycapillary
X-beam Powerflux model, X-ray Optical Systems, Inc., East Greenbush, NY,
US).
Figure 1
Top view schematic of the experimental setup used in the Pb LXRF
measurements.
The x-ray tube had a tungsten (W) target and the x-ray lens was 10 cm in
length and 1 cm outer diameter. The x-ray tube voltage and current could be
varied in 0.1 kV and 1 µA increments, respectively.
Their maximum values of 50 kV and 1 mA were used during the LXRF measurements.
Knife-edge x-ray beam size measurements performed recently in our lab determined
that the x-ray lens had a focal length of 4 mm where the x-ray beam had a 24
µm lateral size measured as full width at
half-maximum (FWHM) at the 10 keV photon energy (Gherase and Vargas 2017). At the 10 keV photon energy, the x-ray
beam divergence at a distance larger than the 4 mm focal length was measured to
be 7.8 mrad (or 0.45°) as shown in figure
2. Hence, an approximate FWHM value of the x-ray beam in mm at a
distance d from the PXL in mm can be calculated from the beam
geometry of figure 2 using the following
equation:
Figure 2
Schematic of the PXL x-ray beam geometry.
It is important to note that the geometrical characteristics of the
x-ray beams formed by PXLs (lateral size, focal length, beam divergence) are all
dependent on the photon energy (Sun and Ding
2005). Upstream of the PXL the x-ray beam was filtered using an
eight-slot wheel in which custom-made filters could be placed. A 1.8 mm-thick
aluminum (Al) filter was used to attenuate the W L-shell XRF emissions in the
8–12 keV energy range. This step reduced, but did not eliminate the
spectral background in this energy range. X-ray photon counting and energy
measurements (i.e. x-ray spectra acquisition) were accomplished using a
silicon-drift x-ray detector with integrated pulse-height analyzer (X-123 SDD,
Amptek, Bedford, MA, US). The circular active area of the detector was 25
mm2 (or 5.6 mm diameter) and 0.5 mm thickness and the window was
a 12.7 µm-thickness beryllium (Be) sheet. The counting
rate capability of the detector provided by the manufacturer was 105
counts s−1.The samples were placed on a T-shaped Al plate securely mounted on an
automated XYZ modular motorized linear positioning stage assembly (Newport,
Irvine, CA, US). Sample positions in the x-ray beam with 1
µm accuracy were achieved with this system. The
entire XRF setup was placed on an x-ray shield consisting of a 56 × 62
cm2 and 6.35 mm-thickness Al plate. The plate was covered by a
stainless steel box with a slightly smaller base area and 46 cm in height which
could be opened to allow operating access to the XRF setup. The cover could be
manually opened when the x-ray beam was turned off. The integrated x-ray tube
and PXL unit was mounted in a fixed position inside the x-ray shield while the
x-ray detector and the positioning stage could be securely positioned anywhere
around the x-ray beam direction using a magnetic base holder and magnetic
strips, respectively. The integrated x-ray tube and PXL unit was also equipped
with an electromechanical beam shutter and a green and yellow light beacon. When
the green light turned on, it indicated that the power supply was on and that
the beam shutter was closed. When the yellow light turned on, it indicated that
the power supply ramped to the voltage and current values selected by the user.
When the yellow light was on and green light turned off, it indicated that the
shutter was open and the x-ray shield lid should be closed to avoid radiation
exposure of nearby users. Two L-shaped metallic pipes guided the connecting
cables with the x-ray tube power unit and the laptop computer which was used to
operate all devices. The entire x-ray shield and XRF assembly was mounted on an
optical table (Newport, Irving, CA, US) which also served as mechanical support
for the x-ray tube power supply and the laptop computer.
2.3. LXRF Pb measurements
Based on the bone/soft tissue phantom combinations three different sets
of LXRF measurements were performed: (1) bare bone phantoms, (2) bare bone and 1
mm thickness soft tissue phantom, (3) bare bone and 3 mm thickness soft tissue
phantom. For each combination, all seven Pb-doped bone phantoms were used. For
brevity, identifying labels were used throughout this article as summarized in
table 2. Five 180 s trials were
acquired for each Pb concentration. The dead time indicated by the x-ray
detector software interface was less than 1% for each measurement. For
each trial, the peak height values of the two observed Pb
Lα and Pb Lβ peaks were
measured. In all, six calibration lines (peak height versus Pb concentration)
resulted. The sample was positioned just outside the x-ray beam on the
positioning stage. The middle of the PXL circular end—as seen from the
direction of the x-ray beam—was used as an approximate visual guide to
the location of the x-ray beam.
