Performance characteristics of a new 3-dimensional continuous-emission and spiral-transmission high-sensitivity and high-resolution PET camera evaluated with the NEMA NU 2-2001 standard.

Keiichi Matsumoto1, Keishi Kitamura, Tetsuro Mizuta, Kazumi Tanaka, Seiichi Yamamoto, Setsu Sakamoto, Yuji Nakamoto, Masaharu Amano, Kenya Murase, Michio Senda.   

Abstract

UNLABELLED: The SET-3000 G/X (clinical tomograph with high resolution and a large axial field of view) is a 3-dimensional (3D) (only) dedicated PET camera with germanium oxyorthosilicate (GSO) and bismuth germanate (BGO) scintillators. The main characteristic of the SET-3000 G/X PET scanner is 3D continuous-emission and spiral-transmission (CEST) scanning, yielding a reduction in whole-body scan time. We evaluated the physical performance of the SET-3000 G/X PET scanner with the National Electrical Manufacturers Association (NEMA) NU 2-2001 standard.
METHODS: A GSO 3D emission scanner is combined with a BGO transmission scanner separated axially by a lead shield. In the GSO scanner, small and thick scintillators (2.45 x 5.1 x 30 mm(3)) are arranged in small blocks (23.1 x 52 mm) to achieve high resolution and a high counting rate. The detector ring has a large solid angle with a diameter of 664 mm and an axial coverage of 260 mm (50 rings). The transmission scanner consists of BGO block detectors with a diameter of 798 mm and an axial width of 23.1 mm and is equipped with a rotating (137)Cs point source of 740 MBq and a tungsten collimator. The low- and high-energy thresholds are set to 400 and 700 keV, respectively, in the emission system. The coincidence time window is set to 6 ns. In CEST acquisition, the patient couch moves continuously through the emission and transmission scanners in a 1-way motion. Emission coincidence data are acquired in the histogram mode with on-the-fly Fourier rebinning, and transmission single data are acquired with emission contamination correction.
RESULTS: With the NEMA NU 2-2001 standard, the main performance results were as follows: the average (radial and tangential) transverse and axial spatial resolutions (full width at half maximum) at 1 cm and at 10 cm off axis were 3.49 and 5.04 mm and 4.48 and 5.40 mm, respectively; the average sensitivity for the 2 radial positions (0 and 10 cm) was 20.71 cps/kBq; the scatter fraction was 50%; the peak noise equivalent count rate was 62.3 kcps at 9.8 kBq/mL; and the peak random rate was 542.1 kcps at 37.6 kBq/mL.
CONCLUSION: The new integrated SET-3000 G/X PET scanner has good overall performance, including high resolution and sensitivity, and has the potential of reducing whole-body acquisition time to less than 10 min while improving small-lesion detectability with a low radiation dose.

Entities:  

Mesh:

Year:  2006        PMID: 16391191

Source DB:  PubMed          Journal:  J Nucl Med        ISSN: 0161-5505            Impact factor:   10.057


  23 in total

Review 1.  [Molecular imaging with new PET tracers].

Authors:  A J Beer; M Schwaiger
Journal:  Radiologe       Date:  2007-01       Impact factor: 0.635

2.  NEC density and liver ROI S/N ratio for image quality control of whole-body FDG-PET scans: comparison with visual assessment.

Authors:  Tetsuro Mizuta; Michio Senda; Terue Okamura; Keishi Kitamura; Yuichi Inaoka; Munehiro Takahashi; Keiichi Matsumoto; Makoto Abe; Yoshihiro Shimonishi; Susumu Shiomi
Journal:  Mol Imaging Biol       Date:  2009-03-28       Impact factor: 3.488

3.  Superiority illusion arises from resting-state brain networks modulated by dopamine.

Authors:  Makiko Yamada; Lucina Q Uddin; Hidehiko Takahashi; Yasuyuki Kimura; Keisuke Takahata; Ririko Kousa; Yoko Ikoma; Yoko Eguchi; Harumasa Takano; Hiroshi Ito; Makoto Higuchi; Tetsuya Suhara
Journal:  Proc Natl Acad Sci U S A       Date:  2013-02-25       Impact factor: 11.205

Review 4.  Radiolabelled RGD peptides for imaging and therapy.

