Frank G Shellock1, Samuel Valencerina.
Abstract
Specialized implants are used for transcatheter closure of septal defects, including atrial and ventricular septal defects, and patent foramen ovale. These metallic devices may pose a risk to patients undergoing magnetic resonance imaging (MRI) procedures especially if performed at 3 T. Therefore, this investigation evaluated MRI safety at 3 T for septal repair implants (CardioSEAL Septal Repair Implant and STARFlex Septal Repair Implant, NMT Medical, Boston, MA, USA) by characterizing magnetic field interactions, heating and artifacts. These implants exhibited minor magnetic field interactions; heating was not excessive (+0.5 degrees C); and artifacts will only create a problem if the area of interest is in the same area as or near these devices. Thus, the findings indicated that it would be safe for a patient with these implants to undergo MRI at 3 T or lower. Importantly, because of the minor magnetic field interactions, MRI may be performed immediately after implantation.Entities:
Mesh:
Substances:
Year: 2005 PMID: 16376188 DOI: 10.1016/j.mri.2005.10.010
Source DB: PubMed Journal: Magn Reson Imaging ISSN: 0730-725X Impact factor: 2.546