Table 2
Notations of the three sets of Pb LXRF measurements.
Pb LXRF experiment
Notation
Bare bone (poP)
bb
Bare bone and 1 mm thickness cylindrical soft tissue (POM)
phantom assembly
bb-1 mm
Bare bone and 3 mm thickness cylindrical soft tissue (POM)
phantom assembly
bb-3 mm
Using the positioning stage, the sample was then moved towards the x-ray
beam in small steps in the 0.1– 0.5 mm range. Smaller 0.1 mm steps were
used for the bb phantom measurements while the larger 0.5 mm steps were used for
the bb-3 mm phantoms. At each location, a single 30 s x-ray spectrum was
acquired and the optimal position was then chosen based on the maximum observed
peak height of the strontium (Sr) Kα peak.
2.4. X-ray beam measurements
An approximate order of magnitude dose calculation was performed based
on the measured photon fluence rate for each energy. The experimental setup
shown in figure 2 was re-arranged such that
the x-ray detector window faced the PXL with a 5 cm distance between the two. At
this distance the x-ray beam lateral size calculated using equation (1) was ~0.8 mm
which was well within the dimensions of the detector. In order to mitigate dead
time counting losses of the detector at levels below 1%, the x-ray tube
current was set at 2 µA which was 50 times lower than
the 1 mA current used in the LXRF measurements.
2.5. Data analysis
The raw x-ray spectra were analyzed using OriginPro 2015 (OriginLab
Northampton, MA, US) statistical data analysis and plotting software. The Pb
Lα and Pb Lβ peaks were
fitted separately by selecting only the data around the peaks. The peaks were
modeled as a Gaussian function with a linear background according to the
following equation:In equation (2),
variables y and x represent the number of
counts and photon energy, respectively. The variables a and
b are the parameters of the linearly-modelled background
and H, x0, and σ are the
peak height, center, and standard deviation parameters of the Gaussian peak
model, respectively. The observed Sr Kα and Sr
Kβ peaks were fitted simultaneously using a model
as in equation (2) to which a
second Gaussian function was added. Nonlinear curve fitting component of the
OriginPro software based on the Levenberg–Marquardt numerical
minimization algorithm was employed to fit the data to equation (2) model. For each
channel with N counts, the standard deviation of the
measurement was taken to be equal to according to the Poisson statistics. Thus, the
corresponding statistical weight was equal to 1/N.
Goodness-of-fit was simultaneously monitored using the following built-in
options of the OriginPro software: (i) reduced chi-square value
(χ2/n), (ii) the
coefficient of determination (R2), and (iii) visualization of the
fitted function and data plots. In general, all parameters in equation (2) were treated as free
parameters in the fitting procedures. However, for low Pb concentrations near
the detection limit, the Pb Lα and Pb
Lβ peaks corresponding to the bone and soft tissue
phantom measurements were hardly distinguishable from the background, and,
therefore, the peak fitting routine was aided by assigning the center
(x0) and standard deviation (σ)
parameters with the measured values of the well-resolved Pb peaks from the bare
bone phantom LXRF measurements. The same approach was employed for the null Pb
concentration spectra analysis.The final results were calculated as the mean and the standard deviation
of the mean (SDOM) of the five fitted peak height values corresponding to the
180 s trials. These calculations and organization of the final results were
performed using Microsoft Office Excel software (Microsoft, Redmond, WA, US).
Sample and final results plots were also done using the OriginPro software.
X-ray atomic transitions of the observed XRF peaks were identified using the
tabulated data from Deslattes . The Sr Kα peak
height in identical experimental conditions should be constant since Sr was a
homogeneous contaminant of the plaster-of-Paris compound. The observed
variations of the Sr Kα peak height data indicated that
there were systematic uncertainties in the measurement procedure. These were
likely caused by density differences amongst the bone phantoms due to the porous
nature of the poP material. This observation led to corrections of the Pb
Lα and Pb Lβ peak height
data. The Pb peak height values denoted generic by
HPb were corrected using the corresponding Sr
Kα peak height denoted by
HSr and its maximum value
HSr,max out of the set of seven Pb concentration
measurements as follows:From the calibration lines of the form y =
y0 + sx obtained with the Pb
Lα and Pb Lβ peak height
data, the DL of the Pb concentration measurement was calculated using the
following equation (Fleming ):In equation (4) the
σ0 represents the SDOM obtained from the
five trials of the null Pb concentration bone phantom and s
represents the slope of the calibration line. Using error propagation of
statistically-independent variables, an uncertainty on DL,
σDL, can be calculated using the slope,
s, and the error on the slope,
σs, as follows:
3. Results
3.1. Pb LXRF measurements
Sample plots of the x-ray spectra obtained in the Pb LXRF experiments
are shown in figure 3.
