Septal repair implants: evaluation of magnetic resonance imaging safety at 3 T.

Frank G Shellock1, Samuel Valencerina.   

Abstract

Specialized implants are used for transcatheter closure of septal defects, including atrial and ventricular septal defects, and patent foramen ovale. These metallic devices may pose a risk to patients undergoing magnetic resonance imaging (MRI) procedures especially if performed at 3 T. Therefore, this investigation evaluated MRI safety at 3 T for septal repair implants (CardioSEAL Septal Repair Implant and STARFlex Septal Repair Implant, NMT Medical, Boston, MA, USA) by characterizing magnetic field interactions, heating and artifacts. These implants exhibited minor magnetic field interactions; heating was not excessive (+0.5 degrees C); and artifacts will only create a problem if the area of interest is in the same area as or near these devices. Thus, the findings indicated that it would be safe for a patient with these implants to undergo MRI at 3 T or lower. Importantly, because of the minor magnetic field interactions, MRI may be performed immediately after implantation.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 16376188     DOI: 10.1016/j.mri.2005.10.010

Source DB:  PubMed          Journal:  Magn Reson Imaging        ISSN: 0730-725X            Impact factor:   2.546


  4 in total

1.  Optimized protocols for cardiac magnetic resonance imaging in patients with thoracic metallic implants.

Authors:  Laura J Olivieri; Russell R Cross; Kendall E O'Brien; Kanishka Ratnayaka; Michael S Hansen
Journal:  Pediatr Radiol       Date:  2015-06-04

2.  3-T MRI safety assessments of magnetic dental attachments and castable magnetic alloys.

Authors:  M Hasegawa; K Miyata; Y Abe; T Ishii; T Ishigami; K Ohtani; E Nagai; T Ohyama; Y Umekawa; S Nakabayashi
Journal:  Dentomaxillofac Radiol       Date:  2015-03-18       Impact factor: 2.419

Review 3.  MRI in patients with cardiac devices.

Authors:  Edward T Martin; David A Sandler
Journal:  Curr Cardiol Rep       Date:  2007-03       Impact factor: 2.931

Review 4.  MR fluoroscopy in vascular and cardiac interventions (review).

Authors:  Maythem Saeed; Steve W Hetts; Joey English; Mark Wilson
Journal:  Int J Cardiovasc Imaging       Date:  2011-02-26       Impact factor: 2.357

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.