Literature DB >> 16326745

Anticoagulant-free Genius haemodialysis using low molecular weight heparin-coated circuits.

Rolf Dario Frank1, Ute Müller, Regina Lanzmich, Christian Groeger, Jürgen Floege.   

Abstract

BACKGROUND: Regional citrate anticoagulation or saline flushes are often used in haemodialysis patients at high risk of bleeding. In an alternative approach we evaluated the effects of covalent circuit coating with low molecular weight heparin (LMWH) for intermittent haemodialysis.
METHODS: In vitro, we compared the thrombogenicity of an uncoated polyvinylchloride (PVC) tubing set with LMWH-coated tubing (AOThel) and a reference tubing with end-point attached heparin coating (Carmeda Bioactive surface) under dynamic blood contact. In vivo, five chronic haemodialysis patients were studied using the Genius dialysis system and F60S filters. Each patient underwent three dialysis sessions separated by a standard haemodialysis each: (1) standard dialysis (uncoated circuit and regular dalteparin dosage), (2) dialysis with LMWH-coated circuit and regular dalteparin dosage and (3) dialysis with a completely LMWH-coated circuit without anticoagulant use.
RESULTS: In vitro, both coated tubings showed significantly reduced thrombin-antithrombin (TAT) complex levels compared with PVC. The reference coating (Carmeda) released substantial antifactor Xa (antiXa) activity into the plasma. The LMWH coating (AOThel) released low antiXa activity only during the initial rinsing. In vivo, all dialysis sessions were well tolerated and completed without major clotting. Antithrombin levels and platelet counts were similar in all groups. P-selectin and D-dimer levels increased similarly in all groups. TAT levels were comparable in all groups during the first 3 h and significantly increased in the anticoagulant-free group after the fourth hour.
CONCLUSIONS: LMWH surface coating reduces thrombogenicity in vitro without releasing significant amounts of heparin from the surface. In vivo, anticoagulant-free haemodialysis using a completely LMWH-coated circuit is feasible and safe in stable chronic dialysis patients with normal coagulation.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 16326745     DOI: 10.1093/ndt/gfi293

Source DB:  PubMed          Journal:  Nephrol Dial Transplant        ISSN: 0931-0509            Impact factor:   5.992


  11 in total

Review 1.  Unfractionated heparin for hemodialysis: still the best option.

Authors:  Robert E Cronin; Robert F Reilly
Journal:  Semin Dial       Date:  2010 Sep-Oct       Impact factor: 3.455

Review 2.  What are the anticoagulation options for intermittent hemodialysis?

Authors:  Andrew Davenport
Journal:  Nat Rev Nephrol       Date:  2011-07-05       Impact factor: 28.314

3.  The effect of a polyurethane coating incorporating both a thrombin inhibitor and nitric oxide on hemocompatibility in extracorporeal circulation.

Authors:  Terry C Major; Elizabeth J Brisbois; Anna M Jones; Margaux E Zanetti; Gail M Annich; Robert H Bartlett; Hitesh Handa
Journal:  Biomaterials       Date:  2014-06-10       Impact factor: 12.479

4.  Subcutaneous Enoxaparin Safely Facilitates Bedside Sustained Low-Efficiency Hemodialysis in Hypercoagulopathic Coronavirus Disease 2019 Patients-A Proof-of-Principle Trial.

Authors:  Elke Neumann-Haefelin; Eugen Widmeier; Joachim Bansbach; Kai Kaufmann; Sebastian Heinrich; Gerd Walz; Hartmut Bürkle; Johannes Kalbhenn
Journal:  Crit Care Explor       Date:  2020-06-15

Review 5.  Development and hemocompatibility testing of nitric oxide releasing polymers using a rabbit model of thrombogenicity.

Authors:  Terry C Major; Hitesh Handa; Gail M Annich; Robert H Bartlett
Journal:  J Biomater Appl       Date:  2014-06-16       Impact factor: 2.646

Review 6.  Alternatives to standard unfractionated heparin for pediatric hemodialysis treatments.

Authors:  Andrew Davenport
Journal:  Pediatr Nephrol       Date:  2012-02-29       Impact factor: 3.714

7.  The Effects of the Combined Argatroban/Nitric Oxide-Releasing Polymer on Platelet Microparticle-Induced Thrombogenicity in Coated Extracorporeal Circuits.

Authors:  Tiffany R Bellomo; Mark A Jeakle; Mark E Meyerhoff; Robert H Bartlett; Terry C Major
Journal:  ASAIO J       Date:  2021-05-01       Impact factor: 3.826

Review 8.  Biopolymers for Biomedical and Pharmaceutical Applications: Recent Advances and Overview of Alginate Electrospinning.

Authors:  Jolanta Wróblewska-Krepsztul; Tomasz Rydzkowski; Iwona Michalska-Pożoga; Vijay Kumar Thakur
Journal:  Nanomaterials (Basel)       Date:  2019-03-10       Impact factor: 5.076

9.  Intermittent saline flushes or continuous saline infusion: what works better when heparin-free dialysis is recommended?

Authors:  Edward Zimbudzi
Journal:  Int J Nephrol Renovasc Dis       Date:  2013-04-15

10.  Hemodialysis without systemic anticoagulation: a prospective randomized trial to evaluate 3 strategies in patients at risk of bleeding.

Authors:  Bruno Guéry; Corinne Alberti; Aude Servais; Elarbi Harrami; Lynda Bererhi; Brigitte Zins; Malik Touam; Dominique Joly
Journal:  PLoS One       Date:  2014-05-13       Impact factor: 3.240

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.