Literature DB >> 16322920

Performance of user independent echocardiographic border detection algorithm: comparison with human observer variability.

Florence H Sheehan1, Edward A Geiser, Brad Munt, Catherine M Otto.   

Abstract

INTRODUCTION: We evaluated a method for autonomous, user-independent automated border delineation (ABD) developed by Geiser and Wilson, by comparing the accuracy of ABD relative to manual border tracing.
METHODS: Short axis echocardiographic images of 84 patients from 3 clinical sites were analyzed using ABD and by manual tracing performed by two observers at each site and two observers at a core laboratory. The centerline method was used to measure the distance between each pair of computer-generated and hand-traced borders. Cardiac parameters were also measured from all sets of borders: LV area, fractional area change, antero-posterior diameter, wall motion, and wall thickening.
RESULTS: The distance between computer-generated and hand-traced borders was slightly but significantly greater than human interobserver variability between the clinical sites and the core laboratory (0.34+/-0.25 (N = 328) vs. 0.26+/-0.16 (N = 320) cm for the endocardium at end diastole, p = 0.0001). Measurements of LV area and fractional area change were similar by ABD and manual tracing. Other cardiac parameters showed greater deviation between ABD and manually traced borders than between human observers.
CONCLUSION: Autonomous ABD provides accurate measurements of LV area and area-derived indices. However measurements dependent on border point location deviate more by ABD.

Entities:  

Mesh:

Year:  2005        PMID: 16322920     DOI: 10.1007/s10554-005-4520-9

Source DB:  PubMed          Journal:  Int J Cardiovasc Imaging        ISSN: 1569-5794            Impact factor:   2.357


  7 in total

1.  Estimation of left ventricular cavity area with an on-line, semiautomated echocardiographic edge detection system.

Authors:  B F Vandenberg; L S Rath; P Stuhlmuller; H E Melton; D J Skorton
Journal:  Circulation       Date:  1992-07       Impact factor: 29.690

2.  Quantitative analysis of regional wall thickening by transesophageal echocardiography.

Authors:  F H Sheehan; M P Feneley; N P DeBruijn; J S Rankin; J W Davis; E L Bolson; P S Glass; F M Clements
Journal:  J Thorac Cardiovasc Surg       Date:  1992-02       Impact factor: 5.209

3.  A multiple active contour model for cardiac boundary detection on echocardiographic sequences.

Authors:  V Chalana; D T Linker; D R Haynor; Y Kim
Journal:  IEEE Trans Med Imaging       Date:  1996       Impact factor: 10.048

4.  Autonomous epicardial and endocardial boundary detection in echocardiographic short-axis images.

Authors:  E A Geiser; D C Wilson; D X Wang; D A Conetta; J D Murphy; A D Hutson
Journal:  J Am Soc Echocardiogr       Date:  1998-04       Impact factor: 5.251

5.  Advantages and applications of the centerline method for characterizing regional ventricular function.

Authors:  F H Sheehan; E L Bolson; H T Dodge; D G Mathey; J Schofer; H W Woo
Journal:  Circulation       Date:  1986-08       Impact factor: 29.690

6.  Ultrasonic backscatter system for automated on-line endocardial boundary detection: evaluation by ultrafast computed tomography.

Authors:  R H Marcus; J Bednarz; R Coulden; S Shroff; M Lipton; R M Lang
Journal:  J Am Coll Cardiol       Date:  1993-09       Impact factor: 24.094

7.  Matched filter identification of left-ventricular endocardial borders in transesophageal echocardiograms.

Authors:  P R Detmer; G Bashein; R W Martin
Journal:  IEEE Trans Med Imaging       Date:  1990       Impact factor: 10.048

  7 in total
  2 in total

1.  Automatic analysis of echocardiographic studies--the final frontier?

Authors:  Johan G Bosch
Journal:  Int J Cardiovasc Imaging       Date:  2005-12       Impact factor: 2.357

2.  A novel algorithm for the morphometric assessment of radiotherapy treatment planning volumes.

Authors:  R Jena; N F Kirkby; K E Burton; A C F Hoole; L T Tan; N G Burnet
Journal:  Br J Radiol       Date:  2009-07-20       Impact factor: 3.039

  2 in total