Figure 3
Two sample x-ray spectra corresponding to the 74
µg g−1 Pb concentration of the
bb (black line) and bb-3 mm (red line) LXRF measurements. The four plots focus
on different energy ranges: (a) 0–50 keV; (b) 0–10 keV energy
range; (c) 10–13.5 keV (Pb LXRF peaks and Gaussian peak fitting); (d)
13–20 keV (Sr Kα and Sr
Kβ peaks and Gaussian fitting of the peaks).
In plot (a) the larger scattered bremsstrahlung wide peak in the bb-3 mm
spectrum can be noticed due to the increased number of scatter events in the
soft tissue phantom. A larger number of XRF peaks can be observed in plot (b) in
the bb spectrum in the absence of the x-ray attenuation within the POM soft
tissue phantom.Figure 4 shows variations of the Sr
Kα (14.1 keV) peak height in two distinct
situations: (i) position of the phantom relative to the x-ray beam was varied in
0.1 mm and 0.5 mm steps (plots (a) and (b)); and (ii) x-ray spectra were
acquired to produce the Pb Lα and Pb
Lβ calibration lines at the optimal positions of
the three phantom sets (plot (c)). In figure
4(b), sample plots of the Sr Kα data which
determined the optimal phantom position are shown. It can be seen that optimal
phantom position was achieved roughly within the 1, 1.5, and 3 mm distances
corresponding to the bb, bb-1 mm, and bb-3 mm phantom measurements,
respectively. The increasing trend was related to the experimental approach in
which the thickness of the POM soft tissue phantom was not accounted for in the
initial phantom positioning procedure. The Sr Kα peak
height data shown in the plots from figure
4(c) were used to perform Pb Lα and Pb
Lβ peak height corrections using equation (3) as described in
section 2.5.
Figure 4
(a) Plot of the Sr Kα peak height variations
when the position of the phantoms is varied in small 0.1 mm steps in the
y-axis direction shown in figure 1. The 74 µg g−1
Pb concentration of the bb and bb-3 mm phantom assembly were used. (b) Sample
plots of the Sr Kα peak height variations obtained
using 30 s x-ray spectra acquisitions. The optimal phantom positions indicated
by arrows correspond to the maximum values of these curves. (c) Sr
Kα peak height values calculated as the mean of the
five 180 s trials for each phantom assembly and Pb concentration. The error bars
were calculated as standard deviation of the mean and were too small to be
distinguished from the data points.
The Pb LXRF measurements led to the raw and corrected calibration lines
shown in figure 5. The slope and
y-axis intercept parameter values and their standard
deviations obtained from the linear fitting routines are shown in table 3. The corresponding reduced
chi-square values for each calibration line are also provided in table 3. A visual inspection of the plots
in figure 5 and the reduced chi-square
values from table 3 show that the equation (3) corrections improved
the calibration line, particularly for the bb measurements. The
p value column in table
3 indicates if the observed chi-square value departure from the
expected unity value is statistically significant at the 5% level. The
p values for the corresponding d = 7
− 2 = 5 degrees of freedom chi-square probability distribution function
were compared to the 5% level using the statistical tables from the
Taylor textbook (Taylor 1997).
Figure 5
Raw (panel plots (a) and (b)) and corrected (panel plots (c) and (d))
calibration lines of Pb measurements based on the Pb Lα
(panel plots (a) and (c)) and Pb Lβ (panel plots (b)
and (d)) peak height measurements.
Table 3
Results of the linear fitting of the raw and corrected Pb
Lα and Pb Lβ peak height
data. The numbers in the round parentheses represent the uncertainties in the
last significant figure of the corresponding value.
