Survival estimates and risk factors for failure with 6 x 5.7-mm implants.

Michael A Gentile1, Sung-Kiang Chuang, Thomas B Dodson.   

Abstract

PURPOSE: Short dental implants facilitate prosthetic restoration in the setting of limited alveolar bone height. The study objectives were to (1) estimate the 1-year survival of Bicon 6 x 5.7-mm implants, (2) compare the 1-year survival of 6 x 5.7-mm implants with that of non-6 x 5.7-mm implants, and (3) identify risk factors associated with implant failure.
MATERIALS AND METHODS: A retrospective cohort study design was used. The sample was composed of patients who had received at least one 6 x 5.7-mm implant. Predictor variables were categorized as demographic, health status, anatomic, implant-specific, prosthetic, perioperative, and reconstructive. The outcome variable was implant failure, defined as explantation. Appropriate descriptive, bivariate, and multivariate survival statistics were computed.
RESULTS: The sample was composed of 35 patients in whom 172 implants had been placed (45 of which were 6 x 5.7-mm). The 1-year survival rates for 6 x 5.7-mm and non-6 x 5.7-mm implants were 92.2% and 95.2%, respectively (P = .76). After adjusting for covariates in a multivariate model, implant size was not associated with failure (P = .95). DISCUSSION: The comparable survival estimates for 6 x 5.7-mm implants and non-6 x 5.7-mm implants in this study suggested that 6 x 5.7-mm implants can become osseointegrated and bear a functional load after placement.
CONCLUSIONS: The survival of 6 x 5.7-mm implants was comparable to that of non-6 x 5.7-mm implants.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 16392351

Source DB:  PubMed          Journal:  Int J Oral Maxillofac Implants        ISSN: 0882-2786            Impact factor:   2.804


  10 in total

Review 1.  Short Implants: New Horizon in Implant Dentistry.

Authors:  Neha Jain; Manisha Gulati; Meenu Garg; Chetan Pathak
Journal:  J Clin Diagn Res       Date:  2016-09-01

2.  Short implants: A systematic review.

Authors:  I Karthikeyan; Shrikar R Desai; Rika Singh
Journal:  J Indian Soc Periodontol       Date:  2012-07

3.  Survival rate of short, locking taper implants with a plateau design: a 5-year retrospective study.

Authors:  Kemal Özgür Demiralp; Nihat Akbulut; Sebnem Kursun; Didem Argun; Nilsun Bagis; Kaan Orhan
Journal:  Biomed Res Int       Date:  2015-04-16       Impact factor: 3.411

4.  2D FEA of evaluation of micromovements and stresses at bone-implant interface in immediately loaded tapered implants in the posterior maxilla.

Authors:  Shrikar R Desai; Rika Singh; I Karthikeyan
Journal:  J Indian Soc Periodontol       Date:  2013-09

5.  Successful Management of Peri-Implantitis around Short and Ultrashort Single-Crown Implants: A Case Series with a 3-Year Follow-Up.

Authors:  Giorgio Lombardo; Mauro Marincola; Andrea Cicconetti; Miguel Angel Simancas-Pallares; Jacopo Pighi; Jeffrey Lehrberg; Annarita Signoriello; Giovanni Corrocher; Xiomara Serpa-Romero; Luis Armando Vila Sierra; Luisa Arevalo-Tovar; Pier Francesco Nocini
Journal:  Int J Dent       Date:  2019-09-15

6.  Does the Length of Mini Dental Implants Affect Their Resistance to Failure by Overloading?

Authors:  Rafif Alshenaiber; Nick Silikas; Craig Barclay
Journal:  Dent J (Basel)       Date:  2022-07-01

7.  Short Dental Implants (≤8.5 mm) versus Standard Dental Implants (≥10 mm): A One-Year Post-Loading Prospective Observational Study.

Authors:  Guillermo Pardo-Zamora; Antonio José Ortiz-Ruíz; Fabio Camacho-Alonso; José Francisco Martínez-Marco; Juan Manuel Molina-González; Núria Piqué-Clusella; Ascensión Vicente-Hernández
Journal:  Int J Environ Res Public Health       Date:  2021-05-26       Impact factor: 3.390

8.  Increased crown-to-implant ratio may not be a risk factor for dental implant failure under appropriate plaque control.

Authors:  Shinsuke Okada; Katsunori Koretake; Yasunari Miyamoto; Hiroshi Oue; Yasumasa Akagawa
Journal:  PLoS One       Date:  2013-05-30       Impact factor: 3.240

9.  Peri-implant crevicular fluid levels of cathepsin-K, RANKL, and OPG around standard, short, and mini dental implants after prosthodontic loading.

Authors:  Raif Alan; İsmail Marakoğlu; Seyfullah Haliloğlu
Journal:  J Periodontal Implant Sci       Date:  2015-10-26       Impact factor: 2.614

10.  A retrospective clinical study of single short implants (less than 8 mm) in posterior edentulous areas.

Authors:  Sang-Yun Kim; Jeong-Kui Ku; Hyun-Suk Kim; Pil-Young Yun; Young-Kyun Kim
Journal:  J Adv Prosthodont       Date:  2018-06-12       Impact factor: 1.904

  10 in total