Literature DB >> 16386322

The EQUAL-ESTRO audit on geometric reconstruction techniques in brachytherapy.

Amélie Roué1, Ivaldo H Ferreira, Jan Van Dam, Hans Svensson, Jack L M Venselaar.   

Abstract

BACKGROUND AND
PURPOSE: A geometric check procedure of the reconstruction techniques used in brachytherapy treatment planning systems was developed by the EQUAL (European Quality Laboratory) Laboratory in the framework of the ESTRO's (European Society for Therapeutic Radiology and Oncology) project 'ESQUIRE' (Education Science and QUality assurance In Radiotherapy in Europe [Baumann M, Brada M. Towards equity in turbulent Europe ESTRO, European cooperation and the European Commission. Radiother Oncol 2005;75:251-2. Heeren G. The bright but ephemeral life of a rainbow. A chronical of seventeen years of intensive ESTRO-EU cooperation. Radiother Oncol 2005;75:253-7]) by the task group Braphyqs (Brachytherapy physics quality system). PATIENTS AND METHODS: The check is performed by using the so-called 'Baltas' phantom, mailed to the participating centres in order to check the local technique of geometric reconstruction used in dose calculation.
RESULTS: To validate the procedures, the check was first tested among the members of the Braphyqs Network. Since November 2002, the system is open to other centres. Until now 152 reconstructions have been checked. Eighty-six percent of the results were within an acceptance level after the first check. For the remaining 14%, a second check has been proposed. The results of the re-checks are in most cases within an acceptance level, except for 2% of the reconstructions.
CONCLUSIONS: The geometric check is available from the EQUAL Laboratory for all the brachytherapy centres. The decrease of the deviations observed between the two checks demonstrates the importance of this kind of external audit as some errors were revealed, which were not discovered before with techniques used in clinical quality control routines.

Entities:  

Mesh:

Year:  2005        PMID: 16386322     DOI: 10.1016/j.radonc.2005.12.004

Source DB:  PubMed          Journal:  Radiother Oncol        ISSN: 0167-8140            Impact factor:   6.280


  7 in total

Review 1.  Dosimetric audit in brachytherapy.

Authors:  A L Palmer; D A Bradley; A Nisbet
Journal:  Br J Radiol       Date:  2014-05-06       Impact factor: 3.039

2.  Technical feasibility of integrating 7 T anatomical MRI in image-guided radiotherapy of glioblastoma: a preparatory study.

Authors:  Inge Compter; Jurgen Peerlings; Daniëlle B P Eekers; Alida A Postma; Dimo Ivanov; Christopher J Wiggins; Pieter Kubben; Benno Küsters; Pieter Wesseling; Linda Ackermans; Olaf E M G Schijns; Philippe Lambin; Aswin L Hoffmann
Journal:  MAGMA       Date:  2016-03-30       Impact factor: 2.310

3.  Geometric error of cervical point A calculated through traditional reconstruction procedures for brachytherapy treatment.

Authors:  Liyun Chang; Sheng-Yow Ho; Shyh-An Yeh; Tsair-Fwu Lee; Pang-Yu Chen
Journal:  J Appl Clin Med Phys       Date:  2015-09-08       Impact factor: 2.102

4.  An innovative method to acquire the location of point A for cervical cancer treatment by HDR brachytherapy.

Authors:  Liyun Chang; Sheng-Yow Ho; Shyh-An Yeh; Tsair-Fwu Lee; Pang-Yu Chen
Journal:  J Appl Clin Med Phys       Date:  2016-11-08       Impact factor: 2.102

5.  Electromagnetic tracking (EMT) technology for improved treatment quality assurance in interstitial brachytherapy.

Authors:  Markus Kellermeier; Jens Herbolzheimer; Stephan Kreppner; Michael Lotter; Vratislav Strnad; Christoph Bert
Journal:  J Appl Clin Med Phys       Date:  2017-01       Impact factor: 2.102

6.  Independent assessment of source position for gynecological applicator in high-dose-rate brachytherapy.

Authors:  Hiroyuki Okamoto; Satoshi Nakamura; Shie Nishioka; Kotaro Iijima; Akihisa Wakita; Yukinao Abe; Naoki Tohyama; Shinji Kawamura; Toshiyuki Minemura; Jun Itami
Journal:  J Contemp Brachytherapy       Date:  2017-10-19

7.  End-to-end test for computed tomography-based high-dose-rate brachytherapy.

Authors:  Fabian Krause; Franziska Risske; Susann Bohn; Marc Delaperriere; Jürgen Dunst; Frank-André Siebert
Journal:  J Contemp Brachytherapy       Date:  2018-12-28
  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.