Experiment
Pb
Lα
Pb
Lβ
s(µg−1
g)
y0(counts)
χ2/n
p
s(µg−1
g)
y0(counts)
χ2/n
p
bb
0.28(1)
1.2(3)
2.0
>0.05
0.24(2)
0.6(5)
4.1
<0.05
bb-1 mm
0.084(5)
0.3(2)
1.4
>0.05
0.10(2)
−0.2(5)
2.3
<0.05
bb-3 mm
0.026(5)
0.6(2)
1.8
>0.05
0.030(4)
0.07(1)
0.9
>0.05
Corrected Pb Lα
Corrected Pb Lβ
bb
0.322(8)
0.8(2)
0.5
>0.05
0.28(2)
0.7(4)
2.0
>0.05
bb-1 mm
0.087(5)
0.5(2)
1.2
>0.05
0.11(2)
0.3(5)
1.1
>0.05
bb-3 mm
0.024(5)
0.7(2)
1.5
>0.05
0.030(4)
0.1(1)
0.7
>0.05
The uncertainty in the null Pb concentration phantom trials and the
values of the MDL and its uncertainty calculated using equations (4) and (5) are provided in table 4.
Table 4
Null Pb concentration uncertainties
σ0 and calculated DL results. The
numbers in the round parentheses represent the uncertainties in the last
significant figure of the corresponding value.
Corrected Pb
Lα
Corrected Pb
Lβ
Experiment
σ0(counts)
DL(µg
g−1)
σ0(counts)
DL(µg
g−1)
bb
0.3
3.1(1)
0.3
2.9(2)
bb-1 mm
0.1
4.9(3)
0.6
18(3)
bb-3 mm
0.3
39(8)
0.2
23(3)
3.2. X-ray beam measurements and absorbed dose calculations
The results of the x-ray beam measurements and subsequent corrections
are summarized in the plots of figure 6.
The count rate calculated from the raw spectrum data (black line in the figure 6(b) plot) was 1.3 ×
103 counts s−1, well within the 105
counts s−1 processing rate of the x-ray detector. The
measured x-ray spectra were corrected for the detection efficiency of the 0.5 mm
Si layer and were also adjusted from the 2 µA tube
current used during x-ray beam measurements to 1 mA x-ray tube current used
during Pb LXRF measurements. An x-ray beam rate of ~ 2 ×
106 photons s−1 was calculated by summing all
corrected counts shown in figure 6(c) and
dividing by the 300 s acquisition time.
Figure 6
(a) Plot of the 0.5 mm thick Si x-ray detector efficiency expressed as
the photoelectric absorption factor 1 −
exp(−ρτt) where
ρ and t are the Si mass density
and thickness and the τ is the photoelectric absorption
part of the mass attenuation coefficient of Si taken from the XCOM database. (b)
Plots of the measured and detector-efficiency-corrected x-ray spectra. (c) Plot
of the x-ray spectrum adjusted for the 1 mA x-ray tube current used in all Pb
LXRF measurements. (d) Semilogarithmic plot of the corrected x-ray spectrum in
the 0–14 keV energy range.
The XRF peaks shown in figure 6(d)
plot indicate the presence of Ni, Cu, and W elements. The first two were
identified as contaminants of the Al collimator, while W lines originated from
the x-ray tube W target material.A rough estimate of the absorbed dose was calculated employing a number
of simplifying assumptions. First, the absorbed dose in the poP bone phantom
material was neglected. The energy Ebeam carried by
the x-ray beam in the 3 min (or 180 s) time of the Pb LXRF measurement trial was
the sum of N counts of energy
E over all 2048 channels:For each photon energy E the fraction of
absorbed energy f along path length
L = 1.11 cm for bb-1 mm and L = 1.99 cm
for the bb-3 mm in the POM soft tissue material was calculated using the
following approximation:In equation (7)
ρ = 1.42 g cm−3 was the density of
the POM material also provided in table 1
and µ (E) is the mass
attenuation coefficient of the POM material corresponding to the photoelectric
and Compton interactions and was calculated using the XCOM tables.This is an approximation, since not all Compton-interacting photons
deposit their entire energy within the volume of interest. Hence, the
approximate average dose D was calculated as the absorbed
energy Eabs divided by the POM mass
m:In equation (8) it was
assumed the energy absorption occurred within the small x-ray beam volume. The
x-ray beam radius r = FWHM/2 = 0.137mm was calculated using
equation (1) at distance
d = 20mm. For bb-1 mm and bb-3 mm 180 s trials the absorbed
dose was estimated at 0.7 and 0.5 mGy, respectively.
