Residual air in the venous cannula increases cerebral embolization at the onset of cardiopulmonary bypass.

Rosendo A Rodriguez1, Fraser Rubens, Dean Belway, Howard J Nathan.   

Abstract

OBJECTIVE: When the right atrium (RA) cannula is connected to the venous return line of the cardiopulmonary bypass (CPB) circuit, air is often introduced. Air in the venous cannula may increase cerebral air embolization at initiation of CPB despite the arterial line filter. We measured the volume of air present in the venous cannula after cannulation of the RA. Transcranial Doppler quantified emboli as high-intensity transient-signals (HITS) in both middle-cerebral arteries (MCA) at the beginning of CPB.
METHODS: After RA cannulation, the air column in the venous line was measured and the total volume calculated using the known lumen diameter. CPB onset was defined as the instant when the CPB machine started moving the patient's blood from the RA into the venous reservoir. Starting from CPB onset, HITS were counted: (a) until completion of the first minute on CPB (1-min count) and (b) until aortic cross clamping (pre-clamping count).
RESULTS: We studied 135 patients during coronary artery bypass surgery operated on by 10 cardiac surgeons. HITS during onset of CPB were detected in 95% of patients. Median counts were 10 HITS (25th, 75th percentiles: 3, 26) at 1-min and 21 HITS (8, 51) during pre-clamping. A significant correlation was found between the volume of air in the venous cannula and the HITS counts (r=0.524, p<0.0001). Absence of retained air was associated with lower HITS counts [3 HITS (1, 11)] compared with any amount of air [13 HITS (4, 29), p=0.002)]. The volume of air in the venous cannula, the MCA mean blood flow velocity and the pre-clamping time were the only independent predictors of the pre-clamping HITS counts (p<0.001).
CONCLUSION: Air in the venous cannula can result in HITS in the MCA. Minimizing the volume of air introduced into the venous cannula after cannulation of the RA can decrease cerebral air embolization at the beginning of CPB.

Entities:  

Mesh:

Year:  2006        PMID: 16376562     DOI: 10.1016/j.ejcts.2005.11.006

Source DB:  PubMed          Journal:  Eur J Cardiothorac Surg        ISSN: 1010-7940            Impact factor:   4.191


  6 in total

1.  Microbubble generation and transmission of medtronic's affinity hardshell venous reservoir and collapsible venous reservoir bag: an in-vitro comparison.

Authors:  Kieron C Potger; Darryl McMillan; Mark Ambrose
Journal:  J Extra Corpor Technol       Date:  2011-09

2.  In vitro evaluation of the air separation ability of four cardiovascular manufacturer extracorporeal circuit designs.

Authors:  Timothy A Dickinson; Jeffrey B Riley; Jeffrey C Crowley; Paul M Zabetakis
Journal:  J Extra Corpor Technol       Date:  2006-09

Review 3.  Vacuum-assisted venous drainage and gaseous microemboli in cardiopulmonary bypass.

Authors:  Shigang Wang; Akif Undar
Journal:  J Extra Corpor Technol       Date:  2008-12

4.  Gaseous microemboli in a pediatric bypass circuit with an unprimed venous line: an in vitro study.

Authors:  Andrea Hudacko; Alicia Sievert; Joseph Sistino
Journal:  J Extra Corpor Technol       Date:  2009-09

5.  Biofeedback: making the science real.

Authors:  Robert A Baker; Richard F Newland; Jayme Bennetts
Journal:  J Extra Corpor Technol       Date:  2009-12

6.  Vacuum-assisted venous drainage, angel or demon: PRO?

Authors:  Yves Durandy
Journal:  J Extra Corpor Technol       Date:  2013-06
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.