Authors:  F C Gaertner; H Kessler; H-J Wester; M Schwaiger; A J Beer
Journal:  Eur J Nucl Med Mol Imaging       Date:  2012-02       Impact factor: 9.236

5.  Cerebral serotonin transporter measurements with [11C]DASB: A review on acquisition and preprocessing across 21 PET centres.

Authors:  Martin Nørgaard; Melanie Ganz; Claus Svarer; Ling Feng; Masanori Ichise; Rupert Lanzenberger; Mark Lubberink; Ramin V Parsey; Marios Politis; Eugenii A Rabiner; Mark Slifstein; Vesna Sossi; Tetsuya Suhara; Peter S Talbot; Federico Turkheimer; Stephen C Strother; Gitte M Knudsen
Journal:  J Cereb Blood Flow Metab       Date:  2018-04-13       Impact factor: 6.200

6.  Interindividual variations of cerebral blood flow, oxygen delivery, and metabolism in relation to hemoglobin concentration measured by positron emission tomography in humans.

Authors:  Masanobu Ibaraki; Yuki Shinohara; Kazuhiro Nakamura; Shuichi Miura; Fumiko Kinoshita; Toshibumi Kinoshita
Journal:  J Cereb Blood Flow Metab       Date:  2010-02-17       Impact factor: 6.200

7.  Cerebrospinal Fluid Clearance in Alzheimer Disease Measured with Dynamic PET.

Authors:  Mony J de Leon; Yi Li; Nobuyuki Okamura; Wai H Tsui; Les A Saint-Louis; Lidia Glodzik; Ricardo S Osorio; Juan Fortea; Tracy Butler; Elizabeth Pirraglia; Silvia Fossati; Hee-Jin Kim; Roxana O Carare; Maiken Nedergaard; Helene Benveniste; Henry Rusinek
Journal:  J Nucl Med       Date:  2017-03-16       Impact factor: 10.057

8.  Paradoxical reduction of cerebral blood flow after acetazolamide loading: a hemodynamic and metabolic study with (15)O PET.

Authors:  Tadashi Watabe; Eku Shimosegawa; Hiroki Kato; Kayako Isohashi; Mana Ishibashi; Mitsuaki Tatsumi; Kazuo Kitagawa; Toshiyuki Fujinaka; Toshiki Yoshimine; Jun Hatazawa
Journal:  Neurosci Bull       Date:  2014-08-06       Impact factor: 5.203

9.  Risk Factors for Predicting Occult Lymph Node Metastasis in Patients with Clinical Stage I Non-small Cell Lung Cancer Staged by Integrated Fluorodeoxyglucose Positron Emission Tomography/Computed Tomography.

Authors:  Kaoru Kaseda; Keisuke Asakura; Akio Kazama; Yukihiko Ozawa
Journal:  World J Surg       Date:  2016-12       Impact factor: 3.352

10.  Imaging of amyloid deposition in human brain using positron emission tomography and [18F]FACT: comparison with [11C]PIB.

Authors:  Hiroshi Ito; Hitoshi Shinotoh; Hitoshi Shimada; Michie Miyoshi; Kazuhiko Yanai; Nobuyuki Okamura; Harumasa Takano; Hidehiko Takahashi; Ryosuke Arakawa; Fumitoshi Kodaka; Maiko Ono; Yoko Eguchi; Makoto Higuchi; Toshimitsu Fukumura; Tetsuya Suhara
Journal:  Eur J Nucl Med Mol Imaging       Date:  2014-04       Impact factor: 9.236

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.