4. Discussion
4.1. Pb LXRF measurements
The data plots of figures 4(a) and
(b) can be explained by the x-ray attenuation of the incident beam
and that of the emergent Sr Kα photons as the phantom
is incrementally positioned towards the x-ray beam. The data in plot 4(c) can be
explained by several possible factors. (1) poP bone phantoms had microscopic air
pockets and surface irregularities on the submillimeter scale similar to that of
the x-ray beam which led to a different XRF experiment for each sample despite
the fact that Sr could be considered uniformly distributed throughout the bone
phantom. (2) Optimal geometry was perhaps not achieved equally well for all
samples given that the initial sample positioning was only determined visually
and not determined in a reproducible manner. (3) The Sr
Kα peak height variation appears to decrease with
increasing thickness of the POM soft tissue phantom as shown plot (c). This is
also related to the much broader maximum noticed in the bb-3 mm experiment
compared to that of the bb measurements shown in the plots from figure 4(a). The systematic uncertainties
noted for the Sr Kα peak data were very likely
responsible for the larger-than-unity
χ2/n values of the bb
calibration line shown in table 3. Ad hoc
corrections given by equation (3)
reduced these values and improved the linearity as can be seen in the panel
plots (c) and (d) of figure 5. Sr and Pb
have different XRF photon energies and photoelectric absorption cross sections.
There were also differences in the x-ray attenuation of the incident and
emergent XRF photons. This means an optimal XRF detection of Sr was not entirely
optimal for Pb detection. A useful test of reproducibility in optimal
positioning would have been to perform the LXRF measurements with the same Pb
concentration phantom several times starting from the positioning stage of the
experiment, and not just mere automated repetitions of the x-ray spectra
acquisition. Due to the relatively long experimental effort required, as well as
the improvement demonstrated by the Sr Kα correction
for the bb calibration line measurements, this test was not performed. The
results of this study and the present discussion indicate the utility of such
effort in future investigations using a similar microbeam technique.An important part of improving Pb LXRF measurement is the slope of the
calibration line which is an important component of the sensitivity of any
applied XRF method. The slope values, s, for each phantom set
of Pb LXRF measurements are provided in table
3. The observed decrease in its value with the increasing of the
overlying POM thickness denoted by t was expected and was due
to the exponential attenuation of the incident and emergent x-ray photons. Let
us assume that incident x-ray photons of energy E
cross a thickness t of the POM soft tissue phantom
to reach the bone and the linear attenuation coefficient of POM at this x-ray
photon energy is µ. Also, the emergent Pb
Lα and Pb Lβ photons have
linear attenuation coefficients µ
and µ and cross a POM soft tissue
phantom layer of thickness t. Then, the relationship between
the Pb Lα or Pb Lβ slope of
the bb phantom, s (0), and that
of a bb and soft tissue layer of thickness t,
s (t) isEquation (9) assumes
well-defined paths of monoenergetic photons. While the direction of the incident
microbeam and the Pb LXRF photon energies support this assumption, the incident
photon energies of the microbeam and the paths of emergent photons are shaped by
the physics of the x-ray tube and the transmission properties of the PXL, and
the solid angle of the detector, respectively. Equation (9) also assumes the same x-ray attenuation in the
poP bone phantom material for all three phantom sets, which is equivalent to
having identical grazing-incidence conditions. Figure 7 shows the relationship between the soft tissue thickness
values t and t in the Pb LXRF
measurements. The selected 0.181 mm−1 x-ray linear
attenuation coefficient of the 15.2 keV photons corresponds to the Pb
L3 edge energy.
Figure 7
Geometry of the Pb LXRF measurements.
Figure 8 shows the calibration line
slope data from table 3 and the
single-parameter attenuation model fitted curves. The observed relationship for
bb calibration lines s (0) >
s (0) can be explained by the
measured L = 0.894
for Pb (Garg ). The L
ratio was larger than unity for the bb-3 mm calibration lines due to the lower
x-ray attenuation of the higher energy Pb Lβ photons.
The x-ray linear attenuation coefficients of the 15.2 keV incident monoenergetic
photons and the Pb Lα and Pb
Lβ photons for both poP and POM materials are
provided in table 5. The fitting results
are shown in table 6. The two
a parameter values from the fourth column of table 6 represent the effective x-ray
linear attenuation coefficient of the POM plastic material for the Pb
Lα and Pb Lβ energies.
Figure 8
Calibration line slope data and approximate x-ray attenuation model
fitted curves for the Pb Lα (black) and Pb
Lβ (red).
Table 5
X-ray linear attenuation coefficients of the phantom materials for
relevant x-ray photon energies.
Photon energy (keV)
10.5
12.6
15.2
Atomic transition
Pb Lα
Pb Lβ
Pb L3 edge
µpoP
(mm−1)
3.15
1.88
1.10
µPOM
(mm−1)
0.511
0.303
0.181
Table 6
Results of the calibration line slope data fitting with the x-ray
attenuation model.
