Assessing wear of the acetabular cup using computed tomography: an ex vivo study.

L Olivecrona1, A Jedenmalm, P Aspelin, L Weidenhielm, M E Noz, G Q Maguire, M P Zeleznik, H Olivecrona.   

Abstract

PURPOSE: To validate a clinically useful method for measuring acetabular cup wear using computed tomography (CT).
MATERIAL AND METHODS: Eight uncemented acetabular cups were scanned twice ex vivo using CT. The linear penetration depth of the femoral component head into the cup and the thickness of the remaining polyethylene liner were measured in the CT volumes using dedicated software. Two independent examiners twice assessed each volume. The CT measurements were compared to direct measurements using a coordinate measuring device and micrometer measurements.
RESULTS: Accuracy of wear measurements expressed as penetration depth was +/-0.6 and +/- 1.0 mm for the two examiners, respectively, with no significant differences between examiners, trials, and CT scans. Accuracy of measurements of remaining polyethylene was +/- 1.3 and +/- 1.0 mm, respectively, for the two examiners. Systematic differences between examiners were found, but no significant differences between trials and CT scans. These differences were due to different interpretations of metal artifacts in the volumes.
CONCLUSION: The proposed CT method for evaluating wear as head penetration depth allows for reliable wear detection at a clinically relevant level. Measurements of remaining polyethylene on CT volumes are not as reliable as wear measurements owing to metal artifacts.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 16392610     DOI: 10.1080/02841850500271074

Source DB:  PubMed          Journal:  Acta Radiol        ISSN: 0284-1851            Impact factor:   1.990


  4 in total

1.  Validation of a 3D CT method for measurement of linear wear of acetabular cups.

Authors:  Anneli Jedenmalm; Fritjof Nilsson; Marilyn E Noz; Douglas D Green; Ulf W Gedde; Ian C Clarke; Andreas Stark; Gerald Q Maguire; Michael P Zeleznik; Henrik Olivecrona
Journal:  Acta Orthop       Date:  2011-02-01       Impact factor: 3.717

2.  Measurement of femoral head penetration in polyethylene using a 3-dimensional CT-scan technique.

Authors:  Eric Vandenbussche; Mohammed Saffarini; Ulrich Hansen; Fabienne Taillieu; Céline Mutschler; Bernard Augereau; Thomas M Gregory
Journal:  Acta Orthop       Date:  2010-10       Impact factor: 3.717

3.  Risk factors for periacetabular osteolysis and wear in asymptomatic patients with uncemented total hip arthroplasties.

Authors:  Buster Sandgren; Joakim Crafoord; Henrik Olivecrona; Göran Garellick; Lars Weidenhielm
Journal:  ScientificWorldJournal       Date:  2014-11-16

4.  A new automated way to measure polyethylene wear in THA using a high resolution CT scanner: method and analysis.

Authors:  Gerald Q Maguire; Marilyn E Noz; Henrik Olivecrona; Michael P Zeleznik; Lars Weidenhielm
Journal:  ScientificWorldJournal       Date:  2014-01-22
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.