Literature DB >> 16385475

Kinetics of antigen binding to antibody microspots: strong limitation by mass transport to the surface.

Wlad Kusnezow1, Yana V Syagailo, Sven Rüffer, Konstantin Klenin, Walter Sebald, Jörg D Hoheisel, Christoph Gauer, Igor Goychuk.   

Abstract

It is well documented that diffusion has generally a strong effect on the binding kinetics in the microtiter plate immunoassays. However, a systematic quantitative experimental evaluation of the microspot kinetics is still missing in the literature. Our work aims at filling this important gap of knowledge on the example of antigen binding to antibody microspots. A mathematical model was derived within the framework of two-compartment model and applied to the quantitative analysis of the experimental data obtained for typical antibody microspot assays. A strong mass-transport dependence of the antigen-antibody microspot kinetics was identified to be one of the main restrictions of this new technology. The binding reactions are slowed down in the microspot immunoassays by several orders of magnitude as compared with the corresponding well-stirred bulk reactions. The task to relax the mass-transport limitations should thus be one of the most important issues in designing the antibody microarrays. These limitations notwithstanding, the detection range of more than five orders of magnitude and the high sensitivity in the low femtomolar range were experimentally achieved in our study, demonstrating thus an enormous potential of this highly capable technology.

Mesh:

Substances:

Year:  2006        PMID: 16385475     DOI: 10.1002/pmic.200500149

Source DB:  PubMed          Journal:  Proteomics        ISSN: 1615-9853            Impact factor:   3.984


  16 in total

1.  Pre-binding dynamic range and sensitivity enhancement for immuno-sensors using nanofluidic preconcentrator.

Authors:  Ying-Chih Wang; Jongyoon Han
Journal:  Lab Chip       Date:  2008-01-14       Impact factor: 6.799

2.  Resonant Cavity Imaging: A Means Toward High-Throughput Label-Free Protein Detection.

Authors:  David A Bergstein; Emre Ozkumur; Arthur C Wu; Ayça Yalçin; Jeremy R Colson; James W Needham; Rostem J Irani; Jonathan M Gershoni; Bennett B Goldberg; Charles Delisi; Michael F Ruane; M Selim Unlü
Journal:  IEEE J Sel Top Quantum Electron       Date:  2008       Impact factor: 4.544

3.  Device considerations for development of conductance-based biosensors.

Authors:  Kangho Lee; Pradeep R Nair; Adina Scott; Muhammad A Alam; David B Janes
Journal:  J Appl Phys       Date:  2009-05-19       Impact factor: 2.546

4.  Accelerated immunoassays based on magnetic particle dynamics in a rotating capillary tube with stationary magnetic field.

Authors:  Jun-Tae Lee; L Sudheendra; Ian M Kennedy
Journal:  Anal Chem       Date:  2012-09-12       Impact factor: 6.986

5.  Evaluation of surface chemistries for antibody microarrays.

Authors:  Shannon L Seurynck-Servoss; Amanda M White; Cheryl L Baird; Karin D Rodland; Richard C Zangar
Journal:  Anal Biochem       Date:  2007-07-25       Impact factor: 3.365

6.  Mass-transport limitations in spot-based microarrays.

Authors:  Ming Zhao; Xuefeng Wang; David Nolte
Journal:  Biomed Opt Express       Date:  2010-09-20       Impact factor: 3.732

7.  Eliminating Size-Associated Diffusion Constraints for Rapid On-Surface Bioassays with Nanoparticle Probes.

Authors:  Junwei Li; Pavel Zrazhevskiy; Xiaohu Gao
Journal:  Small       Date:  2016-01-08       Impact factor: 13.281

Review 8.  Facing current quantification challenges in protein microarrays.

Authors:  Robert Wellhausen; Harald Seitz
Journal:  J Biomed Biotechnol       Date:  2012-04-24

9.  Microfluidic processor allows rapid HER2 immunohistochemistry of breast carcinomas and significantly reduces ambiguous (2+) read-outs.

Authors:  Ata Tuna Ciftlik; Hans-Anton Lehr; Martin A M Gijs
Journal:  Proc Natl Acad Sci U S A       Date:  2013-03-11       Impact factor: 11.205

Review 10.  Validation processes of protein biomarkers in serum--a cross platform comparison.

Authors:  Katja Köhler; Harald Seitz
Journal:  Sensors (Basel)       Date:  2012-09-18       Impact factor: 3.576

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.