Literature DB >> 16376100

Use of an electrothermal bipolar tissue sealing system in lung surgery.

Mario Santini1, Giovanni Vicidomini, Alfonso Baldi, Giuseppe Gallo, Paolo Laperuta, Luigi Busiello, Maria Pia Di Marino, Vincenzo Pastore.   

Abstract

OBJECTIVE: We evaluated the validity of an electrothermal bipolar tissue sealing system (LigaSure, Valleylab Inc., USA) in lung surgery.
METHODS: Our study was divided into two phases. EXPERIMENTAL: We performed sutures of pulmonary vessels and bronchi and lung wedge resections by LigaSure in 28 lungs of adult pigs; subsequently, we quantitated the sealing capacity of the system detecting the burst pressure for each anatomical structure. Clinical: LigaSure was used in 36 patients undergoing lung surgery. We performed 23 thoracotomic procedures in 16 patients (19 wedge resections, 2 segmentectomies, and 2 fissure separations), and 20 thoracoscopic procedures (13 wedge resections, 5 bullectomies, and 2 adherence dissections). EXPERIMENTAL: Bronchi and vessels were divided into seven groups (diameter: 1-7 mm); 10 burst pressure measurements for each group were performed. A total of 84 wedge resections were performed; lung specimens were divided into seven groups (weight: 0.2-1.4 g). The percentage of bronchial sutures resistant to the pneumatic critical pressure (60 mmHg) was 100% in the 1-mm and 2-mm groups. No bronchi with 6-mm or 7-mm diameter reached the critical pressure. All pulmonary vessel sutures were resistant to the critical hydrostatic pressure (150 mmHg). The average burst pressure of wedge resection margins was higher than the critical pressure, and the percentage of suture margins resistant to the critical pressure decreased from 95% (0.2-g group) to 68% (1.4-g group). Histology confirmed the sealing of vessels, with a mean depth of thermal injury limited to 1.1mm. Clinical: In all patients, hemostasis obtained by LigaSure was effective, with minimal perioperative bleeding. The mean operating time was 77.2 min (range: 60-97) for thoracotomies and 60.3 min (range: 46-80) for thoracoscopies. The mean drainage duration was 3.1 days (range: 1-8). Two patients had prolonged air leaks (>7 days). The mean postoperative stay was 7.3 days (range: 5-13) for thoracotomies and 4.6 days (range: 1-6) for thoracoscopies.
CONCLUSIONS: Use of LigaSure in lung surgery appears feasible and easy. It provides satisfactory hemostasis and air-leak prevention; results are comparable to those of stapling devices, but this system seems to have a better benefit/cost ratio. Larger series are needed to confirm these data.

Entities:  

Mesh:

Year:  2006        PMID: 16376100     DOI: 10.1016/j.ejcts.2005.11.017

Source DB:  PubMed          Journal:  Eur J Cardiothorac Surg        ISSN: 1010-7940            Impact factor:   4.191


  14 in total

Review 1.  Video-assisted thoracoscopic surgery for complex mediastinal mass resections.

Authors:  M Blair Marshall; Lorenzo DeMarchi; Dominic A Emerson; Matthew L Holzner
Journal:  Ann Cardiothorac Surg       Date:  2015-11

2.  Exploring new technologies to facilitate laparoscopic surgery: creating intestinal anastomoses without sutures or staples, using a radio-frequency-energy-driven bipolar fusion device.

Authors:  J F Smulders; I H J T de Hingh; J Stavast; J J Jackimowicz
Journal:  Surg Endosc       Date:  2007-05-19       Impact factor: 4.584

3.  Use of the LigaSure device and the Stapler for closure of the small bowel: a comparative ex vivo study.

Authors:  Mario Santini; Alfonso Fiorelli; Gaetana Messina; Paolo Laperuta; Antonio Mazzella; Marina Accardo
Journal:  Surg Today       Date:  2012-09-15       Impact factor: 2.549

4.  Use of a vessel sealing system versus conventional electrocautery for lung parenchymal resection: a comparison of the clinicopathological outcomes in porcine lungs.

Authors:  Seiichiro Sugimoto; Shinichi Toyooka; Norichika Iga; Masashi Furukawa; Ryujiro Sugimoto; Kazuhiko Shien; Hitoshi Nishikawa; Junichi Soh; Masaomi Yamane; Takahiro Oto; Shinichiro Miyoshi
Journal:  Surg Today       Date:  2013-03-15       Impact factor: 2.549

5.  Bipolar sealing of lung parenchyma: tests in an ex vivo model.

Authors:  A Kirschbaum; A Clemens; T Steinfeldt; A Pehl; C Meyer; D K Bartsch
Journal:  Surg Endosc       Date:  2014-07-16       Impact factor: 4.584

6.  Efficacy and Safety of Lateral Approach-Video Endoscopic Inguinal Lymphadenectomy (L-VEIL) over Open Inguinal Block Dissection: a Retrospective Study.

Authors:  Sandeep P Nayak; Harshwardhan Pokharkar; Jaiprakash Gurawalia; Kapil Dev; Srinivas Chanduri; M Vijayakumar
Journal:  Indian J Surg Oncol       Date:  2019-06-14

Review 7.  Bipolar sealing devices in video-assisted thoracic surgery.

Authors:  Filippo Longo; Pierfilippo Crucitti; Fabio Quintarelli; Raffaele Rocco; Giuseppe Mangiameli; Gaetano Rocco
Journal:  J Vis Surg       Date:  2017-02-13

8.  Pitfalls related to the use of endostaplers during video-assisted thoracic surgery.

Authors:  Dominique Gossot; Geoni Merlusca; Alexandre Tudor; Mohamed Sadok Boudaya; Costin Radu; Pierre Magdeleinat
Journal:  Surg Endosc       Date:  2008-03-06       Impact factor: 4.584

Review 9.  Non-vascular experimental and clinical applications of advanced bipolar radiofrequency thermofusion technology in the thorax and abdomen: a systematic review.

Authors:  Shobhit Arya; Hugh Mackenzie; George B Hanna
Journal:  Surg Endosc       Date:  2014-11-19       Impact factor: 4.584

10.  (Video Assisted) thoracoscopic surgery: Getting started.

Authors:  Tamas F Molnar
Journal:  J Minim Access Surg       Date:  2007-10       Impact factor: 1.407