Abstract
Cotinine biomarker measurements involving both smokers and nonsmokers must accommodate a broad range of concentrations. Thus, we have routinely preclassified unknown samples as being either "high" or "low" by using an enzyme-linked immunoassay for cotinine prior to analysis by tandem mass spectrometry (MS). Although this method is effective, it is also time-consuming and complex; a simpler and faster approach would be useful. Consequently, a screening assay for urine cotinine using an immunochromatographic test strip (NicAlert) followed by a computerized analysis of the data was examined as a possible alternative. The results indicate that this approach can provide useful classification efficiency when using our target cutoff value of approximately 20 ng/mL. In the analysis of 50 urine samples from nonsmokers with varying degrees of exposure to environmental tobacco smoke, the classification sensitivity and specificity were 88% and 92%, respectively, for cotinine measured by the test strips relative to total cotinine concentrations measured by atmospheric-pressure ionization tandem MS. However, the relatively high cost of the strips may be a limiting factor.Entities:
Mesh:
Substances:
Year: 2005 PMID: 16374940 DOI: 10.1093/jat/29.8.814
Source DB: PubMed Journal: J Anal Toxicol ISSN: 0146-4760 Impact factor: 3.367