| Literature DB >> 16384748 |
Lourdes Bosch1, Amparo Alegría, Rosaura Farré.
Abstract
The validation of a pre-column derivatization procedure with 6-aminoquinolyl-N-hydroxysuccinimidyl carbamate (AQC) to the determination of the amino acid content by RP-HPLC with fluorescence detection (lambda excitation 250 nm, lambda emission 395 nm) in milk-cereal based infant foods was carried out. The analytical parameters: linearity (0.0025-0.2mM), precision of the method (0.2-3.5% variation coefficients), accuracy (derivatization: 86-106% average recovery and method: 88.3-118.2% average recovery) and the limits of detection (0.016-0.367 microM) and quantification (0.044-1.073 microM) were determined. Glutamic acid, proline and leucine were the most abundant amino acid whereas the lowest contents corresponded to tyrosine and cysteine.Entities:
Mesh:
Substances:
Year: 2005 PMID: 16384748 DOI: 10.1016/j.jchromb.2005.12.002
Source DB: PubMed Journal: J Chromatogr B Analyt Technol Biomed Life Sci ISSN: 1570-0232 Impact factor: 3.205