Slope data
Fitting function: s =
s (t)
sα,β (0)(counts
µg−1 g)
Parameter
a(mm−1)
χ2/n
Pb Lα
sα(t)=sα(0)exp[−at−0.181·(30t+t2)12]
0.322
0.29 ± 0.01
0.11
Pb Lβ
sβ(t)=sβ(0)exp[−at−0.181·(30t+t2)12]
0.28
0.12 ± 0.03
0.71
These values are lower than the 0.511 mm−1 and 0.303
mm−1, respectively as provided in the last row of table 5. The discrepancy has two plausible
explanations: (i) the effective path length of emergent Pb LXRF photons is
larger than the soft tissue phantom thickness t, and (ii) the
x-ray attenuation in the poP bone phantom was not the same amongst all
measurements. Although the simplified attenuation model did not include all
details of the experimental conditions, it is worth mentioning that the data
from figure 8 did not fit well an
exponential decay function in which the exponent was a linear function in
t.The agreement between the slope data and the simplified x-ray
attenuation model brings about the discussion of the scatter background. The
peak fitting routine modeled the spectral background underneath the observed
peaks over a narrow energy range. Hence, the relevant background data for the Pb
LXRF measurements were readily available in the results of the peak fitting
routines.Figure 9 is a scatter plot of the
background underneath the Pb Lα and Pb
Lβ peaks calculated at the 10.5 keV and 12.6 keV
energies. For each phantom, there were 35 data points corresponding to the five
trials for each of the seven Pb-doped bone poP bone phantoms. The
Pearson’s correlation coefficient is also provided in the plot’s
legend. Substantial overlap between the bb-1 mm and bb-3 mm background data of
both peaks can be observed. The difference between the average values of the two
data sets was not statistically significant. In other words, the background in
the 10–13 keV energy range was essentially the same for the bb-1 mm and
bb-3 mm phantoms.
Figure 9
Scatter plot of the fitted background underneath the Pb
Lα and Pb Lβ peaks as
calculated at the 10.5 keV and 12.6 keV energy, respectively.
This is a positive outcome of this study and evidence of the scatter
reduction due to the employed microbeam grazing-incidence approach. There is,
however, clear separation between the bb data (black points) and the bb-1 mm and
bb-3 mm data (red and blue data points) for the background underneath the Pb
Lβ peak located at the 12.6 keV energy. This
increase is due to a larger number of Compton-scattered photons within the POM
layer that reached the detector. The incident x-ray beam spectrum from figure 6(d) shows a significant increase in
the number of photons for energies above 11 keV. Filtration by the added 1.8 mm
Al filter was clearly not sufficient. Additional Al filtration, perhaps at the
output-end of the PXL, will have to be investigated in future studies using this
microbeam system. An alternative option in background reduction is the use of a
confocal XRF method. In this approach, a secondary PXL (Malzer and Kanngießer 2005) or collimating channel
array (CCA) (Choudhury ) would guide the x-ray photons emergent from a small targeted
volume to the x-ray detector. The limited photon pathways allowed by the
secondary PXL would effectively enhance the signal-to-background ratio in the
x-ray spectrum even when PXL transmission is taken into account.
4.2. Literature comparisons
The DL values of Pb are about two times better than the reported values
of Fleming
who used a portable XRF spectrometer, 1.2 and 2.7 mm thickness of resin as soft
tissue phantom material, and the same acquisition times and DL definition. Table 7 summarizes the calibration data
from the paper of Fleming in which the count rate results were converted to number of
counts by multiplying them with their corresponding 180 s acquisition time.
Table 7
Calibration data summary from Fleming
.
Pb
Lα
Pb
Lβ
Phantom
σ0(counts)
Slope(counts ·
µg−1g)
σ0(counts)
Slope(counts ·
µg−1g)
bb-circular
4.3
1.08 ± 0.04
3.6
1.13 ± 0.04
bb-square
4.1
0.97 ± 0.05
4.0
1.0 ± 0.05
bb-square and 1.2 mm resin
4.7
0.58 ± 0.07
5.0
0.61 ± 0.07
bb-circular and 2.7 mm resin
7.9
0.25 ± 0.04
6.3
0.40 ± 0.04
Comparing table 7 values with the
current study ones provided in tables 3
and 4, one can notice that the slope
values from Fleming et al are 5–10 times larger than
the current study, while the corresponding uncertainties on the null Pb
concentration measurements are more than one order of magnitude larger than the
current study. This is perhaps an indicator that the gain in the lower DL values
from table 4 could be related to lower
scatter background levels in the energy region of the observed Pb LXRF peaks.
