The effect of wear on ultrasonic scaler tip displacement amplitude.

Simon C Lea1, Gabriel Landini, A Damien Walmsley.   

Abstract

AIM: During clinical usage, scaler tips may become worn and reduced in length. It is unknown what effect wear has on the magnitude of scaler tip vibrations when they are utilized under typical clinical loads. The aim of this investigation was to assess the effect of simulated wear on ultrasonic scaler tip displacement amplitude, using a scanning laser vibrometer.
MATERIALS AND METHODS: A Cavitron SPS (Dentsply) ultrasonic generator and three scaler insert designs (FSI-100, FSI-1000 and FSI-SLI-10S) were selected for the investigation. Tip vibration displacement amplitude was assessed unloaded and then contacting against tooth surfaces with loads of 0.5 and 1.0 N. Tips were then ground down by 1 mm and then 2 mm and scans were repeated.
RESULTS: For all tips, load and length were found to be significant variables (p<0.0001). The scaler tips showed a fall in displacement amplitude with a reduction in tip length. However, all scaler tips showed variability in the amount oscillation that occurred. This was most pronounced with FSI-SLI-10S.
CONCLUSIONS: This investigation demonstrated that tip wear could affect the performance of dental ultrasonic scaler inserts by reducing their vibration displacement amplitude. Clinicians should be aware of this variability, which may be significant enough to affect clinical procedures.

Entities:  

Mesh:

Year:  2006        PMID: 16367854     DOI: 10.1111/j.1600-051X.2005.00861.x

Source DB:  PubMed          Journal:  J Clin Periodontol        ISSN: 0303-6979            Impact factor:   8.728


  7 in total

1.  Static biofilm removal around ultrasonic tips in vitro.

Authors:  Thomas Thurnheer; Elodie Rohrer; Georgios N Belibasakis; Thomas Attin; Patrick R Schmidlin
Journal:  Clin Oral Investig       Date:  2013-12-08       Impact factor: 3.573

2.  Potential of shock waves to remove calculus and biofilm.

Authors:  Philipp Müller; Bernhard Guggenheim; Thomas Attin; Ernst Marlinghaus; Patrick R Schmidlin
Journal:  Clin Oral Investig       Date:  2010-09-07       Impact factor: 3.573

3.  Cleaning ability and induced dentin loss of a magnetostrictive ultrasonic instrument at different power settings.

Authors:  Kathrin Lampe Bless; Beatrice Sener; Jürg Dual; Thomas Attin; Patrick R Schmidlin
Journal:  Clin Oral Investig       Date:  2010-02-03       Impact factor: 3.573

4.  Instrumentation during the second stage of periodontal therapy: a European survey.

Authors:  Isabelle Laleman; Laurence Seidel; Gilles Gagnot; Michèle Reners; France Lambert
Journal:  Clin Oral Investig       Date:  2022-03-19       Impact factor: 3.606

5.  Application of ultrasound in periodontics: Part II.

Authors:  Vivek K Bains; Ranjana Mohan; Rhythm Bains
Journal:  J Indian Soc Periodontol       Date:  2008-09

6.  Mechanical Properties and Wear Resistance of Commercial Stainless Steel Used in Dental Instruments.

Authors:  Hye-Bin Go; Jae-Yun Bang; Kyoung-Nam Kim; Kwang-Mahn Kim; Jae-Sung Kwon
Journal:  Materials (Basel)       Date:  2021-02-09       Impact factor: 3.623

7.  Influence of scaler tip design on root surface roughness, tooth substance loss and patients' pain perception: an in vitro and a randomised clinical trial.

Authors:  Nur Ayman Abdul Hayei; Noor Azlin Yahya; Syarida Hasnur Safii; Roslan Saub; Rathna Devi Vaithilingam; Nor Adinar Baharuddin
Journal:  BMC Oral Health       Date:  2021-03-31       Impact factor: 2.757

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.