The portable spectrometer had only a 20 µA current
x-ray tube which was 2% of the 1 mA current used in the current study. A
direct side-by-side comparison becomes even more complicated when the 10 cm long
PXL microbeam collimation and transmission properties and the unknown detector
capabilities of the portable spectrometer from Fleming study are taken into
account.A Pb DL value of 8.4 µg g−1
was reported by Nie for 120 s LXRF measurements on poP Pb-doped phantoms with an
overlying 2 mm Lucite soft tissue phantom. In their study, the bone and soft
tissue phantoms as well as cadaver tibia bones were used to test a novel
calibration method and detection limit definition which did not include the
traditional peak fitting routine. Therefore, a direct measurement sensitivity
comparison is, again, difficult. For a 2 mm POM layer a Pb
Lα calibration line slope
s (2mm) = 0.042 counts
µg−1 g can be calculated using
the corresponding function and data from table
6. Combined with a σ0 = 0.2
counts (similar to the values from table
4), a 9.5 µg g−1 DL
estimate was calculated using equation
(4) definition.
4.3. In vivo Pb concentration LXRF measurement
considerations
Several considerations regarding the feasibility of in
vivo Pb LXRF measurements can be made. The Pb LXRF measurement
method presented here used the relatively large Sr Kα
peak to find the optimal grazing-incidence geometry and in systematic
uncertainty corrections. A direct translation of the measurement method
presented in this study to in vivo bone Pb applications poses
certain difficulties. It is known that Sr concentration in unexposed human bone
ranges in the 0.1–0.3 mg Sr/g Ca (Pejović-Milić 2004). The Sr level in the poP bone
phantoms used in the current investigation was unknown. Pejović-Milić
determined that a commercial poP sample contained about 4 mg Sr/g Ca,
approximately an order of magnitude higher than the expected human bone Sr
concentration. Therefore, direct implementation would require proportionally
longer times which would increase the overall radiation dose cost.The scattered bremsstrahlung peak does not offer, at least in a brief
superficial analysis, additional useful information. Figure 10 shows the normalized number of counts in the
10–50 keV scattered bremsstrahlung energy region and the normalized
fitted Sr Kα peak height as a function of the relative
position of the bb-3 mm phantom. While the Sr Kα peak
height data plot indicates the optimal grazing-incidence position of the
microbeam, the sum of the bremsstrahlung counts values decrease monotonically
with the relative position and the curve does not indicate the position for
which the microbeam is incident on the bone (at grazing-incidence or not).
Alternatively, the grazing-incidence geometry could be determined by a separate
soft tissue measurement procedure such as ultrasound (Pejović-Milić )
as done in the past LXRF studies. This approach, however, does not eliminate the
issue of intrinsic variations of the x-ray linear attenuation coefficients of
the human soft tissues.
Figure 10
Plots of the normalized (i) sum of the scattered bremsstrahlung photons
in the 10–50 keV energy (black line and data points), and (ii) fitted
peak height of the Sr Kα peak (red line and data
points) versus the relative position of the bb-3 mm phantom. The corresponding
30 s x-ray spectra also generated data shown in figure 4(a). Normalization was done by dividing the values in each
data set by its corresponding maximum value.
The calculated absorbed radiation dose to the POM material corresponding
to the 180 s trials was less than 1 mGy. There were unavoidable counting losses
in the detection process (dead time was ~1%) and the dose
absorbed in the poP bone phantom was neglected. An important approximation
overestimating the calculated dose was the radiation absorption volume being
taken as equal to the volume along the x-ray beam path in the POM material which
was estimated using the lateral beam size measurements in air. The 1 mGy order
of magnitude of the radiation dose is the same as reported in the early KXRF
study of Somervaille , but lower than the 10 mGy dose to the skin reported in the
pioneering LXRF study of Wielopolski . As discussed earlier, an added radiation cost
was the finding of the optimal grazing-incidence geometry. However, due to the
smaller radiation-exposed volumes, six spectral acquisitions of 30 s duration
delivered a lower dose to the phantoms than the five 180 s trials used in the Pb
LXRF measurements.An important part of the LXRF Pb concentration measurement applicability
to in vivo human studies is the calibration method. The early
calibration method proposed by Wielopolski
, 1989) was based on ultrasound measurements of soft tissue thickness
overlying the bone.After more than three decades of research, a novel calibration method is
needed to improve the precision and accuracy of Pb LXRF measurements to the
levels required by in vivo studies. A first step was taken by
Nie who
developed a novel calibration method based on the increase of the observed
Compton peak with the overlying soft tissue thickness. Nie et
al also replaced the traditional peak fitting and calibration line
method used in Pb KXRF measurements with a self-titled ‘background
subtraction’ method. In it, the authors implicitly assumed that the net
increase in the number of counts under the Pb peaks, after the null Pb
concentration background is subtracted, is due to the presence of the Pb in the
analyzed sample even when the Pb peaks themselves were not resolved. The
assumption is not necessarily wrong. A criticism which can be brought forward is
that the authors did not provide the reference for a similar precedent in the
past or current XRF literature. This data analysis procedure was not tested in
this work, but, perhaps, it deserves a closer look. The expected increase in the
number of scattered bremsstrahlung photons with increasing thickness of the soft
tissue phantom layer was also noticed in this study as shown in figure 11. A similar trend, albeit with more
data points and reversed axes, was also found by Nie . The use of such data to correct
the in vivo Pb LXRF data for the soft tissue x-ray attenuation
is appealing, but requires a careful investigation of all uncertainties involved
in such approach.
Figure 11
The number of counts in the 10–50 keV energy region of the
acquired x-ray spectra. The values and corresponding error bars were calculated
as the mean and standard deviation of the mean of the 35 data points for each
phantom set.
A different soft tissue x-ray attenuation correction method can be
implemented using the grazing-incidence approach proposed in this study. Instead
of using the Sr Kα peak to achieve the optimal
grazing-incidence geometry, the microbeam could scan the soft tissue from the
top to the location of the bone with the x-ray detector opposite to the PXL. The
setup would resemble a traditional x-ray projection scanning imaging method. The
scanning distance would provide the soft tissue thickness in the perpendicular
direction to the microbeam. The average linear x-ray attenuation coefficient of
the soft tissue could be calculated from these measurements provided that the
soft tissue thickness in the microbeam direction could be measured by mechanical
means such as the use of thin plastic parallel plates. This could mitigate the
negative effects of the assumptions regarding x-ray attenuation properties of
the soft tissue employed in earlier LXRF studies and discussed by Todd (2002a). At this stage, this proposal
is speculative and its radiation dose and additional details can be determined
in future LXRF experiments employing this method.
5. Conclusions
A table-top microbeam unit was used to implement a grazing-incidence
approach to Pb LXRF measurements in poP bone and POM soft tissue phantoms. The
calculated Pb DL values based on the calibration line LXRF measurements were in the
range 3–40 µg g−1 for the
0–3 mm POM thickness range. The radiation dose in the POM material for the
180 s irradiation was estimated to be lower than 1 mGy. The data analysis also
demonstrated that the backscatter background corresponding to the observed Pb
Lα and Pb Lβ peaks did not
increase with increasing POM soft tissue phantom thickness and was estimated to be a
direct effect of the proposed method. The discussion of the methodology and results
indicated that a direct implementation of this approach to in vivo
measurements would be difficult, but viable solutions and ideas using a microbeam
were proposed to motivate and encourage future bone Pb LXRF investigations.
Authors: Andrew C Todd; Spencer Carroll; Ciaran Geraghty; Fuad A Khan; Erin L Moshier; Shida Tang; Patrick J Parsons Journal: Phys Med Biol Date: 2002-04-21 Impact factor: 3.609
Authors: S Behinaein; D R Chettle; M Fisher; W I Manton; L Marro; D E B Fleming; N Healey; M Inskip; T E Arbuckle; F E McNeill Journal: Physiol Meas Date: 2017-01-09 Impact factor: 2.833
Authors: Marc G Weisskopf; Nitin Jain; Huiling Nie; David Sparrow; Pantel Vokonas; Joel Schwartz; Howard Hu Journal: Circulation Date: 2009-09-08 Impact factor: 29.690
Authors: Bruce P Lanphear; Richard Hornung; Jane Khoury; Kimberly Yolton; Peter Baghurst; David C Bellinger; Richard L Canfield; Kim N Dietrich; Robert Bornschein; Tom Greene; Stephen J Rothenberg; Herbert L Needleman; Lourdes Schnaas; Gail Wasserman; Joseph Graziano; Russell Roberts Journal: Environ Health Perspect Date: 2005-07 Impact factor: 9.031